40m QIL Cryo_Lab CTN SUS_Lab CAML OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  SUS Lab eLog, Page 29 of 38  Not logged in ELOG logo
ID Date Author Typeup Category Subject
  1767   Mon Jun 4 11:15:17 2018 gautamLab InfrastructureEquipmentLoanCrackle lab ---> 40m

Johannes and I have taken the following equipment from the crackle lab to the 40m between Friday, June 1, and Sunday, June 3 2018. 

  • 1 Lightwave M126N-1064-200 NPRO (S/N 259) + controller unit. Measured power output was 320mW.
  • 1 AcoustoOptic Modulator, box was labelled "TNI Isomet AOM 1205C-843"
  • 1 AcoustoOptic Modulator RF driver, box was labelled "AOM Modulator Driver Model 232A-1"
  • 1 broadband EOM (Unlabeled, but looks like the NewFocus)
  • 1 Faraday isolator (Unlabeled)

Apart from the NPRO which was taken from the optical bench, everything else was taken from the storage cabinet by the lab entrance.

  1856   Fri Nov 6 11:58:02 2020 KojiLab InfrastructureGeneralECDL Gratings / Lenses delivered. Powermeter lending from OMC Lab

The gratings and aspheric lenses glued on the mounts were delivered to the lab on Thu.

The powermeter controler + S401C head was lent from OMC Lab. Returned to OMC Jul 15, 2020 KA

Attachment 1: P_20201105_195258.jpg
P_20201105_195258.jpg
  1861   Wed Nov 11 12:55:18 2020 PacoLab InfrastructureGeneral1064 pump power curve

Log of the output power vs current in the 1064 nm (Innolight) pump laser. The crystal temperature was set to 45.5 C, and the current limit is set to 2.1 A

Attachment 1: 2020_11_10_mephisto_test.png
2020_11_10_mephisto_test.png
  1874   Fri Dec 11 16:04:36 2020 PacoLab InfrastructureOpticsPPKTP crystals

Two crystals from Raicol arrived. Picked them up from Downs today and inspected them (see photos below). The lengths are nominal (20 mm), they are serialized as 123 and 124, and the ends look like they have the specified (AR) coating. I reached out for Covesion two days ago to track the ovens so we can mount these guys, but have yet to hear back from them.

Attachment 1: raicol_124.jpg
raicol_124.jpg
Attachment 2: raicol_123.jpg
raicol_123.jpg
  1876   Tue Dec 15 13:00:41 2020 anchalLab InfrastructureEquipmentLoanReceived Marconi and Rb clock from CTN

Received one Marconi 2023A (#539) from CTN and an SRS FS725 Rb clock. (See CTN/2605)

  1880   Thu Dec 17 12:01:54 2020 PacoLab InfrastructureOpticscrystal ovens, clips and controllers

Covesion order arrived, containing 2x

  • Crystal oven (20 mm long) (below)
  • Clips (for mounting crystals) (below)
  • Blank crystal (to press on the ppktp crystal) (below)
  • OC2 oven controller
  • Controller cable and power cable
Attachment 1: covesion_oven_clip_blank.jpg
covesion_oven_clip_blank.jpg
  1893   Fri Jan 15 10:57:39 2021 PacoLab InfrastructureEquipmentLoanBorrow AG4395A and EOM Driver TF re-tune

See equipment borrowing note here.

Attempting TF measurement for resonant EOM driver, but not having luck reproducing the measurements done recently (Dec-03), so I started debugging the circuit. Both power supply connections (+- 18 VDC) seem nominal. The MAX2470 buffer regulated input is nominal at 5VDC. Looking at MMBT5551 HF transistor, base-emitter voltage is -0.60 VDC (nominal wrt -0.66 V). Using a scope, I feed a single tone (36 MHz, 190 mVpp) and look at the RFmon output and it looks ok (gain ~ 1). I changed the RFmon SMA cable and that seemed to do the trick... Bad cable (now in trash) stole my morning.

Tune EOM driver resonance to 35.993 MHz (shown below for reference).

Attachment 1: EOMDriver_TF_Inp_to_RFMon_CloseUp_15-01-2021_160402.pdf
EOMDriver_TF_Inp_to_RFMon_CloseUp_15-01-2021_160402.pdf
Attachment 2: EOMDriver_TF_Inp_to_RFMon_Wide_15-01-2021_151039.pdf
EOMDriver_TF_Inp_to_RFMon_Wide_15-01-2021_151039.pdf
  1902   Wed Feb 17 11:56:48 2021 PacoLab InfrastructureElectronicsUPDH box zero model and SR560 "lock"

UPDHv3 box (serial 17142) is bogus. While retrieving values of some of the components to plug into working zero model, saw the VGA stage is bypassed by a previously unnoticed hack. Verified this by taking TF and not seeing any changes with respect to the gain knob (shown below are zero's model TFs suggesting a tunable UGF from ~ 10 Hz to 1 kHz), so this box is not good for a standalone servo.

As suggested a few meetings ago, made a quick and dirty lock using a single SR560 and took measurement of something* CLTF (SR560 gain = 10) below. New goal is to find a decent replacement, for which decided to use RedPitaya's python API "pyRPL". Just using the GUI out of the box can also lock the cavity relatively quickly but neither method results in longer than 1 minute lock... so took one step back to polish the pdh error signal.

* Something = Use SR785 TF measurement with source on Ch1, and to B input in SR560. The SR560 in (A-B) mode, and demodulated signal connected to A. The loop was closed with the SR560 output driving the PZT, and Ch2 of SR785. Wouldn't call this CLTF...

Attachment 1: updhv3_VGA_gain.pdf
updhv3_VGA_gain.pdf
Attachment 2: SR560_OLTFSR785_17-02-2021_164500.pdf
SR560_OLTFSR785_17-02-2021_164500.pdf
  1907   Tue Mar 30 15:18:06 2021 PacoLab InfrastructureEquipmentLoan2um CCD borrowed

With Aidan's assistance, I borrowed

  • WiDy SWIR camera (Pembroke WiDy SWIR 640U-A) from the QIL (Attachment 1)
  • Heimann thermopile sensor (HTPA80x64d) with microUSB-A and ethernet cables from the Adaptive optics lab (Attachment 2)

for ~ 2 um imaging in the Crackle lab.

