40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  SUS Lab eLog  Not logged in ELOG logo
Message ID: 1923     Entry time: Thu Sep 2 17:31:38 2021
Author: Radhika 
Type: DailyProgress 
Category: 1418 nm AUX ECDL 
Subject: 1418 nm ECDL Relative Intensity Noise 

I took a relative intensity noise (RIN) measurement of the ECDL, by feeding the 0th order output of the AOM to the SR785. The RF power driving the AOM was set to 0 dBm. The RIN at 1 Hz is about 3x10-5, which is consistent with informal measurements we took on 08/13. From my understanding this noise looks pretty low, which is good. I will consult with Paco and add more discussion or conclusions, if any.

Attachment 1: ECDL_RIN_02-09-2021_165151_alone.pdf  34 kB  Uploaded Thu Sep 2 18:34:43 2021  | Hide | Hide all