40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  ATF eLog, Page 10 of 56  Not logged in ELOG logo
ID Date Author Type Category Subject
  2353   Sun May 26 01:12:33 2019 AnjaliUpdate2micronLasersFrequency noise measurement of 2 micron source using PLL
  • I added one more amplifier stage (ZFL-500 LN) after the detector. Since noise figure of ZFL-500LN (2.9 dB) is lower than that of ZHL-3A (5 dB), ZFL-500LN is the first amplifier stage after the photo detector and it is followed by ZHL-3A.

  • Attachment # 1 shows the beat note spectrum measured from the spectrum analyser. There was a 30 dB attenuator in the path during the measurement. So, the output RF power from the MZI (with two stages of amplification) is now about 3 dBm and the SNR of 37 dB is preserved even after two stages of amplification.

  • So, now the RF power to the RF port of the mixer is 3 dBm. I have attached the video of signal from the PLL loop at different gain (G=1, G=2,G=5) values in SR 560. The time domain trace seems to very noise. I suspect this is because of the inherent large noise in 2-micron laser diode with a broad line width of 2 MHz.

  • I then attempted to do the closed loop transfer function in the present PLL configuration by injecting the signal from SR 785. Attachment 2 shows the closed and open loop transfer functions at different gain values in SR 560 when the actuation slope is 10 kHz/V. Attachment 3 shows the closed and open loop transfer functions at different values of actuation slope when the gain is 5. The magnitude and phase traces are not very smooth as we observed when we did the similar measurement with an arbitrary function generator (AFG) as the RF source. In this case, when MZI output is fed in as the RF source, the RF power is fluctuating.

  • I also tried to do the frequency noise measurement. Attchement # 4 is the FM noise at different gain values when the actuation slope is 10 kHz/V. Attachement 5 is the FM noise at different actuation slope values when the gain is 5. This time, depending on the gain value and the actuation slope value, a short frequency span was considered in SR 785 for the frequency noise measurement. The frequency span is considered based on the value of unity gain frequencies  that are approximated from the open loop transfer functions measured from attachment # 2 and #3

Attachment 1: beat_note.pdf
beat_note.pdf
Attachment 2: Transfer_function_S_10kHzV_different_gain.pdf
Transfer_function_S_10kHzV_different_gain.pdf
Attachment 3: transfer_function_G_5_different_actuation.pdf
transfer_function_G_5_different_actuation.pdf
Attachment 4: FM_noise_diff_gain.pdf
FM_noise_diff_gain.pdf
Attachment 5: FM_noise_G_5_different_actuation_slope.pdf
FM_noise_G_5_different_actuation_slope.pdf
Attachment 6: Gain_1.mp4
Attachment 7: Gain_2.mp4
Attachment 8: Gain_5.mp4
  2352   Thu May 23 15:29:58 2019 awadeDailyProgressWOPOMaking time domain measurement of noise power in homoydne with WOPO pump on

In order to see squeezing I want to scan the phase of the homodyne relative to the squeezed light and see the variations in noise power as a function of time. 

Excess 532 nm dumped on detector A

From my initial scans it seems like there is excess noise of a factor of 1.5 above shot noise (see below).  This is only present with 532 nm pump injected. It could be anit-squeezing washing out with some phase noise bluring across the sample time of 0.1 seconds or maybe RIN from residual 532 nm present at one of the photo detectors.

After injecting light into the WOPO for some initial tests it was apparent that there was about 1.4 mW of waste 532 nm light exiting the fiber launch on the detector A path. About 16 µW gets through to the detector from the dichroic mirror reflection. A quick measurement of 532 nm power on detector B show that there was about 49 µW coming out of that fiber end with 0.2 µW making to the photodiode. The 50:50 splitting doesn't apply for non-design spec wavelengths of fiber splitter.   The InGaAs detectors have a pretty poor responsivity at this wavelength but the pumping light on detector A was enough to create a DC voltage of 1.36 mV. After dividing through by detector gain of 2kΩ this is equivalent to about 0.68 µA of DC power on the detector A.  This suggests a responsivity of order 0.014 A/W.  There isn't enough light on detector B to create any DC voltage. 

The imbalance here, with the 532 nm light, means that there is a mechanism for coupling in 532 nm RIN into the measured output signal.  Not sure if the RIN would be all that high at the 600 kHz (that I'm planning to mix down from) but it would be a good idea to remove it anyway.  I'll look for a another HR1064/HT532 dicroic mirror to attenuate this 532 nm component a little more.

 

 

Quote:

For the detection of squeezed light the homodyne detector  is now configured for digital subtraction in post processing. The measurement scheme is now to take the output of the two TIA amplifiers (see QIL:2324 and QIL:2327), digitize​ them directly at 210 MSa/s (5 nV/rtHz input ref noise) with the Zurich box and demodulate both detector streams using its internal FPGA at about 1 MHz.  The Zurich box allows for direct sampling of the signal out of the FPGA which can then be downloaded either through the web interface or the python API. T

Attachment 1: RMSFunctionTimeCompairingSNWithInjected532nmpump.pdf
RMSFunctionTimeCompairingSNWithInjected532nmpump.pdf
  2351   Thu May 23 03:00:24 2019 AnjaliUpdate2micronLasersFrequency noise measurement of 2 micron source using PLL
  • The delay line length is measured to be 15.7 m. 
  • As suggested by Prof.Rana, I used a DC block after the photodetector and the output of it was then fed into a RF amplifier (ZHL-3 A+). Attachment #1 shows the output of the detector (a) MZI output- we see the beat note at 78.8 MHz and SNR is about 37 dB (b) when light is bloked- we see  radiation from RF amplifier with a small amplitude of about -63 dBm at 78.8 MHz  (c) when RF is turned off. 
  • It can be seen that, after the RF amplifier, the noise level is also increased with the signal.
  • The RF level that is now fed into the mixer is about -23 dBm. But the PLL was not getting locked-oscilloscope trace is oscillating. A video is attached, which is captured when the there were no gain in SR 560 (G=1) and actuation slope was 10 kHz/V. Attachment 2 and 3 shows the corresponsing traces from SR 785 in full span mode and short spn mode respectively. It is observed that , the oscillation strength was increasing as I increased the gain value in SR 560. I doubt, this is happening because of the higher noise level at the RF port of the mixer in this case. The trace on SR 785 was more stable and it was not swinging as much as we observed before. 
Attachment 1: Detector_output.pdf
Detector_output.pdf
Attachment 2: FM_noise_full_span.pdf
FM_noise_full_span.pdf
Attachment 3: FM_noise_short_span.pdf
FM_noise_short_span.pdf
Attachment 4: oscilloscope.mp4
  2350   Thu May 23 02:53:19 2019 AnjaliUpdate2micronLasersCharacterization of new detector

FC/UPC to FC/APC patch cord has arrived. I repeated the DC characterisation of the photodetector with this patchcord. The couping is improving by about 2 dB (Table below shws the result)

Laser diode current (mA) Input power (mW) Output voltage (mv)
50 0.5 11
70 0.9 20.7
90 1.3 30.3
Quote:

In this case, the input power was measured after the isolator.The power to voltage conversion is linear. The voltage levels are very low because this is a non-amplified detector. Also, the detector is coupled to a FC/UPC patch cord and we have all FC/APC fiber connectors. So, there could be some coupling loss from FC/APC to FC/UPC. FC/APC to FC/UPC conversion patch cord is ordered. We can check the performance again after it is arrived.

  2349   Wed May 22 15:07:09 2019 awadeDailyProgressWOPODemodulation and subtraction for homodyne detector with Zurich box

For the detection of squeezed light the homodyne detector  is now configured for digital subtraction in post processing. The measurement scheme is now to take the output of the two TIA amplifiers (see QIL:2324 and QIL:2327), digitize​ them directly at 210 MSa/s (5 nV/rtHz input ref noise) with the Zurich box and demodulate both detector streams using its internal FPGA at about 1 MHz.  The Zurich box allows for direct sampling of the signal out of the FPGA which can then be downloaded either through the web interface or the python API. The signal chain is illustrated below:

Whiteboard illustration of WOPO HD detector chain
WOPO homodyne detection chain (see attachments at bottom for high resolution)

The basic approach is to mix down the shot noise limited light from some higher frequency (somewhere in the range 100kHz - 8 MHz) and then compute the ideal subtraction factor from a short (10 sec) segment of data.  The method for optimization is to compute the time domain subtracted signal for a given balancing factor (SF) and compute the goodness of subtraction from the RMS of the signal.  An fmin search is done to find the ideal subtraction factor (SF) for shot noise limited light.

It wasn't immediately clear to me what demodulation​ frequency to choose, the low pass filter and the digitization sampling rate. So I set up a python notebook and did a few sweeps of these parameters to find the best combination of parameters. This book is attached below as 20190514notebook_ZurichDemodeCompairTests.ipynb in a zip. I've pickled data for replotting if needed later.  

Sweeping the demodulator 4th order LP corner frequency

At present the detectors are illuminated with 4.5 mA worth of light (as measured from DC voltages / TI gain).  This should give about 25 nV/rtHz at the output of the photodetector, as measured by the Zurich box.  Fixing the demodulation frequency at 500 kHz and the sampling rate at 30 kHz I found empirically that the when the LP filter (4th order) was kept at 10 kHz and below the minimum subtracted signal converged to about 30 nV/rtHz (about what I expected).  With the demodulation frequency set much higher than this, the minimum subtracted signal no longer converged to the predicted SN level.  Maybe there is some aliasing thing going on here.  The outcome of this experiment is that clearly the demodulation LP filter should be set to at least 1/3 of the sampling rate for the time domain rms minimization to work.  