Attachment 1: IMG_20210330_151046.jpg
IMG_20210330_151046.jpg
Attachment 2: IMG_20210330_150958.jpg
IMG_20210330_150958.jpg
  1909   Tue May 18 10:28:50 2021 PacoLab InfrastructureEquipmentLoanheimann sensor update

Heimann (HTPA80x64d) thermopile array;

- First test to grab frames was done in my personal Win10 machine, with no success. Either I was unable to configure the server correctly, or the software "ArraySoft" is not supported in Win10. Upon contacting Heimann, I received instructions to update to a newer version but was warned that it's just a new GUI, nothing really changed from v1 --> v2. So didn't even bother.

- Instead, wrote a simple python-socket UDP server to catch the video stream. Most trouble happened when using temperature mode (command "K"). The client streams a bunch of zeros... My guess is that this unit does not have an internal temperature calibration, and one could in principle be uploaded but we probably don't care. Streaming in raw voltage mode (command "t") works well, as shown by the sample frame shown in Attachment 1.

- After recovering the CTN Win7 laptop from Radhika, I gave "ArraySoft" another change just to know the frames I was getting in python were not bogus. For this I pointed a 532 nm laser pointer straight to the sensor and got an image shown in Attachment 2. The key difference is the processing of the video stream. Attachment 1 is a single frame, while Attachment 2 is the average of 30 frames with no offsets present. 

- Another issue present during this task was a faulty USB connection. Sometimes moving the sensor around would interrupt the stream (power lost). I carefully removed the case and exposed the TO-39 package and surrounding electronics to inspect and search possible failures but after seeing none, I swaped the USB power cable with my portable battery charger and had a more robust operation... So I dumped the old USB cable, and will get a new one.

- Since this one was borrowed from TCS lab, I placed an order for another one which will be set up permanently in the lab. Hopefully this will be enough for the OSA.

Attachment 1: no_light.png
no_light.png
Attachment 2: light.png
light.png
  1913   Thu Jun 10 09:59:52 2021 PacoLab InfrastructureDOPODisassembly for new optical table

Today the DOPO v0 got disassembled to make way for the optical table swap. Most components have been stored in the white cabinet's bottom panel.

Attachment 1: IMG_20210610_092713.jpg
IMG_20210610_092713.jpg
  1929   Thu Jun 23 16:34:46 2022 PacoLab InfrastructureDOPORelocated DOPO setup

Following Koji's request, I took some time to clear the area surrounding the crackle chamber so it can be migrated to the former TCS lab.

I moved the DOPO setup which was sitting on a breadboard for easy transportation (Attachment #1) and placed into the other table in the lab. Attachments #2-3 shows the cleared area. Several instruments from the DOPO experiment still remain around the other side of the crackle chamber, if they need to be relocated I can move them as well.

Attachment 1: PXL_20220623_222426584.jpg
PXL_20220623_222426584.jpg
Attachment 2: PXL_20220623_223623414.jpg
PXL_20220623_223623414.jpg
Attachment 3: PXL_20220623_224259785.jpg
PXL_20220623_224259785.jpg
  1930   Tue Jan 10 23:40:17 2023 KojiLab InfrastructureGeneralHeavy item transport - preparation 

See http://nodus.ligo.caltech.edu:8080/Mariner/121

  1931   Thu Jan 12 11:51:49 2023 KojiLab InfrastructureGeneralHow to move the large engine hoist through the narrow door

How to move the large engine hoist through the narrow door

See http://nodus.ligo.caltech.edu:8080/Mariner/122

  1933   Mon Sep 25 15:56:17 2023 PacoLab Infrastructure1418 nm AUX ECDL2nd ECDL housing assembled

I assembled the second ECDL with a 19XX SAF gain chip (all pins remain disconnected), and a 14XX nm angle facet plate with a grating.

Attachments #1-4 show the final state of the housing along with the thorlabs current and temperature drivers. The remaining spare components (including screws, angle facet plates for the gratings and two SAF gain chips, one for each wavelength) are stored in a single box in the right top cabinet on the north end.

Attachment 1: PXL_20230925_223237259.MP.jpg
PXL_20230925_223237259.MP.jpg
Attachment 2: PXL_20230925_223622837.MP.jpg
PXL_20230925_223622837.MP.jpg
Attachment 3: PXL_20230925_222547746.MP.jpg
PXL_20230925_222547746.MP.jpg
Attachment 4: PXL_20230925_223541276.MP.jpg
PXL_20230925_223541276.MP.jpg
  1940   Sat Oct 28 22:46:55 2023 PacoLab InfrastructureEquipmentLoanThorlabs HV pzt driver

I took the Thorlabs PZT driver to the 40m for use with the BHD OMC.

  107   Mon Dec 17 19:39:52 2007 waldmanLaserOMCFiber seems to be broken
The 50 m fiber running from Rana's lab to 056 seems to be broken. I can't get any light through it to save my life. A 5 meter fiber couples like child's play. I think we should acquire a fiber coupler - then I will couple light into the 5 m fiber that works fine and couple it to the 50 m fiber and prove that its broken. Only then will I go pull the installed fiber from the 40m clean room.

sam
  108   Wed Dec 19 15:29:14 2007 DmassLaserGeneralTip Tilts Being Baked
The Tip Tilt mirrors are now being baked with Bob. They should be ready in time.

The traveler has DCC #E070354-00-X.

