Message ID: 1906     Entry time: Tue Mar 9 19:21:38 2021
Author: Paco 
Type: Misc 
Category: Optics 
Subject: DOPO cavity pole 

- Noticed that the cavity transmission peaks @ 1064 nm were much wider than originally estimated by the dopo cavity design notebook suggesting a lower Finesse. So using the PDH error signal, and knowing the EOM sidebands are at 36 MHz estimated the current DOPO cavity linewidth to be 19.5 MHz, well in excess of the target 10.4 MHz.

- Updated the crystal AR coating specs from Raicol (R < 0.3% @ 1064/2128), but more importantly, I included the absorption coefficient of KTP, alpha=0.005/cm (often quoted as < 0.01 / cm) into the roundtrip loss and the design now gives 17.97 MHz. So, given the uncertainty in the absorption coefficient of the NL crystal, and all the coatings in the experiment, this adjustment might be enough to explain this observation.