Selection of sampling frequency

I did a similar sweep with the sampling rate while keeping the demodulation frequency fixed at 500 kHz and the demodulation low pass fixed at 10 kHz.  The conclusion was basically the same: as long as the demodulator low pass was kept within 1/3 of the sampling frequency then the method of computing the ideal subtraction factor using the total time domain rms as a cost function returned the minimum possible differential noise.

Selection of demodulation frequency

The demodulation frequency choice is a tricker one.  Initially I selected 1 MHz but was getting excess noise above what I expected for shot noise. I then looked at a single channel on the Zurich's scope in FFT mode.  Data is plotted below

Basically the plot shows that there is a strong peak at about 900 kHz. This is due to the relaxation oscillation of the laser.  When the laser's noise eater is turned on this peak is damped with some tradeoff of slightly higher noise at lower frequencies. So when choosing a demodulation frequency it is best to select something well below 1 MHz. I'm not sure if it is best to turn the noise eater off if the demodulation frequency is set well clear of the 900 kHz peak.  It seems like there is a noise penalty for the noise eater loop at all other frequencies. 

With the noise eater on I then did a sweep of the demodulation frequency while keeping the low pass filter at 10 kHz and the sampling rate at 30 kSa/s.  I've plotted the median ASD value in the band between 100 Hz and 1 kHz as an indication of the minimum SN level after digital subtraction verses the demodulation frequency.  Plot below.  There is a hump at about 1 MHz that should be avoided, and spikes at 87 kHz and just below 300 kHz but anything else in the range of 100 kHz to 700 kHz should be fine.  

Current selected configuration

So for now, based on the above sweeps of parameters, I have selected demodulation frequency of 600 kHz that is low passed at 10 kHz with a 4th order filter and sampled at 30 kSa/s.  As an example of optimized subtraction I have plotted the two homodyne channels along with the subtracted differential signal (optimized by minimizing subtracted time domain rms) and the dark noise (using the same subtraction value).

Here the clearance from the dark noise floor doesn't seem so great.  The input referred noise of the Zuirch box is 5 nV/rtHz per channel.  Also the equivalent output noise from the PDs at this point is 2.5 nV/rtHz.  As these are uncorrelated noise processes this would place the noise floor at about sqrt(5^2*2+2.5^2*2) = 7.9 nv/rtHz, pretty close to what we are actually seeing. 

One option here is to pre-amplify the PDs before going into the zurich box.  A standard mini-circuits amplifier like ZFL-500LN+ has an input referred noise of 1.6 nV/rtHz which would bring the dark noise floor down to something closer to 4.2 nV/rtHz.  I'll keep going without any pre-pre-amplification for now and see what I can do with 532 nm pumping.  There should be enough clearance to see at least something if it is there.

 

 

Attachment 2: Sweep_DemodLPFValues.pdf
Sweep_DemodLPFValues.pdf
Attachment 3: SingleHDChannel_RelaxOscillations_CompairNoiseEaterONOFF.pdf
SingleHDChannel_RelaxOscillations_CompairNoiseEaterONOFF.pdf
Attachment 4: DemodFreqSweep_median100Hzto1kHz_vsdemodFreq.pdf
DemodFreqSweep_median100Hzto1kHz_vsdemodFreq.pdf
Attachment 5: CompairSubtractedSignalToDarkNoise.pdf
CompairSubtractedSignalToDarkNoise.pdf
  2348   Mon May 20 02:23:14 2019 AnjaliUpdate2micronLasersFrequency noise measurement of 2 micron source using PLL
  • Attachment # 1 shows the schematic of the experimental setup for the frequency noise measurement of 2-micron laser source using PLL. Instead of Brimrose driver, another Marconi is used to provide the RF power to the AOM. We know from the characterisation of AOM that we need to give RF power of 28 dBm at 78.8 MHz to achieve maximum diffraction efficiency to the first order port of AOM. The maximum output power from Marconi is 13 dBm. Hence, we used another RF amplifier (ZHL-3-A+) to amplify the RF power from Marconi. We initially tested the RF output from RF amplifier on spectrum analyser (RF power fed into spectrum analyser with proper attenuation in the path) and adjusted the RF frequency and power in Marconi such that we get 28 dBm output power from the RF amplifier at 78.8 MHz. The two marconis are set such that they are share the same time standard.

  • Now, the output power from the photodetector in MZI (Laser diode operated at input current of 90 mA) is fed into the RF input port of the mixer, instead of AFG. The 600 Ohm output of SR 560 is observed on oscilloscope and SR 785 simultaneously.

  • We observed dc line in the oscilloscope when the gain in SR 560 is set to 13 dB (20 times). Gain value below this ( 10 dB) or above this (17 dB) was showing oscillations in the oscilloscope with frequency varying with the actuation slope in Marconi. Attachment #2 shows the frequency noise measurement from SR 785 (V/rt Hz value from SR 785 multiplied with the actuation slope).  

  • It is observed that, the time domain trace on the oscilloscope was not very stable. In between, we could see the oscillation was popping up. Also, the trace on SR 785 was swinging a lot (attached the video). As we observed in the case of AFG, the FM noise measured increses with the value of actuation slope in Marconi. 

  • In this case, the RF power that is fed into the RF port of the mixer is very small (~ -40 dBm) compared to our previous experiment with AFG. So, I would like to repeat the sample experiment (locking AFG to Marconi) with AFG set to RF power comparable with that from the actual experiment.I should then find out the unity gain frequency of that particular combination of gain and actuation slope,which would help us to find the frequency range upto which the PLL measurment is valid.

  •  I also need to measure the actual delay line length. I will also clean up the fiber connectors again and we can also use the FC/UPC to FC/APC patch cord for the detector after it is arrived. I still must understand the results better.Since we are using Non-PM fibers, the polarisation fluctuation might have also affected the measurement .Kindly give me further suggestions.

Attachment 1: FM_noise_measurement_setup.png
FM_noise_measurement_setup.png
Attachment 2: 2_micron_FM_noise.pdf
2_micron_FM_noise.pdf
Attachment 3: Oscilloscope.mp4
Attachment 4: SR_785.mp4
  2347   Sun May 19 22:08:09 2019 AnjaliUpdate2micronLasersPLL loop for frequency noise measurement
  • After the discussion with Prof.Rana, we realised the mistake in our analysis. It was also suggested to make the measurement at the output of SR560. Attachment #1 shows the schematic of the setup for the measurement of closed loop transfer function. The RF power from AFG is 0 dBm and that  from Marconi is 7 dBm.

  • The open loop transfer function is calculated from closed loop transfer function using the expression G_{ol}=\frac{G_{cl}+G_{560}}{G_{cl}} , where G_{560}} is the gain value set in SR560.

  • Attachment #2 : Closed loop and open loop transfer functions at different values of gain in SR 560 when the actuation slope in Marconi is 10 kHz/V. The unity gain frequencies are respectively 1kHz, 3 kHz and 5 kHz when the gain values are 3 dB, 7 dB and 10 dB.

  • Attachment #3 : Closed loop and open loop transfer function at different values of actuation slope in Marconi when the gain in SR 560 is 10 dB.The unity gain frequencies are respectively 5kHz, 16 kHz and 43 kHz when the actuation slope values are 10 kHz/V, 30 kHz/V and 100 kHz/V. It can be seen that the characteristics are significantly different for a larger value for the product of gain and actuation slope (G=10, S=100 kHz/V).

  • The probing signal from SR 785 was then disconnected. In this case, the oscilloscope measure the error signal in time domain and the measurement from SR 785 essentially gives the frequency noise of AFG. The measurement from SR 785 has the unit of V/rt Hz, which is then multiplied with actuation slope to get the frequency noise in Hz/rt Hz. During our measurements, the oscilloscope signal was showing a low level ( mV) DC line , confirming that the PLL is  locked.

  • Attachment # 4 : Frequency noise of AFG at different gain value in SR 560 when the actuation slope in Marconi is 30 kHz/V. In the full span mode, the line width (resolution) is 128 Hz where as the line width is 2 Hz in short span mode. The peak at 60 Hz visible in short span plot corresponds to AC mains. 

  • Attachment # 5 : Frequency noise of AFG at different  actuation slope in Marconi when the gain in SR 560 is 10 dB.  I was thinking, we should measure the same frequency noise irrespective of the setting in the PLL. It can be seen from attachment # 5 that the frequency noise measurement is affected by the value of actuation slope in Marconi. It was earlier observed that the phase noise of Marconi increases with increase in the actuation slope and , from these measurment shown in attachment #5, we are seeing increase in frequency noise value at larger values of actuation slope in Marconi.

Quote:

 

  • We also tried to estimate the open loop transfer function from the closed loop transfer function using the equation G_{ol}=\frac{G_{cl}}{1-G_{cl}}

  • Attachment # 7 : Comparison of Open loop transfer function that is measured from SR 785 and that is estimated from the closed loop transfer function using the above expression. These two values are significantly different. Kindly correct me.

 

Attachment 1: setup_close_loop_transfer_function.png
setup_close_loop_transfer_function.png
Attachment 2: Actuation_slope_10kHzV_diff_gain.pdf
Actuation_slope_10kHzV_diff_gain.pdf
Attachment 3: gain_10_different_actuation_slope.pdf
gain_10_different_actuation_slope.pdf
Attachment 4: FM_noise_AFG_Actuation_30kHzV_diff_gain.pdf
FM_noise_AFG_Actuation_30kHzV_diff_gain.pdf
Attachment 5: FM_noise_AFG_gain_10_different_actuation_slope.pdf
FM_noise_AFG_gain_10_different_actuation_slope.pdf
  2346   Sat May 18 22:25:30 2019 AnjaliUpdate2micronLasersCharacterization of new detector

The new photo detector has arrived (https://www.newport.com/p/818-BB-51F) . We did the DC and AC characterization of the same.