As per Bob's know-how, they are cooking at:
80 C for magnets
200 C for steel;
120 C for the Al, Cu, Bronze, and teflon

davidym@caltech.edu
  109   Wed Dec 19 23:09:43 2007 waldmanLaserOMCOMC relocked
The OMC has been relocked in preparation for final diode alignment, final QPD aligment, and adding the beam blocks. The mode matching is really terrible, so it makes alignment a little difficult because there is a high, high order mode close to the 00 that is making problems for the vertical alignment.

sam
  113   Thu Dec 20 19:12:11 2007 waldmanLaserOMCStressful reattachment of heater
Photos may follow eventually, but for now here's the rundown. I scraped the heater clean of the thermal epoxy using a clean razor blade. Then I stuffed a small piece of lint free cloth in the OTAS bore and wrapped the OMC in tin foil. With a vacuum sucking directly from the face of the OTAS, I gently scraped the glue off the OTAS aluminum. I wiped both the OTAS and the heater down with an isoproponal soaked lint-free cloth. I put a thin sheen of VacSeal on the face of the heater, wiping off the excess from the edges with a cloth. Then I clamped the heater to the OTAS using 2" c-clamps from the tombstone back to the heater front, making sure the alignment of the OTAS was correct (connector on the absolute bottom, concentric with the OTAS outer diameter). I added a second clamp, then beaded the outside of the joint with a little bit extra VacSeal, just for kicks. I'll leave it covered at least overnight, and maybe for a day or two.

sam
  114   Sat Dec 22 18:05:07 2007 waldmanLaserOMCHeater bonding (almost) successful
The heater is now (re)bonded to the OTAS using VacSeal instead of the thermal epoxy. Unlike the expensive high-quality epoxy, VacSeal has supported 4 cycles of connector on/off.

Unfortunately, I didn't bond the heater as straight onto the aluminum as I had hoped. The connector is still about 5 degrees from the vertical so that the connector body sticks up close to the beam (but not as close as it might be. The mode matching is sooooo terrible (because I used an unlabeled lens) that it is impossible to identify the little bit of scattered light coming off the connector. Some of it could be (and probably is), junk light. I believe the closest approach of the connector is ~3 mm, but in the event we measure the finesse to be low, we might want to consider reworking the connector to get it down out of the beam.
  115   Sun Dec 23 17:21:57 2007 waldmanLaserOMCQPD3 centering
In the current cabling scheme, QPD3 is the Far field QPD mounted to the breadboard. With the cavity on a "pretty good alignment" I can physically move the QPD and get the signals to go +/- 1. In other words, I can physically move the QPD to zero its signal, which is the whole purpose of this test. The QPD matrix looks like [1 1 1 1; 1 1 -1 -1; 1 -1 -1 1]; Pitch looks right, yaw might be backwards. Note that the QPD is in a "plus" orientation.

I can center QPD4, the "near field" QPD just fine as well. In this configuration.

One caveat going forward - This alignment is going to be a real pain. In order to be in the linear range of the QPD, we need to be +/- omega/10 or something. In other words, with our 500 micron waist size, we need to get within 50 microns. We should get a translation stage made for this.
  116   Sun Dec 23 18:40:12 2007 waldmanLaserOMCDCPD centering
I can use a CCD camera and a macro lens to see the transmitted cavity beam on the DC photodiodes. I don't know which camera b/c there is no name on it, but it is a square edged camera with a 1/4-20 thread on the mount and has a many pinned cable running to a PS-12SU power supply. I also don't know which macro lens, but tis C-mount, about 4 inches long by 1.5 inches diameter with zoom, focus and f-stop adjustments (all unmarked).

When I look at the face of the transmitted PD, I see the three pins behind the diode chip on the left side. I can move the PD such that the beam is centered on the diode, but this is all the way to one side of the radiator clearance holes. When I maximally misalign the diode, the beam overlaps the diode edge on the side of the feed thru pins.

For the reflected DCPD, I cannot center the beam. Viewing the face of the PD with the two visible feedthru pin son the left side of the diode chip, the beam is low to the left and I cannot center it. I will try again tomorrow with the radiator with the oversize thru holes and see if I can center it. I think it may be a good idea to use an oversized radiator on the transmitted DCPD as well.
  117   Wed Dec 26 15:46:22 2007 waldmanLaserOMCDCPD centering
I used the drilled out radiator with both DCPDs and confirmed that I could center the beams. I also found the reflected beams and confirmed that a) there is clearance for the beamdumps and b) the location of the beams. Note that by some freak of fate, the transmitted DCPD ghost beam is only at about +15mm above the breadboard. I should mount the beamdump a bit low to accomodate it.


Quote:
I can use a CCD camera and a macro lens to see the transmitted cavity beam on the DC photodiodes. I don't know which camera b/c there is no name on it, but it is a square edged camera with a 1/4-20 thread on the mount and has a many pinned cable running to a PS-12SU power supply. I also don't know which macro lens, but tis C-mount, about 4 inches long by 1.5 inches diameter with zoom, focus and f-stop adjustments (all unmarked).

When I look at the face of the transmitted PD, I see the three pins behind the diode chip on the left side. I can move the PD such that the beam is centered on the diode, but this is all the way to one side of the radiator clearance holes. When I maximally misalign the diode, the beam overlaps the diode edge on the side of the feed thru pins.

For the reflected DCPD, I cannot center the beam. Viewing the face of the PD with the two visible feedthru pin son the left side of the diode chip, the beam is low to the left and I cannot center it. I will try again tomorrow with the radiator with the oversize thru holes and see if I can center it. I think it may be a good idea to use an oversized radiator on the transmitted DCPD as well.
  118   Wed Dec 26 22:39:52 2007 waldmanLaserOMCDCPD beam dumps
I manufactured 2x 1" x 1" black glass beam dumps by vac-sealing the appropriate sized squares to DCPD tombstones. I then locked the cavity and looked at where the DCPD reflections went. The REFL DCPD is no problem, but the choices for the TRANS DCPD aren't so hot. As shown in the enclosed photos, the three options are ugly. I lean towards option 1, but the other two are possibilities as well.

To be clear,

Option 1: beam dump close to the PD at an angle to the beam splitter. Spot is centered on the dump and a little of the tombstone hangs off the edge

Option 2: beam dump close to the PD, parallel to the beam splitter. In order to get the spot on the black glass, the tombstone is very close to the beamsplitter. (This option might be more attractive if the tombstone is rotated 180 degrees.) The tombstone doesn't hang off the edge.

Option 3: We can catch the beam after the beam splitter if we hang the tombstone significantly off the edge, like by half. You'll see in the photos

www.ligo.caltech.edu/~swaldman/BeamdumpPics
  120   Fri Jan 25 22:08:45 2008 waldmanLaserOMCOMC ready for shipment
The OMC has been boxed in its spiffy black box and is ready for shipment.

We mounted the OMC in its frame, removing any loose mechanical parts and tightening up the nylon tipped set screws. Then we swaddled each optic in tin foil and lint free cloth before wrapping the whole thing in a double layer of the shiny stuff. Finally we wrapped the whole kaboodle in a double layer of ameristat before shoe-horning (literally) it into the box. The package is about 1/4" too long, so it required some effort to get it into the box, and will doubtless require similar effort to remove it.

The OMC in its box was left on top of its shipping container in 056.
  106   Mon Nov 26 16:05:43 2007 Norna RobertsonMiscSUSMass of OMC silca bench (measured 16 Nov 2007)
For the record here is a note sent round OMC-SUS colleagues on 16th November.

--------------------------------

A team of Chris, Chub, Helena and me ably led by Sam (who did the scary lifting) got the OMC bench dismounted from its suspension and weighed today. It is now sitting on its custom plate with dark side down (minus the preamp boxes) in the centre of the optics table in room 56. It will remain there for the forseeable future. The metal parts of the suspension will be disassembled by Chris with Ken Mailand's help starting Monday.

Results of weighing

1) Optical bench including preamps and the wire bundle to the heater, pzt etc = 6.514 kg
2)Preamp cable (mock-up) = 24.4 g (without the connectors to the preamp)
3) The two leaning towers of counterweights (which were situated close to the far corners of the mass - NOT above the holes where they need to be for fixing in place) = 200.2 g and 187.7 g

I remind you all that there are 2 more tombstones ( fused silica 8 mm thick x 30 mm tall x 32? mm wide with 18 mm diam hole- size to be checked from CAD drawing) and two pieces of black glass 1 inch by 1 inch by 3 mm ( I noted at our telecon that these weigh 8.5 gm total) to be added to the bench - situated close to the DC photodiodes which sit on bright side under the preamps.
  119   Wed Jan 23 22:52:21 2008 Norna RobertsonMiscSUSOMC SUS assembly adjustments made and lessons learnt
Calum and I have been working in the 40 m annex re-assembling and testing the OMC SUS after the parts have been cleaned and baked and various parts modified. We have had several problems which needed to be addressed and made several observations of items which need attention for the next OMC. Also we note several points which need to be followed when the OMC is being prepared at LLO for input into the vacuum chamber. The following are in no particular order.

1) Problem with one of the blades
When we got the double pendulum hanging today we noticed that one of the lower blades was interfering with part of the top mass - hitting a block which holds one of the screws for adjusting the position of the clamp for the upper wire ( for adjusting pitch). The blade was 13S - one of those which had an angled clamp of 3.5 degrees (the larger of the angles - the blades at the other end of the bench have 3 degree angles). These large angles are being used to counteract the fact that the bench weighs more than these blades were designed for. The blades thus have a convex curvature (looking from above) and the crown of the bend was where the interference was occuring. The puzzling bit is why only one blade showed this, and not its partner. We tried another 3.5 degree clamp ( in case it was a fault with the clamp) but saw the same effect. We then tried reducing to 3 degrees for that blade - still almost touching. We then reduced the angle for that blade to 2.5 degrees. This worked Ok - and when the masses were hanging again all the blade tips appeared to be at approximately the same height (measured using a steel rule). So the dynamics should be OK. Puzzle remains why only one blade showed this behaviour. Note that for the LHO OMC we will be getting new blades designed for the mass they will be taking, and so we should not need to use angled clamps as large as for the LLO suspension Hence this interference problem should not arise again.
Lesson for the future - need to characterise all blades with the full range of angled clamps to be used. (We did not have time to do this for this suspension).

2) Test of new method of hanging silica bench (using metal bench)
Yesterday with Ken Mailand's help we tried out the new way of suspending the bench using Ken's bench holder with one of the metal benches and the two lab jacks. We learnt that to get the right range of height adjustment to lift the bench to put the discs on the lower wire clamps and then lower the bench again the mating pieces on the lab jacks had to mate with the central portion of the handles at each end rather than the top part of the handles. Also it looks like the lower EQ stop holder (which sits under the bench) should be put in place before the bench is brought in - otherwise difficult to get into place. Finally we note that this job is a three person job - need one on each end to raise and lower the lab jacks and one to fit the discs and then guide the discs and clamps into the holes and check they are seated correctly as the bench is lowered.

3) Height of bench.
We measured the height of the lower surface of the metal bench above the optics table to be (160.5 - 38.75) = 121.75 mm. Requirement is 101.6 + 20 = 121.6 (+/- 2mm) so we are OK. This was without any extra mass added to the top mass. We note that the tabelcloth is close to lower end of its range, but should be OK. However the final alignment needs to be done when the OSEMs are in place.

3) EQ stops and parts which should be removed before putting into vacuum chamber.
For sending the OMC to LLO we put on double nuts on all the EQ stops. This limits the range of the stops and in particular the ones under the bench were not long enough to reach the bench when double nuts are used. So we had to improvise the position of the EQ holder by putting ~ 1/2 inch "shims" to raise the holder ( the shims were three of the masses made for attaching to the metal bench to increase its mass). We also removed all "loose" parts - the adjustment mechanisms for the top blade yaw positions, the magnets and flags, the set screws for locking the larger vertical stops used to hold the top mass, and the screws used to alter the pitch adjustment clamp positions.
When the OMC is reassembled at LLO and made ready for installation, all the second nuts on the EQ stops should be removed. Also once the pitch adjustment using the moveable clamps is set correctly, the screws for doing those adjustments should be removed again. Basically all loose screws should be removed.