  • Following table shows the results from DC characterisation            
  • Laser diode current (mA) Input power (mW) Output voltage(mV)
    50 0.5 7.2
    70 0.9 13.6
    90 1.3 19.7
  • In this case, the input power was measured after the isolator.The power to voltage conversion is linear. The voltage levels are very low because this is a non-amplified detector. Also, the detector is coupled to a FC/UPC patch cord and we have all FC/APC fiber connectors. So, there could be some coupling loss from FC/APC to FC/UPC. FC/APC to FC/UPC conversion patch cord is ordered. We can check the performance again after it is arrived.

  • We then assembled the Mach-Zehnder interferometer (MZI) for the 2-micron laser source. Attachment #1 shows the schematic of the same. We measured a power level of 0.37 mW when the AOM was not turned on (RF power to AOM off). When AOM is turned ON, the power level measured at the output of MZI is 0.5 mW. Power meter was then replaced with the new photodetector and the beat note was observed on spectrum analyser.

  • Attachment # 2 shows the RF beat note at 78.8 MHz on spectrum analyser. So, the band width of the detector is enough to work at this frequency range. The signal to noise ratio is about 37 dB. There are some small peaks appearing at around 77 MHz and 80 MHz, but they are about  29 dB below the main peak.
  • Attachment # 3 shows the noise floor. There are no other RF interferences at this frequency range.
Attachment 1: setup.png
setup.png
Attachment 2: beat_note.pdf
beat_note.pdf
Attachment 3: Noise_floor.pdf
Noise_floor.pdf
  2345   Thu May 16 16:52:46 2019 awadeHowtoWOPOZurich box python access

The Zurich box can also be accessed via a python API.  Its a little bit opaque in some points on how this works as their documentation is a little thin on actual examples.  I've included a jupyter notebook and equivalent .py file for future reference.  This code is tested and has worked with the Zurich box data server at IP address 10.0.1.23.  It shows exactly how settings can be modified and how data can be read back using their poll command.

Notebook includes all instructions for setting the Zurich HF2LI box up from scratch. 

 

Attachment 1: ZurichPythonAccess_MWE.zip
  2344   Thu May 16 13:40:13 2019 anchalUpdate2micronLasersPLL loop for frequency noise measurement

I think you have made some coding error in your attachment 7 plot. Just pick a point in your plot and calculate by hand if your estimate is correct. Otherwise, we need to see your code to pinpoint the error. You can attach your code in a .zip file here.

  2343   Wed May 15 20:08:01 2019 AnjaliUpdate2micronLasersPLL loop for frequency noise measurement
  • We are trying to setup the phase locked loop (PLL) for the frequency noise measurement of 2 micron laser. We started with a sample experiment in which we were trying to lock an arbitrary function generator (AFG) with the Marconi using PLL. Attachment #1 shows the schematic of the PLL setup. We are using a level 7 mixer (ZFM-3-S+). The RF port is connected to  AFG and LO port to Marconi. Output of mixer is passing through a low pass filter with cut off frequency at 1.9 MHz (SLP-1.9 +). The output of LPF is fed into input A of SR 560. SR560 is set with 1 MHz low pass. Initially, we set a gain value of 10 dB and actuation slope of 100 kHz/V in SR 560 and Marconi respectively. The 50 Ohm output  of SR 560 was connected to Marconi and the 600 Ohm output was connected to an oscilloscope to check the performance of PLL. The carrier frequency from AWG and Marconi were set close to each other (~ 11 MHz). We observed  a dc output at about 60 mV on the oscilloscope. This ensures that PLL is working.
  • We then attempted to measure the band width. To do that, the source output from SR785 was fed into input B of SR560. Part of the source output was fed into channel 1 of SR 785, through T connector, for the transfer function measurement. We also used a T connector at the input A port of SR 560 and one of the ports of this T connector was fed into channel B of SR785. I still must interpret most of the results that we got.

  • Attachment # 2: Closed loop transfer function (a) Magnitude (b) Phase, at different gain values  in SR 560 when the Marconi actuation slope is 10 kHz/V.

  • Attachment # 3: Closed loop transfer function at different actuation slope value in Marconi when the gain is 7 dB.  The increase in noise at lower frequency in phase plot (b) may indicate that the phase/frequency noise of the Marconi increases if the actuation slope value is increased. 

  • Attachment # 4: Closed loop transfer function at different actuation slope value set in Marconi when the gain is 10 dB. The transfer function measured for the case of gain = 10 dB and actuation slope = 100 kHz/V (that is the product of gain and actuation slope is larger) shows significantly different characteristics.

  • Using the SSUserFn option in SR785, we tried to get the open loop transfer function as well from SR 785. The functional form was \frac{x}{1-x}

  • Attachment # 5: Open loop transfer function at different gain values set in SR 560 when the Marconi actuation slope is 10 kHz/V. The unity gain band width are 0.9 kHz, 2.8 kHz and 5.8 kHz respectively when the gain values are 3 dB, 7 dB and 10 dB

  • Attachment # 6 : Closed loop transfer function at different actuation slope value set in Marconi when the gain is 7 dB. The unity gain band width are 3 kHz, 9 kHz and 30 kHz respectively when the actuation slope values are 10 kHz/V, 30 kHz/V, and 100 kHz/V.

  • We also tried to estimate the open loop transfer function from the closed loop transfer function using the equation G_{ol}=\frac{G_{cl}}{1-G_{cl}}

  • Attachment # 7 : Comparison of Open loop transfer function that is measured from SR 785 and that is estimated from the closed loop transfer function using the above expression. These two values are significantly different. Kindly correct me.

Attachment 1: PLL_setup.png
PLL_setup.png
Attachment 2: closed_loop_FM_dvn_10kHz_diff_gain.pdf
closed_loop_FM_dvn_10kHz_diff_gain.pdf
Attachment 3: closed_loop_gain_5_diff_FM_dvn.pdf
closed_loop_gain_5_diff_FM_dvn.pdf
Attachment 4: closed_loop_gain_10_diff_FM_dvn.pdf
closed_loop_gain_10_diff_FM_dvn.pdf
Attachment 5: open_loop_FM_dvn_10kHz_diff_gain.pdf
open_loop_FM_dvn_10kHz_diff_gain.pdf
Attachment 6: open_loop_gain_5_diff_FM_dvn.pdf
open_loop_gain_5_diff_FM_dvn.pdf
Attachment 7: Comparison_measured_estimated_open_loop.pdf
Comparison_measured_estimated_open_loop.pdf
  2342   Wed May 15 19:24:52 2019 JonLab InfrastructureCymacsPermanent cymac hardware installed and tested

[Jon, Chub]

Today Chub brought over the remaining parts for the cymac and we installed them. Specifically:

  • ADC and DAC adapter boards permanently mounted inside the fb4 chassis
  • SRS DS345 function generator mounted on a new shelf installed near the top of the rack
  • New, higher-quality timing cables installed
  • 160' timing cable run from the cryolab to the QIL, to provide a 10 MHz GPS reference signal

Initially I found that no timing signal was reaching the lab. The function generator for the cryolab cymac was also not receiving a signal. I traced the problem back to the RF distribution box in the cryolab, whose power cable appears to have come unplugged. I reconnected it and both labs are now receiving the timing signal.

After installation, I repeated the manual loopback test in #2337 for all ADC and DAC channels. All of the channels still work, and the ADC gain was found to identical to before (0.38 ct/ct). For future reference, the second attachment shows the state of the system overview screens when the cymac is in normal operation. Chris has confirmed that not every indicator light on these screens will turn green. 

The last piece of hardware work is to replace the bench power supply currently powering the AA and AI chassis. Chub and I plan to replace it with two rackmount Sorensenn supplies, one of which is already installed in the rack. Chub is going to procure a second.

Attachment 1: IMG_3300.jpg
IMG_3300.jpg
Attachment 2: IMG_3299.jpg
IMG_3299.jpg
  2341   Tue May 14 13:10:36 2019 ranaLab InfrastructureHVACHVAC debuggin

fixed (?) HVAC problems (~1700 Monday) by better attaching a pneumatic hose that actuates the hot water flow for the south blowers

update on Tuesday: problem not fixed. Bumping the actuator there makes the thing work for a few minutes, but then it freezed again.

Attachment 1: HVAC-B265B.pdf
HVAC-B265B.pdf
  2340   Sat May 11 15:47:37 2019 JonLab InfrastructureCymacsMEDM user interface set up on QIL cymac

I created a basic MEDM sitemap modeled after the one the cryolab is running. It is accessed by connecting to fb4 (10.0.1.156) and executing the usual command $ sitemap. Eventually the fb4:/opt/rtcds/caltech/c4/medm directory should be shared with all QIL workstations via NFS, so that directly connecting to fb4 is not necessary. The master screen provides full control of the signal routing and conditioning supported by the user model. The 32 input signals from ADC0 are individually conditioned via standard LIGO filter bank modules and routed through an output coupling matrix, which constructs the 16 output signals sent to DAC0.

Attachment 1: Screen_Shot_2019-05-11_at_3.21.07_PM.png
Screen_Shot_2019-05-11_at_3.21.07_PM.png
  2339   Fri May 10 18:39:10 2019 ranaLab InfrastructureHVAChowto get temperature trends

Modesto came back this morning. We were still getting 1 hot blower and 1 cold blower.

As it turned out, although hot water was reaching the south heating coils input, it was not getting into the actual heating duct (as we could tell from touching those coils).

Modesto tapped the actuation valve with a hammer and it seemed to loosen it up. We will record temperature over the weekend.

Also we noted that the damper valve near the south-east vent was adjustable, but had no effect on the output air velocity so we assume its busted inside - Modesto will crack it open to investigate.

We also reduced the flow of hot water to the north blowers since they always seemed to be hotter. This may balance the actuators somewhat.

Modesto also replaced the thermostat on the west wall, but no real effect. I have adjusted the setpoint from 74 to 71F around 1800 today.