4) Parts needing modifications
i) EQ cross pieces holding the plate which goes over the bench need to be reduced in length to fit between the structure legs.
ii) Blind holes in EQ corner brackets.

5) Point to note for next OMC.
i) The slots in the structure which take the dowel pins for alignment need to be lengthened to allow the tableloth the full range of movement which the slots for the attachment screws would allow.
ii) The targets for aligning OSEMS need a hole in the centre for the flag.

More to follow tomorrow
Also we took lots of pictures today. will put relevant ones into installation document.

OMC is due to be crated tomorrow for pick-up on Friday am.
  121   Thu Feb 28 13:46:16 2008 David YMMiscOMCTip Tilts - Susp. Wire shipped
As per Adam Mullavey's request, I have shipped the remainder of the (cleaned) suspension wire for the tip tilt assemblies to LLO for the out of vacuum tip tilts. Unclean the wire as necessary, as Bob says the bake job is trivial for it.
-Dmass
  122   Tue Nov 17 10:33:37 2009 ZachMiscSUSAOSEM test progress

 It's dusty in here...

--------------------------------

 

I was recently commissioned to do some noise measurements on the new  AOSEMS. I set up a humble experiment in the LIGO e-lab to do some preliminary measurements:

 

I made a simple current-to-voltage converter out of an OP27E (using a 100-kohm feedback resistor) to use as the transimpedance amplifier for the readout. This results in a transimpedance of 0.1 V / uA. A simple schematic of the important elements is attached below.

 

DC power was provided without regulators directly from the laboratory DC supply in the lab. The value of 1.7 V across the LED was set such that the current through it was ~35 mA.

 

Rana and I took a few important PSDs (one of the DC supply, one of the OP27E with no supplied current, and two with the setup fully connected--one each with and without the PD covered), all from 250 mHz - 200 Hz, AC coupled. Using a sophisticated estimation method (called, by some, the "pick two points and approximate with a power law" method for lack of something fittingly elegant), we obtained a rough estimate of these spectral densities in order to compare them.

 

These were all converted into equivalent PD current noise. For all but the "supply" noise, this was done trivially by dividing by the transimpedance of the OP27E. For "supply", LED voltage noise had to be converted to PD current noise in the following way:

 

Z_LED = 1.7 V / 35 mA ~ 50 ohm

 

equivalent PD current noise = (I_PD / I_LED) * (measured supply voltage noise / 50 ohm)

 

where the PD-LED current ratio was found empirically to be (I_PD / I_LED) ~ 1 / 1000 by measuring the voltage out of the amp with full brightness (i.e. I_LED = 35 mA, no obstruction) and dividing by the transimpedance (see 2nd figure).

 

The third figure below is a plot of these spectral densities in common units. Somewhat expectedly, the noise of the "dark" configuration seems limited by the supply noise. However, the "bright" line seems to be dominated by something else. I'm not sure I see how it could be anything but the LED itself, but it is worthwhile to repeat this "test" with a better setup.

 

On the to-do list:

 

1. Voltage regulator/reference

Rana thinks that the AD587LN is a good choice of reference given its performance on some LISA tests. I am in contact with AD, and there is no longer a 'LN' package, but I am trying to get samples of the currently manufactured one that is most similar (AD587KNZ).

In the meantime, I am going to find some simple regulators downstairs or at the 40m.

 

2. Bandpass filter

I was advised that it is a good idea to build your own high/bandpass filter instead of relying on the spectrum analyzer's AC coupling function. I will be doing just this.

 

3. Switch to a better op amp

        Like the AD743

 

4. Calibration

I need to find a good way to hold the OSEM in place while I stick something in there with a micron drive without it being unreliably shaky.

 

 

 

 

 

 

Attachment 1: schematic.jpg
schematic.jpg
Attachment 2: 2009-11-12_17.34.58.jpg
2009-11-12_17.34.58.jpg
Attachment 3: noise_comparison.png
noise_comparison.png
  123   Tue Nov 17 21:23:08 2009 KojiMiscSUSAOSEM test progress

We have LT1021-7 at the 40m, next to the Alberto's desk. This is the VREF for 7V.

Quote:

1. Voltage regulator/reference

Rana thinks that the AD587LN is a good choice of reference given its performance on some LISA tests. I am in contact with AD, and there is no longer a 'LN' package, but I am trying to get samples of the currently manufactured one that is most similar (AD587KNZ).

In the meantime, I am going to find some simple regulators downstairs or at the 40m. 

 

  124   Thu Nov 19 03:41:12 2009 ZachMiscSUSAOSEM calibration

 Tonight, I calibrated the AOSEM's response in [A/m]. I used a sophisticated rig consisting of:

 

1. One of those anodized Faraday isolator mounts to hold the OSEM

 

2. A translation stage with a screw gauge to jam something into it (gracefully)

 

3. Some DC power supplies and multimeters

 

4. My simple transimpedance amplifier sketched in the previous post (I did not bother upgrading the readout circuit for this measurement since I was just taking a relatively rough DC measurement)

 

---------------

 

I used a 9/64 hex key to simulate the shadowmaking magnet. A picture of the setup is attached below.

 

Some pertinent info:

 

- The current through the LED was maintained at 35 mA throughout the measurement. The measured voltage across it was 1.62 V, giving Z_LED = 46.3 ohm.

 

- The op amp supply voltage was +/- 10 V, and the PD bias was +10 V.

 

- The output voltage of the amplifier with the PD fully lit was 3.04 V (measured before and after the test). Note that this voltage increases slightly as the key is inserted due to reflections. 

 

 

The second attachment is a plot of the photocurrent versus the position of the key (the x axis is shifted such that the key is roughly centered at x = 0, and x < 0 corresponds to the key being further inside). The response of the OSEM in the linear region is roughly 0.05 A/m.