  2338   Fri May 10 16:22:45 2019 AnjaliUpdate2micronLasersCharacterization of RF driver and AOM .

Following are the results from RF driver and AOM characterisation.

  • Attachment #1 shows the results from  characterisation of Brimrose RF driver . The RF power and frequency are measured on Agilent spectrum analyser. A 30 dB attenuator was also used in the path from RF driver to spectrum analyser. This attenuation value is taken care in RF power output calculation. The RF driver has two BNC connectors labelled as “Modulation” and “Frequency” , located on the front panel. Varying the modulation input (in the range 0 V to 1 V) changes the RF output power from the RF driver as shown in attachment #1 (a). The maximum RF output power is about 0.6 W and the input RF power to the AOM is limited to this value as exceeding the same might cause damage to the AOM. Varying the frequency input (in the range 0 V to 10 V) changed the RF frequency from the RF driver as shown in attachment #1 (b). The AOM centre frequency is at 80 MHz with a frequency shift range of 8 MHz.

  • Attachment #2 and #3 shows the output power from the zeroth order and first order port of the AOM when the frequency input voltage to the RF driver (thus the RF frequency from the driver) is varied. The output power from the first order port is maximum (output power from zeroth order is minimum) when the frequency is about 78.8 MHz.  As expected, the power in the zeroth order port is completely transferred to the first order port. This happens when the frequency input voltage to the RF driver is about 8.8 V.

  • Attachment #4 and #5 shows the output power from the zeroth order port of the AOM when the Modulation input voltage to the RF driver (thus the RF power from the driver) is varied. The output power from the first order port is maximum (output power from zeroth order is minimum) when the RF power to the AOM is maximum. This happens when the modulation input voltage to the RF driver is 1 V.

  • Attachment #5 shows the diffraction efficiency to the first order port (we use output power from first order port for the heterodyne measurement) as a function of frequency and RF power. The diffraction efficiency is calculated from the ratio of Power in first order port to the input power to the AOM.The power at the input end of AOM is 1.29 mW. So the percentage value calculated includes the insertion loss ( as per data sheet : 3-4 dB for the first order port ) as well . So the conclusion is , inorder to get maximum diffraction efficiency to the first order port of AOM, we should supply RF power of about 0.6 W at 78.8 MHz . If we are using the Brimrose driver, this can be set by giving modulation input voltage of 1 V and frequency input voltage of 8.8 V.

Attachment 1: RF_driver_characterisation.pdf
RF_driver_characterisation.pdf
Attachment 2: AOM_zero_order_output_Vs_Frequency.pdf
AOM_zero_order_output_Vs_Frequency.pdf
Attachment 3: AOM_first_order_output_Vs_frequency.pdf
AOM_first_order_output_Vs_frequency.pdf
Attachment 4: AOM_zeroorder_output_Vs_RF_power.pdf
AOM_zeroorder_output_Vs_RF_power.pdf
Attachment 5: AOM_first_order_output_Vs_RF_power.pdf
AOM_first_order_output_Vs_RF_power.pdf
Attachment 6: Diffraction_efficiency.pdf
Diffraction_efficiency.pdf
  2337   Fri May 10 13:54:28 2019 JonLab InfrastructureCymacsQIL cymac installed in rack, passes loopback testing

After a bit more debugging, Chris and I determined the RTS software was fully functional. At this point I moved all the cymac hardware from underneath the ws2 desk to the neighboring server rack. I permanently mounted the linux machine, fb4, and the anti-alias and anti-image chassis (pictured). Installing fb4 required removing an existing machine, fb1, which Chris and Andrew confirmed is not in current use. For now fb1 is sitting against the wall near the rack, as we may be able to repurpose it. The rest of the hardware is connected temporarily for testing while we await a few remaining cables, etc. (more on this below).

I performed a simple loopback test on the cymac input/output channels. I set a large DC offset (10k counts) in one DAC channel at a time, connected this signal to an input ADC channel, and measured the input channel response. All 16 DAC channels and all 32 ADC channels were confirmed to drive/sense the signal, and in every case the ADC gain was found to be identical: 0.38 counts/count.

There are still a few items to wrap up the installation:

  • [Jon] Create a sitemap MEDM screen
  • [Chub] Two 5' SMB-BNC cables, for connecting the timing reference to the adapter boards. The ones in use now are poor-quality. Chub has already ordered these cables.
  • [Chub] Two backplates for the D060060/D060061 adapter boards, to let them mount in the computer chassis like a PCIe card
  • [Chub] Install a long BNC cable between the cryolab and ATF server racks, for delivering the 10 MHz GPS timing reference. Chub and I walked through the labs this week and determined the cable route and required length.
  • [Chub] +/-15V rackmount power suppy, to replace the bench supply currently power the AA and AI chassis
  • [Chris, Jon] Configure the framebuilder functionality
Attachment 1: IMG_3289.jpg
IMG_3289.jpg
  2336   Tue May 7 19:39:39 2019 JonLab InfrastructureCymacsUser model set up on QIL cymac

This afternoon I set up a user model on the QIL cymac (fb4), which appears to be running. It required the following steps.

First I created a simple SimuLink model of the signal flow from the one ADC card to the one DAC card (pictured below). The 32 input channels from the ADC are each routed through a standard LIGO filter module (10 IIR filter slots, ramped gain, and test/excitation points). From there the 32 conditioned signals are routed into an output matrix constructing the 16 output signals to the DAC. I saved this model to /opt/rtcds/userapps/local/c4tst.mdl.

Next I compiled the model and installed the build files to the RTS target directories:

$ cd /opt/rtcds/rtscore/release
$ make CC=gcc-4.4 c4tst
$ make CC=gcc-4.4 install-c4tst

Finally I started the new model services and enabled them to auto-start on boot.

$ sudo systemctl start rts@c4tst.target
$ sudo systemctl enable rts@c4tst.target

The cymac I/O has not yet been tested.

Attachment 1: IMG_3278.jpg
IMG_3278.jpg
  2335   Sun May 5 20:58:20 2019 ranaLab InfrastructureHVAChowto get temperature trends

the temperature has stabilized, although the fact that there is a sign error in one of the HVAC actuators means that there is a big delta T in the lab. I will contact facilities again.

the change on April 24 is me adjusting the setpoint on the south wall from 70 to 73F.

Quote:

We need to get the temperature data from the sensors that Aidan setup. I have got it by downloading a sketchy HOBO app onto my phone. In order for this to work you have to go stand next to the sensors and download via bluetooth. There is no "pairing" required - the app seems to recognize the sensors nearly immediately.

and please, Aidan and everyone else, please do not use double-sticky tape or glue or bubble gum to attach sensors in a clean optical setup. We're going to have to make a mess to remove these.

I've just downloaded more data (attached). The lab seemed sort of cold and the south HEPA output was blowing colder air than the north HEPA output.

I've changed the temperature from 70F to 73F at 7:45 PM today. Lets see how it goes.

the python script in the attachment will now scrub the data file for bad lines and make this kind of plot. IT works on .csv files, but none of the other HOBO formats.

Attachment 1: HVAC-B265B.pdf
HVAC-B265B.pdf
Attachment 2: plotTemp.tgz
  2334   Sun May 5 20:52:16 2019 ranaNoise Budget2micronLasersFrequency noise measurement of 2 micron source

Looks good! A few comments:

  1. The LPF should be a 1.9 MHz LP from Mini-circuits
  2. The SR560 should be set with a 1 MHz low pass only
  3. No Rubidium needed. We just want to use two IFR/Marconi synthesizers so that they can share the same time standard.
  4. We will need a RF Amp to drive the AOM hard enough - check the manual and see what level is required to get the max diffraction.
  2333   Sun May 5 15:17:33 2019 AnjaliNoise Budget2micronLasersFrequency noise measurement of 2 micron source

We are going to start the heterodyne configuration for the frequency noise measurement of 2 micron laser source. The schematic of the same is shown in attachment #1. The initial plan was to separate the photodetector output to in phase , quadrature component and then extract the frequency noise information from the same. The new plan suggested is to use a phase locked loop for the demodulation part. We can use IFR Marconi 2023A as a VCO and SR560 as the amplifier. The RF generator for the AOM and the VCO need to be locked through Rubidium frequency standard (FS725). One of the issues is that the band width of the photodetector that we have is 14 MHz and the AOM is at 80 MHz. The response of photodetector at 80 MHz is about 20 % of that at DC.I have modified the noise budget after discussing with Anchal. Attachment #2 shows the noise budget. The Marconi data is taken from Anchal. The data plotted in attachment #2 corresponds to Marconi #539 at 48 MHz and actuation slope of 500 Hz. We may have to find the actual numbers based on our settings for the experiment.  From the data sheet, the input noise of SR 560 is 4nV/rtHz. The frequency mixer (Minicircuit ZAD-1-1+) has a conversion loss of 4.83 dB and the thermal noise of it is not given in the data sheet and thus it is not included in the noise budget. The length of the fiber considered for MZI is 10 m.

 

Attachment 1: schmeatic_heterodyne.png
schmeatic_heterodyne.png
Attachment 2: Noise_budget_modified.pdf
Noise_budget_modified.pdf
  2332   Wed May 1 14:34:53 2019 awadeDailyProgressWOPOOutput chain and demodulation of squeezing signal

The detectors have been verified to have a low intercept current, that will give us good clearance of our shot noise to dark noise (at least as measured in the <100 kHz band).  The question is if we can make a ~ 1 Mhz measurement of squeezing without extra noise in that readout chain dominating over our expected 1 dB squeezing.

The plan is to use the Zurich box (model HF2LI) to mix down quantum enhancement from 1 MHz in our homodyne detector.  This will be well above most of the classical noise of the laser and other acoustic pickup from the environment.  

The Zurich box has a input referred noise of 5 nV/rtHz above 10 kHz and operating at at the higher end of its sampling rate. It has a couple of nice features that would make it well suited to the task.  We can use the lock in amplifier instrument to digitally mix down signal from 1 MHz and output one or both of the i/q quadratures to its digital output ports.  From this instrument we can also set the bandwidth of the 

Output noise of Zurich is 25 nV/rtHz, so I'll need some gain factor to boost signal if readout is done from the external ports. Relative gain of the two paths can be changed manually by typing values into boxes, but might be tought to do a direct ballenced subtraction within the box live if the real time drift is large.

 

Laser diode modulation port on Diabolo has a signal bandwidth of 5 kHz, so can't really be used to generate a reference signal.  

  2331   Tue Apr 30 19:52:14 2019 awadeHowtoWOPOConfiguring Zurich box server

As I couldn't find this on the 40m elog/wiki I'm putting notes here on how to configure a fresh webserver + python server for the Zurich box (model HF2LI) that runs an autobooting service job. 

The Zurich box (model HF2LI) has no front panel interface.  All controls are done from a USB interface to a host computer (running windows or Ubuntu/Debian).  This computer then hosts a web-server that runs an interface in a browser (through http) or responds to requests from python/matlab/labview/C API over the network.  

Installing data and webserver 

Set up on a Linux box is fast and easy from a fresh install.  A thin low powered computer is ideal. Start with a fresh install of Debian, use the latest stable release. Then download the latest stable release for the HF2LI, HF2IS, HF2PLL instruments from the Zurich downloads page.  You’ll probably want the 64bit linux. You can untar this file with

> tar xzvf LabOneLinux<arch>-<release>-<revision>.tar.gz

(Insert your filename details above)

Then cd into the directory it created and run

> sudo bash install.sh

It will ask a bunch of questions, just hit enter and y all the way through. The install should immediately boot the HF2 data server at the end.  You can check the data server is running with

> ziService status

Now plug in your Zurich box using the USB cable. Note: there are ethernet ports on the box, but these are NOT for network ethernet.  If the box is powered up the data server should find the box.  

Running the web server

In another terminal you’ll want to to start a Webserver so you can view the box status, settings and live data.  Go to another terminal and run

> ziWebServer

Your Zurich device should now be exposed on port 8006 of that machine.  You are read to go.

Now on that computer you can got to http://localhost:8006, alternatively on any other computer on the network you can go to the host machine’s IP address and see your Zurich interface.  I.e. if the computer IP is 192.168.1.121 then go to http://192.168.1.128:8006.  

Remote access

If you wanted to access the Zurich interface remotely (outside network) you can do some SSH forwarding if you have a gateway machine to outside the lab.  The standard command for this is 

> ssh -nNT -L <clientport>:localhost:<hostport> <controls>@<Host-IP-Address>

where controls is the username, Host-IP-Address is the IP address of the machine to be forwarded from, clientport is the port on your machine (the one you are ssh'ing from) and hostport is the port on the remote machine (for Zurich this is 8006).  Then when you go to http://localhost:<clientport> and you'll see Zurich interface.

Python API

Since the release 2.18 (Nov 2018) the python access tools are now just in Pypi.  You can install them with pip:

> pip install --upgrade zhinst

This should now allow you to use python API to access your Zurich box.

 

Setting up service to keep host alive from when box is booted

You’ll want to set up a service on your Debian box with systemd.  Make a unit

> cd /etc/systemd/system

> sudo vim ziServer.service

You must use sudo to be able to save the file here.

Paste the following in, this assumes that your username and its group is called controls on your machine:

[Unit]
Description=Zurich Instruments HF2LI data server
After=syslog.target network.target

[Service]
User=controls
Group=controls
WorkingDirectory=/home/controls/services/
ExecStart=/bin/bash -c "exec ziServer"
Restart=no
RestartSec=30

[Install]
WantedBy=multi-user.target


Tip: in vim you can type “:set paste” to get exact paste without auto indent etc.

Save the file and exit.  You'll also need to create the working director, or change the above to your desired directory

> mkdir /home/controls/services/

Finally, tell systemd to look for new things:

> systemctl daemon-reload

and enable the service,

> systemctl enable ziServer.service

and manually start the service,

> systemctl start myservice.service

It should be running, try to see if this is so:

> ps | grep ziServer

It should run on reboot of the machine and restart every 30 seconds if it crashes for whatever reason. Try rebooting the host machine to check it is still running

Configuring web-server service

Follow the above steps to also auto load the webserver with a unit file named /etc/systemd/system/ziWebServer.service:

[Unit]
Description=Zurich Instruments HF2LI web server
After=syslog.target network.target

[Service]
User=controls
Group=controls
WorkingDirectory=/home/controls/services/
ExecStart=/bin/bash -c "exec ziWebServer"
Restart=no
RestartSec=30

[Install]
WantedBy=multi-user.target

 Now you have a local network ready server for accessing the Zurich box through a web browser.

  2330   Wed Apr 24 19:30:55 2019 ranaLab InfrastructureHVAChowto get temperature trends

We need to get the temperature data from the sensors that Aidan setup. I have got it by downloading a sketchy HOBO app onto my phone. In order for this to work you have to go stand next to the sensors and download via bluetooth. There is no "pairing" required - the app seems to recognize the sensors nearly immediately.

and please, Aidan and everyone else, please do not use double-sticky tape or glue or bubble gum to attach sensors in a clean optical setup. We're going to have to make a mess to remove these.

I've just downloaded more data (attached). The lab seemed sort of cold and the south HEPA output was blowing colder air than the north HEPA output.

I've changed the temperature from 70F to 73F at 7:45 PM today. Lets see how it goes.

Attachment 1: QIL_temperature_190424.tgz
  2329   Wed Apr 24 14:19:55 2019 awadeLab InfrastructureHVACHVAC QIL: two actuators tied to one sensor now

RXA: I talked with Modesto this morning. He had gotten the information at some point that our only problem was that we had 2 thermostats, and had not been told about the weird temperature fluctuations and the hot air. So I talked with him this morning so that he knows that we're going for stability in addition to using 1 sensor.

AWADE (Additional Info Wed Apr 24 19:42:32 2019 ): Modesto said that he set the temperature to 70 F.  

Modesto returned today to carry out the fix of the AC.  The south side hot actuator on the south vent was tied to the west wall thermostat.  So current status is that both heat/cooling actuators are tied to the one thermostat on the west wall.  Work was completed at 2 pm.

Modesto said he would be back tomorrow to check the system.

Quote:

Someone from Facilities named Modesto came out to the QIL to see the problem with the HVAC. He measured 80F coming out of the hot vent (when its thermostat was set to 68F) and 65F coming out of the cold vent. He was able to drop the hot vent to about 62F by turning the set point down to 55F (so the controller is at least working).

I asked that he set it up that both units are controlled by one temperature sensor. So he’s going to disconnect the hot vent controller from it’s thermostat and tie it into the thermostat for the cold side. This seems reasonable but could be potentially problematic if the HVAC units have significantly different responses.

Anyway, the current plan is for him to come in on Friday to do this work

 

  2328   Wed Apr 24 09:01:49 2019 AnjaliUpdate2micronLasersIntensity noise stabilization of 2 micron source

Attachment #1 shows the schematic of the experimental setup for amplitude stabilization using AOM. The proposed idea is as follows

  • The output power from the  Zeroth order port of the AOM depends on the RF power supplied to the AOM and the  RF power from the RF driver is dependent on the input voltage to the driver. 
  • When amplitude of the laser chnages to due to amplitude noise, the output from photodiode will also changes. The output of one of the photodiodes is connected through a gain element (G) to the RF driver of AOM. So, when PD output changes , the gain of the gain element should be adjusted such that the input voltage to the AOM driver changes and thus the RF power which affect the output power from the AOM. 
  • The gain element could be SR560 (I still have to understand more about it) and the photodiodes that we have are thorlabs DET10D detector which are DC coupled. We also need additional TIA stage after the PD. There were two of them built by the summer student. We tested both and one of them only works fine now.
Quote:

OK...but how will the amplitude stabilization be done? How about a diagram showing the feedback loop and electronics?

  1. what is the feedback circuit?
  2. Can it drive the low impedance of an AOM driver?
  3. What is the photodiode?
  4. It has to be an AC coupled feedback (DC coupled is always terrible).
  5. What is the requirement on the intensity noise as a function of frequency?
Quote:

[Aidan, Anjali]

  • Attachment #1 shows the cleaned up setup for  2 micron experiment

 

Attachment 1: Intensity_stabilisation.png
Intensity_stabilisation.png
  2327   Mon Apr 22 19:35:16 2019 anchal and awadeSummaryWOPOShot Noise Intercept Current measurement of Photodiodes

Andrew measured the output voltage noise of photodiodes through an AC coupled pre-amplifier (G=200, 0.7 nV/rtHz) while varying the intensity of white light falling on it. The setup uses a regular 2xAA flashlight (torch) with an incandescent bulb to illuminate the photodiode. The flashlight focus is adjusted to give a non-expanding beam and a 25.4 mm (1") lens is then used to focus the collimated beam down. The total power of light was controlled by trimming it with an iris.


Data Analysis:

Below is the fitting of data to estimate the shotnoise intercept current. Here the fitted function is the quadrature sum of input referred shot noise and dark noise modeled as an equivalent shot noise quanitiy i_snint (called intercept current):

i_\textrm{in} = \sqrt{2e^-(i_\textrm{pc}+i_\textrm{snint})}

I have included zip file with data and notebook in it as well. With measured shotnoise intercept current of 48 ± 7 uA and 31 ± 7 uA, we are good to go to use these for measuring shot noise with light power in the order of 5.0 mW falling on the photodiodes. With 5.0 mW (~3.9 mA) on each PD, this gives us a clearance of 10dB from the dark noise floor on average for each detector.  Plots with fits are attached below.