Attachment 1: 100_0362.JPG
100_0362.JPG
Attachment 2: calibration_plot_11_18_09.png
calibration_plot_11_18_09.png
  125   Thu Nov 19 12:00:04 2009 ZachMiscSUSbandpass filter

 Attached is the transfer function for the bandpass filter I built for the AOSEM readout. The schematic is primitively outlined below. The corner frequencies are what they should be (HP: 100 mHz, LP: 1 kHz)

The Johnson noise for the 1 k resistor is roughly 4 nV/rt(Hz). Frank says that it doesn't make sense to use much lower than 1 k if I'm putting it into an SR785.

 

 

           10 uF                  1 k

      -------| |-------------/\/\/\/\/\/------------------

                        | |

        < |

         160 k  >        _|_  160 nF

        <        ---

        > |

        | |

                V                 V

 

Attachment 1: BPF_bode_11_18_09.png
BPF_bode_11_18_09.png
  126   Fri Nov 20 06:18:51 2009 ranaMiscSUSbandpass filter

seems OK, as long as the current noise doesn't get you. To make sure you have to terminate the input to this filter and then look at the resulting noise in the SR785.

  127   Sun Nov 22 16:02:09 2009 ZachMiscSUSAOSEM noise measurement

On Friday, Rana and I discovered that my transimpedance amp was oscillating like whoa at about 100 kHz. A little research showed this to be due to the input capacitance of the AD743 (~20 pF). To fix this, I put a 20-pF cap in parallel with the 100k feedback resistor, and that seemed to do the trick.

 

The relevant circuitry is shown in attachment 1. +/- 12 V DC was provided by voltage regulators (7912, 78M12). The voltage across the LED was measured to be V_LED = 1.61 V, and the current through it was I_LED = 31.6 mA, giving Z_LED = 50.9 ohm. The voltage out of the amp with a fully lit PD was V_out = -2.83 V, giving a photocurrent of I_ph = 28.3 uA. 

 

I was concerned about noise that might be imposed by the bandpass filter, so I compared spectra I took with and without it (that is, AC coupled, no BPF, and DC coupled, with BPF). This comparison is shown in attachment 2. There appears to be no difference apart from the aliasing effects at low frequency.

 

After this, I took the real measurement, extending the range to 800 Hz, averaging 100x and with a linewidth of 1 Hz (I realize now that I should probably have done this with a smaller linewidth, so that I could see below ~1 Hz. I will repeat the measurement this week with better low-frequency resolution). The result can be seen in attachment 3, calibrated to displacement noise in m/rt(Hz) using the measured 0.05-A/m response of the OSEM in the linear region. The four lines are:

 

- Bright: noise in the OSEM with a fully lit PD

- Dark: noise in the OSEM with the LED off

- Amp: noise in the transimpedance amp with the input terminated

- V_LED: noise in the LED voltage

 

The first three spectra were taken at the output of the amplifier and calibrated back to meters using the transimpedance gain and OSEM response in A/m. The last was taken across the LED, and calibrated into meters using the values given in paragraph 2. All measurements were taken with the OSEM under a box and with the lights out.

 

It appears that we are still limited by our setup. The "Dark" noise is coincident with the amplifier noise, while the "Bright" noise is coincident with the LED noise. That said, it is fairly comforting that all this noise is at the level of around 10^-10 m or less, as we can probably expect the true noise of the OSEM to be lower than this. We will know this for sure once we have a truly quiet setup (starting with ultra-low-noise voltage references).

Attachment 1: schematic.png
schematic.png
Attachment 2: BPF_noise_comparison.png
BPF_noise_comparison.png
Attachment 3: 0-800.png
0-800.png
  128   Tue Nov 24 12:28:52 2009 ZachMiscSUSAOSEM measurement update

  I was able to take some better measurements last night. I took data in two bands: 0-100 Hz, 0-1.6 kHz, each with 800 lines. This gives us a decent idea of what's going on at low and high frequency. Attached are four plots, two from each band. All measurements were taken with a box over both the OSEM and the readout circuit and the lights out.

The first two are low- and high- frequency comparisons of the noise in the full (bright) configuration as measured with no BPF and AC coupling vs with the BPF and DC coupling. There appears to be no difference apart from the expected effect above the pole at 1 kHz.

The next two are plots of the noise in various components and the full scheme calibrated into equivalent displacement noise. Everything is below ~10e-10 m/rt(Hz) with the exception of line peaks, and again it would appear that we are limited by our measurement equipment.

Some notes:

- The "dark" noise seems to be coincident with the "amp" noise with the exception of some extra pickup that increases at high frequency (seems to be line-related).

- The "LED" noise is coincident with the "supply" noise up until its 8-Hz corner frequency, after which it falls off as expected until it hits an apparent floor around 100 Hz.

- The "bright" noise seems to be coincident with the "supply" noise, while the "dark" and "amp" are much lower. This could be because the supply noise only shows up when there is an appreciable voltage at the output of the amp.

 

Have to think about this for a bit, but the next logical step is to turn the measurement setup into something solid (i.e. soldering, enclosure, etc.).

Attachment 1: BPF_low.png
BPF_low.png
Attachment 2: BPF_high.png
BPF_high.png
Attachment 3: OSEM_low.png
OSEM_low.png
Attachment 4: OSEM_high.png
OSEM_high.png
  129   Tue Nov 24 23:17:02 2009 ZachMiscSUSAD587KN voltage noise

I got a few AD587KN (high-precision 10V reference) samples today from AD. I hooked them up to see how much quieter my DC supply would be. The results are pretty good, with the voltage noise reduced by a factor of 5-10 throughout. The first two attachments below are comparisons of the noise in

1. The +12V regulator (MC78M12) alone

2. The AD785KN reference with V_in = +12 V provided by the regulator

3. The same as in 2, only now with an additional "noise reduction" capacitor (a 1-uF capacitor from pin 8 to ground forms a LPF with an internal 4-k resistor, giving a corner frequency of 40 Hz to reduce high-frequency noise),

plotted with the same frequency ranges and settings as those in the previous post.

The reference comes very close to its noise spec of 100 nV/rt(Hz) @ 100 Hz. The only issue is that it seems to have much more line pickup than the regulator (which seems almost completely insensitive to line noise), and this is worsened by the extra capacitor. Attachment 3 is a close-up of the low-frequency spectrum around 60 Hz. I suspect that this will be alleviated somewhat when I move away from the breadboard phase.

I want to rig this up so that I can stabilize the supply voltage to the transimpedance amp and LED, but in order to do so I will need to build a higher-current source using a power transistor, like either of those shown in attachment 4 (the AD587LN is only able to provide <10mA). 

Attachment 1: ref_noise_low.png
ref_noise_low.png
Attachment 2: ref_noise_high.png
ref_noise_high.png
Attachment 3: line_noise.png
line_noise.png
Attachment 4: high_current.png
high_current.png
  134   Wed Dec 2 18:29:07 2009 ZachMiscSUSLow-noise LED driver

 Yesterday, I rebuilt Rana's low-noise LED driver in the Bridge elab. It is based on a 2nd-order active lowpass filter (using the Sallen-Key topology). The schematic is shown below. The circuit is essentially the same as the one Rana posted a few days ago, only the R and C values are all around twice what they are in his schematic. This results in the same corner frequency of fc = 0.03 Hz.

 

schematic.png

 

I hooked it up and measured Vsk,out = 9.76 V. I then used it to drive a 50-ohm resistor, and measured VLED = 1.71 V, then measured the current to be ILED = 33.5 mA.

After ensuring that it was supplying the correct voltage, I hooked it up to the LED and took a spectrum of the voltage noise across it over the two frequency bands I have been using in previous posts. The following are comparison plots of the noise here and the noise with the simple RC filter used before, calibrated to displacement noise.

Low-freq:

noise_comp_low.png

RA: although these plots have Displacement in the y-axis, they are NOT measurements of actual displacement noise. They are estimates for the contribution to the displacement noise made by the LED RIN based on measurements of the voltage noise across the LED.

High-freq:

noise_comp_high.png

Something is clearly wrong: not only is the new configuration worse at lower frequencies, but the rolloff seems to go as 1/f and not 1/f2. Investigating after dinner...

 

  135   Thu Dec 3 02:35:39 2009 ranaMiscSUSLow-noise LED driver

 

 OP27 current noise is too high - use AD743.

  136   Mon Dec 7 02:59:38 2009 ZachMiscSUSLED Driver noise (with AD743)

 I retook the measurement from my last post, this time using an AD743 in place of the OP27 (per Rana's comment). The results are below.

Low-frequency:

noise_comp_low.png

RA: Although it says m/rHz, this is not measured displacement noise, but rather estimated displacement noise due to the LED noise. The previously measured conversion from the LED RIN to apparent displacement is used to convert from the voltage noise of the LED driver to the contribution to the OSEM's displacement readout.

High-frequency:

noise_comp_high.png

Seems better than before, but not quite what expected. I observed that the transfer function of the S-K filter was what it should be up until a decade or two above the corner frequency, after which it appeared to spring zeroes out of nowhere and level off at high frequency. I tried to see what would happen if I changed the resistor values, and the following plot is what I got.

S-K_respons_vs_R.png

This plot seems familiar from my electronics courses, but I haven't put a finger on what is causing this behavior yet. I'm sure that the answer is somewhere in H&H (or in the brain of a kind soul who happens to be reading this--wink wink).

  137   Mon Dec 7 12:01:14 2009 ranaMiscSUSAOSEM LED Driver noise (with AD743)

The low frequency noise looks pretty good now. The funny shape is most likely a thermal transient due to having not enough insulation. You need to droop some Kleenex over the circuit to stop the thermal air currents and then put a second box over the first box. Then its probably best to sit outside of the room when taking the measurement to reduce the human noise.

  139   Tue Dec 15 01:48:22 2009 ZachMiscSUSAOSEM noise measurement

 I tracked down some more AD743s at Wilson House 2.0 today (thanks to Rich). I was then able to simultaneously use 743s to both drive the LED and amplify the readout. Below is an 800-line DC - 25 Hz noise spectrum of

- The voltage across the LED (DC level: VLED = 1.59 V) -- AC coupled

- The output of the amp with a fully lit PD (DC level: Vout,full = -2.95 V) -- DC coupled, through bandpass filter

- The output of the amp with the LED out -- DC coupled, through bandpass filter

- The output of the amp with an open input -- DC coupled, through bandpass filter

all calibrated to equivalent displacement noise. For the LED plot, this was done by using the measured current ratio between the LED and the PD when fully lit along with the measured OSEM response of 0.05 A/m. The other three were converted using this response along with the transimpedance gain of the amp (100,000 V/A). For all measurements, the OSEM was covered by a box, and the circuit was draped with a cloth and put under a box within another box (to reduce air currents).

AOSEM_noise_12_14_09.png

The funny low-frequency junk from the previous driver spectrum is gone--thanks to the isolation from air currents--but the line seems a bit higher overall (trying to figure out why). There also still seems to be the funny effect of noise added by the LED above the level of the voltage noise across it, and I think it's somewhat strange that the "Dark" noise is lower than the amp noise with no input. We can probably still do better..

  140   Tue Dec 15 09:28:39 2009 ranaMiscSUSPD front end noise

This is the LISO estimate of the PD front end noise. It could be improved somewhat by using a higher value resistor, but there's no point.

The shot noise level for the 20-40 uA of current we have is more than 1 pA/rHz so we should be OK above 30 mHz.

So even the level of the dark noise below seems too high and also the 10,000 V/A statement. The feedback resistor we had used to be 100k...

Attachment 1: dcpd.png
dcpd.png
  141   Tue Dec 15 10:58:56 2009 ZachMiscSUSPD front end noise

10k was a typo--fixed.

Quote:

This is the LISO estimate of the PD front end noise. It could be improved somewhat by using a higher value resistor, but there's no point.

The shot noise level for the 20-40 uA of current we have is more than 1 pA/rHz so we should be OK above 30 mHz.