Attachment 2: WOPO_BHD_Photodiodes_SNInterceptAnalysis.pdf
WOPO_BHD_Photodiodes_SNInterceptAnalysis.pdf WOPO_BHD_Photodiodes_SNInterceptAnalysis.pdf
Attachment 3: 20190419_MeasuredInterceptCurve.zip
  2326   Thu Apr 11 20:50:17 2019 ranaHowtoGeneralhow to learn quickly

https://maglab.caltech.edu/human-magnetic-reception-laboratory/

  2325   Wed Apr 10 18:58:48 2019 ranaUpdate2micronLasersIntensity noise stabilization of 2 micron source

OK...but how will the amplitude stabilization be done? How about a diagram showing the feedback loop and electronics?

  1. what is the feedback circuit?
  2. Can it drive the low impedance of an AOM driver?
  3. What is the photodiode?
  4. It has to be an AC coupled feedback (DC coupled is always terrible).
  5. What is the requirement on the intensity noise as a function of frequency?
Quote:

[Aidan, Anjali]

  • Attachment #1 shows the cleaned up setup for  2 micron experiment
  2324   Wed Apr 10 13:10:45 2019 awadeDailyProgressWOPORebuild of WOPO homodyne detectors using AD829

[awade]

This post details the rebuild I made of the transimpedance amplifiers (TIA) for the homodyne detectors.  My previous iteration of this design failed to take into account the output swing limitations and current draw limitations at the output of the op amp.  It turns out that the OP27 was not the best choice here.  On the face of it its GBP and input referred noise were fine for a ~ 1 MHz detection.  However, operating the chip with 5V output voltage (to get 10 dB shot noise clearance from thermal noise) and with output impedance on order of 150 Ω meant that the frequency response was not as expected (see ATF:2316). 

A better choice is something like AD829 which is designed for driving signals directly into 50Ω terminated outputs.  The updated design is illustrated below.

Here the transimpedance gain has been lowered to 2 kΩ to lower the relative input referred noise contributed by the AD829's slightly higher current noise.  Here the optical power can be increased to give the same effective output signal.  In this case increase optical power and lowering gain keeps the thermal noise clearance below shot noise constant but improves the clearance of the AD829 input current noise.

 

Scaling of thermal noise relative to shot noise given DC output voltage

For reference the thermal noise of the TIA is given by the Johnson noise

\delta n_\mathrm{thR} = \sqrt{\frac{4k_BT}{R_\textrm{fb}}}

where k_B is the  Boltzmann's constant, T is the temperature (300K) and Rfb is the feedback resistor value.  Also here the shot noise is

\delta n_\mathrm{SN} = \sqrt{2 e^- \mathcal{R}P_\textrm{PD}}

Where e is the charge on an electron (1.602e-19 C), R is the photodiode responsivity (~0.8 A/W), and P_PD is the power on the photodetectors (0.5-5 mW).  Keeping in mind that there is a maximum voltage swing (V_max) on the output of the op amp, this comes from its current driving capability into a given load, then the feedback resistance (TIA gain) is limited to 

R_\textrm{fb} = \frac{V_\textrm{Max}}{\mathcal{R} P_\textrm{PD}}

The noise clearance, given a maximum output voltage, is then given by the ratio

NR = \frac{\delta n_\mathrm{SN}}{\delta n_\mathrm{thR}} = \sqrt{\frac{e^-V_\textrm{Max}}{2kT}}

Power on the photodiode and responsivity cancel out for the noise clearance.  Assuming we are not optical power limited then the thermal noise ratio (clearance below shot noise) is wholly determined by V_Max and the physical constants e, k_B and T.  For room temperature (300 K) this reduces to the rule of thumb 

NR \sim 4.4 \sqrt{V_\textrm{Max}}

Keeping the V_max to within 3 V means that we will get a thermal noise clearance factor of 7.6 below shot noise (8.8 dB). V_max of 5.2 volts will give 10 dB clearance. Ok I guess. 

Choosing AD829

Koji suggested the AD829 as a replacement to OP27.  AD829 has applications in driving 50 Ω/75 Ω loads in video applications and has some nice noise characteristics.  The bottom line is that it has 1.7 nV/√Hz input voltage noise, 1.5 pA/√Hz input current noise, can do a ±3 V voltage swing into 150 Ω load (DC coupled), and is fast (600 MHz uncompensated, with 230 V/µs slew).  There are some quirks.  AD829 seems have have a weird 80 MHz feature that causes oscillations if not compensated properly.  I did a bit of modeling in LISO and then just decided to build it once I found that ~2 kΩ gain was about right for ensuring that the dark noise wasn't dominated by the op amp current noise.

Because I was building on proto-board I wanted to ensure that there was as little parasitic capacitance as possible.  I built the whole amplifier + PD onto a single SOIC-14 to 0.1 pitch adaptor (Adafruit SMT Breakout PCB for SOIC-14, Part No. 1210), pictured below.  The small footprint and short trace lengths between critical pins means that there is a lower chance of parasitic impedance causing instability or impact on bandwidth.  

Top view of Unit B TIA with PD mount soldered directly to inverting pin and positive bias
TIA at earlier stage of construction for both unit A and unit B
shows feedback resistor and capacitor soldered on reverse side (these passive
components are stacked on top of each other).

The feedback resistor (2 kΩ) and feedback compensating capacitor (15 pF ceramic + 3 pF nominal pin parasitic) were soldered directly to the reverse side where the TSSOP-14 pins were connected through vias to the top side.  This minimized the path length of this electrical signal.  I scratched off pads that were unused in case they shorted or were a source of crosstalk.  The photodiode mount pin was soldered directly to the edge of the board to minimize distance.  The power bypassing capacitors were SMT 100 nF ceramics (5%) that were soldered directly to the SMT breakout with the shortest path to ground on that board (see pictured).  The other photodiode pin was soldered to +5 V (-5V) supply pin in unit A (unit B).  The opposite biasing makes subtracting the signals using a summing circuit easier but may affect the overall TF response. There is no power regulation, I'm planning to use batteries to power these circuits.

In my initial tests I had no compensating capacitor attached to pin 5.  I inserted a 40 pF ceramic capacitor in place of the photodiode and looked at the AD829 output pin with a high impedance probe on an oscilloscope.  I immediately saw 80 MHz oscillations.  I don't really need a super high bandwidth so I went strait to the maximum recommended choice for pin5 compensating capacitor of 68 pF.  This makes AD829 unity gain stable with 66 MHz bandwidth (slew 16V/µs) but is more than enough for my needs.  This killed the high frequency RF ringing junk. Maybe less capacitance here would have worked, but this performance is enough for my needs and gives some certainty about the op amps stability (ignoring input capacitance compensation).

At the output I added 100 Ω of series resistance to limit the loading on the op amp and current draw when 50 Ω terminated.  With 50 Ω terminating impedance this makes a 150 Ω to ground.  Providing the output DC swing is kept within 3V the op amp should behave as expected.

The whole thing fits together very nicely on a single SMT breakout board. Making it compact will hopefully avoid any issues with stray capacitance messing up the performance.  I have then mounted this on a larger proto board for ease of installing in the experiment. Paths to the output SMA output are thin wire and not routed through the underlying board. Only power and ground is routed through the board, all other pins to the op amp disconnected. 

 

Transfer function PD

I measured the signal transfer function of the above TIA units using the Jenne rig at the 40 m.  It took me a number of tries: I made the mistake of loading the DC port of the NF 1611 detector with 50 ohms (which resulted in too much current draw and affected the LF response of the witness detector); there were also some issues with the polarization/positioning onto the beam splitter there that made it not 50:50 (it was 69:33 ), I moved the fiber launch a bit to get the splitting right and retook measurements after that; also I initially had the current set too high on the laser current driver which I think was saturating its response.  I'll skip over the details and just present the final measurement. I've included a zip in the attachments with the data that logs some more of the details of my failed attempts.

I used the same calibration as used in PSL:2247.  Here the laser diode current was set to 25.0 mA, which is 1.51 mW of 1064 nm, and the AC excitation into the laser was set to -21 dBm (19.9 mV rms).  Power on the witness detector was measured to be 0.79 mW and that on the PD under test was 0.75 mW (as measured with a power meter).  Units of measurement were linear magnitude and phase in degrees. DC voltages were measured with 1 MΩ impedance.

Unit A: using 25 mA current into laser and -21 dBm sweep I get DC on PD of 1.20 V (into 1 MΩ) and a DC voltage of NF1611 detector of -2.13 V (into 1 MΩ).

Unit B: using 25 mA current into laser and -21 dBm sweep I get DC on PD of 1.07 V (into 1 MΩ) and a DC voltage of NF1611 detector of -2.12 V (into 1 MΩ). 

The calibrated TIA current to voltage gain is plotted below:

A notebook detailing all the attempts is attached below.

 

 

Quote:

This set of measurments turned out to be bad. The OP27 (±15V supplies) with 50 Ω series at the output + 50 Ω load impedance seemed to be saturated.