So even the level of the dark noise below seems too high and also the 10,000 V/A statement. The feedback resistor we had used to be 100k...

 

  142   Thu Dec 17 10:50:16 2009 ZachMiscSUSAOSEM noise measurement

 I retook the measurement from the previous post, since the noise in the amp appeared much higher than it should. It looks much better now, but still not great. Above around 3 Hz, the amp noise is at the LISO-predicted level of ~8e-12 m/rHz equivalent displacement noise. Below this, it seems to show some 1/f-ish noise back to DC. LISO predicts some low-frequency noise as well, due to the increase in the AD743's current and voltage noise there, but it doesn't begin until below 1 Hz, and it doesn't seem quite as steep. I'm trying to figure out what is causing this, but the best solution might be to move to a more solid, soldered arrangement.
Details:
LED: Voltage across the LED (DC value 1.61 V), AC coupled
Bright: Noise at output of amp, LED on and PD connected, DC coupled with bandpass
Dark: Noise at output of amp, LED off and PD connected, DC coupled with bandpass
Amp: Noise at output of amp, PD disconnected, AC coupled (I checked this DC coupled and through the bandpass, as well, but there is no significant difference)
Amp (theoretical): LISO-predicted noise at the output of the amp with no input
AOSEM_noise_plot_12_15_09.png

 

  143   Fri Dec 18 11:56:12 2009 ZachMiscSUSNew AOSEM LED problem

 I picked up the other AOSEM from the 40m today, so that I could compare it with the one I've measured in an attempt to get to the bottom of the noisy LED problem. It began uneventfully: I measured the impedance of the LED by connecting an ammeter in series and slowly increasing the voltage. I got ILED = 36 mA at about VLED = 1.7, giving ZLED ~ 47 ohms.

Then, I powered up the LED driver, and tested it with a 50-ohm resistor (as usual), measuring V ~ 1.7 across it. Having confirmed that everything was working properly, I hooked the LED up, and measured NO current. I hooked directly back up to the DC supply and found the same result. The thing appears to be blown, but I have no idea how. I went through every precaution I have been taking with the other OSEM, which worked fine when I switched it back in. Crap.

One thing I noticed before I hooked anything up was that the small white pieces attached to the LED and PD on either side of the OSEM opening were very loose when compared to the other OSEM. When I first measured no current, I tried applying some pressure to the LED side, and some current flowed across, but only about 1/10 of what it should have been.

  146   Wed Jan 13 14:13:03 2010 ZachMiscSUSASOSEM comparison

EDIT: I have calibrated the y axis of the plot to meters

Last night, I got around to testing some of the other AOSEM samples, to see how the noise varied between them. What I found was rather strange: the noise in all the new ones (#s 2-6) was about the same, but they were all quite a bit noisier than the previous one I have been testing (#1). The only difference between them, as far as I can tell, is that the first specimen has a coil wound around it already, while the others just have a rubber band. Also, the newer ones all have an impedance of ~ 44-45 ohms, while I measured 47 ohms for the first (though, among the new ones, the slight variation in Z seems to have no correlation with the small differences in noise level). For those wondering, YES, I did remeasure the noise in the 1st one; I am not using old data.

Either my meddling with the old one has somehow made it quieter or something is amiss.

AOSEM_comparison_m_1_12_10.png

  147   Tue Jan 26 22:22:17 2010 ZachMiscSUSAOSEM LED, PD current comparison

Norna and Rich: I am sorry for taking so long to get you this measurement. I plan to do the noise measurements on the standalone LEDs this week.

The following table gives the current through each OSEM's LED (measured using the voltage drop across the 238-ohm resistor in series), as well as the measured photocurrent (the DC output of the amplifier divided by its transimpedance gain of 100,000 V/A), and the ratio of the two. The plot from the previous post is reproduced below for analysis--I realized that I did still have this plot saved in units of V/rHz. In some cases (e.g. #3, #4), the noise level seems to be correlated to the photocurrent, but not all of them follow this pattern. The issue of #1 being significantly quieter than the other set remains, as well.

AOSEM LED and PD Current Comparison
AOSEM # I_LED (mA) I_PD (uA) I_LED/I_PD
1 35.1 31.4 1120
2 35.5 38.5 923
3 35.5 57.6 616
4 35.4 30.5 1160
5 35.4 42.9 826
6 35.5 48.2 737
7 35.5 39.2 906
8 35.4 39.8

889

9 35.2 34.0 1040
10 35.5 43.8 810

 

AOSEM_comparison_1_12_10.png

 

  148   Fri Feb 12 11:33:04 2010 MottMiscSUSState of the Shaker

The Shaker project is coming along nicely.  I am currently looking into using the built-in ability to download a waveform to the front end to do the sweeps, but we are running into memory problems, and I get the sense from Tony that it was not really designed to do this.  Currently we are able to download a waveform to the frontend, run the generator according to it, and make a measurement over a full run of the waveform.  If we can crack the limited time constraint and figure out the averaging, this is going to be the most straightforward solution.

I am working, in parallel, with Gert (who is out of the office at the moment) on using pure script to do this, although I am worried about starting and stopping the generator so frequently.  Apart from anything else, there is a slight hang in the frontend when the generator start method is called; it is not noticeable when the button is pushed in the app, but I think it adds quite a bit of latency to the program.  I am still waiting to hear from Gert about how to acquire a time series; hopefully we can figure that out by early next week, since it is critical.  I am also not entirely sure how we force the program to do all the analysis on the time series after it is acquired.  Ideally we would want the analysis to run in parallel and update the frontend continuously, but I am not sure this is possible with VBA (I don't think you can do multithreaded programming) and I am not sure I would know how to do so even if it is!

 

 

  149   Fri Feb 12 11:35:19 2010 MottMiscSUSPiezos

Engineering was very helpful showing me how to make the leads we need for the piezos; I will go crimp some more at the beginning of next week. 

The new structures should be coming in soon, so we will have a dedicated structure for the piezo damping, at which point we can really get cracking.

ELOG V3.1.3-