Attachment 1: BasicSchematic_TranImpPD_AD829.pdf
BasicSchematic_TranImpPD_AD829.pdf
Attachment 4: WOPO_HDPD_AD829_2kohmTranImpGain_UnitAandBandLISO.pdf
WOPO_HDPD_AD829_2kohmTranImpGain_UnitAandBandLISO.pdf
Attachment 5: 20190405_TFHDPDs_AD829Design_LargeBoard.zip
  2323   Wed Apr 10 12:17:34 2019 AnjaliUpdate2micronLasersIntensity noise stabilization of 2 micron source

[Aidan, Anjali]

  • Attachment #1 shows the cleaned up setup for  2 micron experiment
  • We also have received the AOM . The initial test is to do the intensity stabilisation of 2 micron laser diode using AOM
  • Today, we started with the laser and AOM characterisation
  • Attachment #2 shows the schematic of the experimental setup for the laser characterisation. Laser output is connected to a Faray isolator and the isolator output is measured using Thorlabs thermal power meter.
  • The current limit on the current controller is set to 95 mA (maximum allowed current to the laser diode is 120 mA) and thermistor is set to 10 kohm (maximum value of thermistor resistance is 10 kohm)
  • Attachment #3 shows the output power as a function of input current to the laser diode.
  • From the data sheet of the laser diode, the threshold current is 20 mA.For a given inut current,  the output power measured is less compared to the value given in the data sheet. This is beacuse of the insertion loss of the isolator in the path.
  • The slope efficieny is calculted to be 0.018 mW/mA. The value given in the data sheet is 0.3 mW/mA. The reduction is because of the insertion loss of the isolator
  • The output of isolator is then connected to the AOM. Attachement #4 shows the schematic of the setup. The zeroth order port of the AOM is connected to the power meter and the output power from the first order port is blocked.
  • We then applied a dc voltage to the modulation input port of the AOM driver. We expect a decrease in the power from the zeroth order port with increase in the input voltage to the  AOM driver as the light is getting deflcted and power is geting coupled to the first order port.
  • Attachmet #5 shows the variation in power from zeroth order port with input voltage to the AOM driver.As expected, the power levels are decreasing with increase in input voltage
  • It is also observed that the output power levels are fluctuating. This could be because of temperature fluctuation and the thermal power meter is sensitive to that. 
Attachment 1: 2_micron_setup.jpg
2_micron_setup.jpg
Attachment 2: Laser_chara_setup.png
Laser_chara_setup.png
Attachment 3: Laser_characteristics.pdf
Laser_characteristics.pdf
Attachment 4: AOM_chara_setup.png
AOM_chara_setup.png
Attachment 5: AOM_characteristics.pdf
AOM_characteristics.pdf
  2322   Mon Apr 8 14:02:06 2019 AidanLab InfrastructureLab MonitoringHOBO MX100 temperature sensors/data loggers added to the optical tables

Here is the last 6 days.

Quote:

Here is the first set of data from the last 24 hours or so. The delta T  on the North table is around 5C versus 2C on the South table.

Quote:

I added a Bluetooth temperature sensor/data logger to each optical table. They're tucked INSIDE the NE corner of the enclosure structure - shielding them from the direct blast from the HVAC. They're recording the temperature once every 5 minutes. The came with double-sided tape to stick them into position.

The data can be accessed from your phone (I don't know yet if downloading the data clears the memory or not).

The idea is to gather data on the temperature of the two tables over the next few weeks.

Photos and instructions are attached.

 

 

Attachment 1: QIL_lab_temperature_6days.pdf
QIL_lab_temperature_6days.pdf
Attachment 2: QIL_South_2019-04-08_13_39_33_-0700.txt
Serial Number:	20572037 --

Date Time, GMT -0700	Temp, (*F)	Host Connect	EOF	
2019-04-02 14:05:45	78.25	 	 	
2019-04-02 14:10:45	76.47	 	 	
2019-04-02 14:15:45	71.84	 	 	
2019-04-02 14:20:45	70.68	 	 	
2019-04-02 14:25:45	70.37	 	 	
2019-04-02 14:30:45	70.30	 	 	
2019-04-02 14:35:45	70.37	 	 	
... 1720 more lines ...
Attachment 3: QIL_North_2019-04-08_13_39_16_-0700.txt
Serial Number:	20572038 --

Date Time, GMT -0700	Temp, (*F)	Host Connect	EOF	
2019-04-02 14:07:53	78.71	 	 	
2019-04-02 14:12:53	74.70	 	 	
2019-04-02 14:17:53	72.84	 	 	
2019-04-02 14:22:53	72.30	 	 	
2019-04-02 14:27:53	72.15	 	 	
2019-04-02 14:32:53	72.23	 	 	
2019-04-02 14:37:53	72.84	 	 	
... 1719 more lines ...
  2321   Wed Apr 3 12:51:55 2019 AidanLab InfrastructureLab MonitoringHOBO MX100 temperature sensors/data loggers added to the optical tables

Here is the first set of data from the last 24 hours or so. The delta T  on the North table is around 5C versus 2C on the South table.

Quote:

I added a Bluetooth temperature sensor/data logger to each optical table. They're tucked INSIDE the NE corner of the enclosure structure - shielding them from the direct blast from the HVAC. They're recording the temperature once every 5 minutes. The came with double-sided tape to stick them into position.

The data can be accessed from your phone (I don't know yet if downloading the data clears the memory or not).

The idea is to gather data on the temperature of the two tables over the next few weeks.

Photos and instructions are attached.

 

Attachment 1: QIL_North_2019-04-03_12_46_44_-0700.csv
Serial Number:20572038 --

"Date Time, GMT -0700","Temp, (*F)","Host Connect","EOF"
2019-04-02 14:07:53,78.71, , 
2019-04-02 14:12:53,74.70, , 
2019-04-02 14:17:53,72.84, , 
2019-04-02 14:22:53,72.30, , 
2019-04-02 14:27:53,72.15, , 
2019-04-02 14:32:53,72.23, , 
2019-04-02 14:37:53,72.84, , 
... 267 more lines ...
Attachment 2: QIL_South_2019-04-03_12_46_11_-0700.csv
Serial Number:20572037 --

"Date Time, GMT -0700","Temp, (*F)","Host Connect","EOF"
2019-04-02 14:05:45,78.25, , 
2019-04-02 14:10:45,76.47, , 
2019-04-02 14:15:45,71.84, , 
2019-04-02 14:20:45,70.68, , 
2019-04-02 14:25:45,70.37, , 
2019-04-02 14:30:45,70.30, , 
2019-04-02 14:35:45,70.37, , 
... 268 more lines ...
Attachment 3: QILtemperature.png
QILtemperature.png
  2320   Tue Apr 2 14:16:00 2019 AidanLab InfrastructureLab MonitoringHOBO MX100 temperature sensors/data loggers added to the optical tables

I added a Bluetooth temperature sensor/data logger to each optical table. They're tucked INSIDE the NE corner of the enclosure structure - shielding them from the direct blast from the HVAC. They're recording the temperature once every 5 minutes. The came with double-sided tape to stick them into position.

The data can be accessed from your phone (I don't know yet if downloading the data clears the memory or not).

The idea is to gather data on the temperature of the two tables over the next few weeks.

Photos and instructions are attached.

Attachment 1: IMG_9165.jpg
IMG_9165.jpg
Attachment 2: IMG_9158.jpg
IMG_9158.jpg
Attachment 3: IMG_9159.jpg
IMG_9159.jpg
Attachment 4: IMG_9160.jpg
IMG_9160.jpg
Attachment 5: IMG_9162.jpg
IMG_9162.jpg
Attachment 6: IMG_9163.jpg
IMG_9163.jpg
Attachment 7: IMG_9164.jpg
IMG_9164.jpg
  2319   Mon Apr 1 11:03:24 2019 JonUpdateComputingCymac assembly started

I know there is some CTN slow channel data on the disk. Is it at all possible to boot so that can be recovered?

Quote:

This afternoon Chris and I installed the ADC and DAC cards in fb4. We connected them to the timing card adapters (left external to the computer chassis for now).

We found fb4 to be running Debian 8 so first attempted to upgrade to 9, as that is the version supported by Jamie's cymac binaries. However, we encountered problems during the upgrade, apparently with gdm (the linux GUI). By switching to consol mode and killing gdm, were able to proceed to the point of updating all the packages. It completed successfully, but then the system failed to reboot, even in recovery mode. During boot, the advligo-rts kernel fails to start, and then boot hangs completely at the point the graphical interface is started.

We may want to start with a fresh install of Debian 9 and just reinstall the LIGO binaries.

 

  2318   Fri Mar 29 16:00:21 2019 awadeDailyProgressWOPORemeasuring TF Homodyne Photo Detector B after switching out foil cap

This set of measurments turned out to be bad. The OP27 (±15V supplies) with 50 Ω series at the output + 50 Ω load impedance seemed to be saturated.

The rest of this post is a stream of conciseness.

As noted in previous post (ATF:2316) the PD unit B in the homodyne had a slightly reduced HF response, possibly due to the use of a foil cap in the PD bias supply low pass.  I removed this and resmeasured.

Unit B TF: measured NF1611 DC voltage of  312 mV (@ 50Ω impedance), PD DC output voltage was - 566 mV (@ 50 Ω impedance).  Here the photodetector has transresistance is 6.8 kΩ. The NF1611 detectors have a DC path gain of 10 kΩ and an AC path gain of 700 Ω. From this the R_PC is xxx.  Here I ignore the overall phase (at ~1 MHz as the pi phase shift length is 150 m, so negligible).  This was a bad measurment, something was up with the circuit saturating way too low.  

Changed 100 Ω series on output to 50 Ω to make it more strait forward to understand how it works with 50 Ω load. 

Unit A TF: measured NF1611 DC voltage of  658 mV (@ 50Ω impedance), PD DC output voltage was 1.42 mV (@ 50 Ω impedance).  Here the photodetector has transresistance of 6.8 kΩ. The NF1611 detectors have a DC path gain of 10 kΩ and an AC path gain of 700 Ω. From this the R_PC is 0.3151.  Here I ignore the overall phase (at ~1 MHz as the pi phase shift length is 150 m, so negligible).  This measurment could have be saturating the current output of the OP27, likely that numbers here are not right.

After examining the output of detector B it became apparent that the output was saturated for powers of more than ~20 mA laser diode drive. 

Unit B TF: measured NF1611 DC voltage of  312 mV (@ 50Ω impedance), PD DC output voltage was -566 mV (@ 50 Ω impedance).  Here the photodetector has transresistance is 6.8 kΩ. The NF1611 detectors have a DC path gain of 10 kΩ and an AC path gain of 700 Ω. From this the R_PC is -0.3748.  Here I ignore the overall phase (at ~1 MHz as the pi phase shift length is 150 m, so negligible).   100 Ω  (50Ω + 50 Ω) loading is too low, OP27 open loop gain and voltage swing range slew starts to sag below 150Ω loading.  These results are bogus.

This time I measured with 50 Ω on the oscilloscope for apples-to-apples comparison with what the Aglient 4395A was seeing.  Turns out I think I was saturating the photodetector current wise in the first round of measurments. All of the above is probably not right and should be disregarded.  Also, later I switched to a AD829 as the op amp of choice so the above is not much use for characterizing the detector performance anymore.  The OP27 isn't really spec'ed to operate with loads this low and only has its full band with for output loads of >2 kΩ.  

  2317   Thu Mar 28 18:16:59 2019 JonUpdateComputingCymac assembly started

This afternoon Chris and I installed the ADC and DAC cards in fb4.

  • We connected them to the timing card adapters (left external to the computer chassis for now).
  • We found fb4 to be running Debian 8 so first attempted to upgrade to 9, as that is the version supported by Jamie's cymac binaries.
  • However, we encountered problems during the upgrade, apparently with gdm (the linux GUI). 
  • By switching to consol mode and killing gdm, were able to proceed to the point of updating all the packages.
  • It completed successfully, but then the system failed to reboot, even in recovery mode.
  • During boot, the advligo-rts kernel fails to start, and then boot hangs completely at the point the graphical interface is started.

We may want to start with a fresh install of Debian 9 and just reinstall the LIGO binaries.

  2316   Wed Mar 27 20:37:38 2019 awadeDailyProgressWOPOMeasuring TF Homodyne Photo Detector for WOPO experiment

Initial measurment of PD TF

I realize I never really measured the signal transfer function for each of these Photo Detectors (PD).  This post summarizes measurements of the optical to electrical signal output signal transfer function. Here I used the Jenne rig at the 40m and took a transfer function using an Aglent 4395A from 30 kHz to 5 MHz.  

I've labeled the two PD in the homodyne as unit A and unit B.  Below (first attached)  is the raw TF as measured by the Aglient taking the ratio of the detector to the reference NF1611 detector:

Fixing differences between output impedance unit A and B

After looking a bit at the circuit I realized that I had added 100 Ω in series in detector B (to limit current draw when driving 50 Ω loads).  This had not been added to the detector A, which means that with more power than 300 µW on this detector the op amp would be drawing more than 30 mA when driving a 50 Ω.  This shouldn't have been an issue when driving the summing circuit, but is good practice.  

I added 100 Ω in series with op27 output (thin film 1206 size) in detector A to match the output to that of detector B.  I remeasured the TF and get a much better match between the paths.

 

Calibrating TIA Measurment

The following is used to calibrate the  transfer function (See PSL:2247):

Z_\textrm{AC,PD} = Z_\textrm{AC,Ref} R_\textrm{PC}e^{i\delta\phi} T_\textrm{meas}

where ZAC,PD is Calculated RF Transimpedance, Z AC,Ref is the known RF transimpedance of the reference photodiode, 𝛿ϕ is arbitrary phase delay due to light and cable length and RPC  is the photocurrent ratio at DC of reference PD to RFPD under test given by

R_\textrm{PC} = \frac{V_\textrm{DC,Ref}/Z_\textrm{DC,Ref}}{V_\textrm{DC,PD}/Z_\textrm{DC,PD}}

For these measurements the calibration factors were as follows:

Unit A TF (after adding 100 Ω series​ output): measured NF1611 DC voltage of 780 mV (@ 1 MΩ impedance), PD DC output voltage was -1.20 V (@ 1 MΩ impedance).  Here the photodetector has transresistance is 6.8 kΩ. The NF1611 detectors have a DC path gain of 10 kΩ and an AC path gain of 700 Ω. From this the R_PC is 0.4496.  Here I ignore the overall phase (at ~1 MHz as the pi phase shift length is 150 m, so negligible).  

Unit B TF: measured NF1611 DC voltage of 791 mV (@ 1 MΩ impedance), PD DC output voltage was -1.18 V (@ 1 MΩ impedance).  Here the photodetector has transresistance is 6.8 kΩ. The NF1611 detectors have a DC path gain of 10 kΩ and an AC path gain of 700 Ω. From this the R_PC is 0.4558.  Here I ignore the overall phase (at ~1 MHz as the pi phase shift length is 150 m, so negligible).  

 

Results of calibrated transimpedance measurements are plotted below*. I also added the LISO estimated noise curve.  The 3 dB of unit A and B were 1.93 MHz and 1.72 MHz respectively.  There is some discrepancy with the LISO model here, some of this might be artifacts in the reference detector and some is likely dirt effects in the proto board circuit that I'm not going to debug for now.  The biggest discrepancy between the two detectors starts around 900 kHz.  When I looked again at the detectors I realized I had used a Wima foil cap in detector B for the bias LPF. It should be fine for these frequencies, but I switched the foil cap out and replaced it with a 1 µF + 100 nF ceramic cap to match the unit A.  I haven't gotten a chance to remeasure this since as Gautam is still using the Jenne rig to do some stuff. I'll remeasure later, we should then find the two responses match.

 

*I didn't do the many IRIS measurement​ uncertainty analysis, its not necessary at this stage. 

 

Attachment 1: TFUnitAandB_BeforeMatchingOutImp.pdf
TFUnitAandB_BeforeMatchingOutImp.pdf
Attachment 2: TFUnitAandB_FixingDetectorAImp.pdf
TFUnitAandB_FixingDetectorAImp.pdf
Attachment 3: WOPO_HD_TranImpGain_UnitAandBandLISO.pdf
WOPO_HD_TranImpGain_UnitAandBandLISO.pdf
  2314   Wed Mar 27 11:27:00 2019 AidanLab InfrastructureHVACHVAC temperature gradients in the QIL

Someone from Facilities named Modesto came out to the QIL to see the problem with the HVAC. He measured 80F coming out of the hot vent (when its thermostat was set to 68F) and 65F coming out of the cold vent. He was able to drop the hot vent to about 62F by turning the set point down to 55F (so the controller is at least working).

I asked that he set it up that both units are controlled by one temperature sensor. So he’s going to disconnect the hot vent controller from it’s thermostat and tie it into the thermostat for the cold side. This seems reasonable but could be potentially problematic if the HVAC units have significantly different responses.

Anyway, the current plan is for him to come in on Friday to do this work

  2313   Tue Mar 26 17:13:53 2019 AnjaliUpdate2micronLasersFrequency stabilization of 2 micron source

Experimental plan 

  • The aim is to repeat the homodyne measurement setup with  balanced detection scheme for the frequency stabilisation of 2 micron laser source.
  • Attachment # 1 shows the schematic of the experimental setup.
  • All the components specified in the schematic are available.
  • For the balanced detection, we can initialy use SR560, but we need to design a new circuitry for the same.
Attachment 1: Homodyne_setup_for_2_micron.pdf
Homodyne_setup_for_2_micron.pdf
  2312   Mon Mar 25 09:40:16 2019 AnjaliNoise Budget2micronLasersNoise budget for the frequency discriminator
  • Different fiber parameters are very important in determining the thermal noise of the fiber. I was considering SM1950 fiber for the calculations, but all the characteristic parameters for this fiber are not known.
  • Glenn ( IEEE Journal of Quantum Electronics25(6), 1218-1224 ) and Jing Dong (Applied Physics Letters108(2), 021108) had given the set of fiber parameters. Attachment #1 shows the comparison of thermoconductive noise of the fiber with fiber parameters taken from Glenn and Jing Dong. The parameters from Jing Dong are for Corning standard SMF 28 fibers and they are the recent resuts compared to Glenn. Hence I think it would be good to follow the parameters given in Jing Dong's paper.
  • It is also found that the thermal noise of fiber has contributions from the thermoconductive noise (https://ieeexplore.ieee.org/stamp/stamp.jsp?tp=&arnumber=255919) and thermomechanical noise (https://ieeexplore.ieee.org/stamp/stamp.jsp?tp=&arnumber=5614023) . Thermoconductive noise  (in rad/rtHz) predicts a frequency independent characteristics at lower frequencies where as thermomechanical noise (in rad/rtHz) has a 1/f characteristics over a certain frequency range. The thermomechanical noise is of concern only over a certain frequency range it is given by the ratio of velocity of sound with the total length of the fiber. This indicates that, in the case of thermomechanical noise, the fluctuation of concern is much slower than the time it takes for the sound wave to travel across the length of the fiber. 
  • Attachment #2 shows the noise budget including the thermomechanical noise of the fiber. With 10 m length of the fiber, the thermoconductive noise is of concern for frequencies lower than 364 Hz . Thus a low pass transfer function is assumed with corner frquency of 364 Hz in the calculation of thermomechanical noise. 
Attachment 1: Comparison_of_thermoconductive_noise_in_Hz_per_rtHz.pdf
Comparison_of_thermoconductive_noise_in_Hz_per_rtHz.pdf
Attachment 2: Noise_budget_with_thermomechanical_noise.pdf
Noise_budget_with_thermomechanical_noise.pdf
  2311   Fri Mar 22 10:00:30 2019 AnjaliNoise Budget2micronLasersNoise budget for the frequency discriminator

Attachement #1 shows the modified noise budget with modification on dark current noise.

I still have to find out the PSD for room temperature to plot the room thermal noise

 

Attachment 1: Noise_budget_modified.pdf
Noise_budget_modified.pdf
  2310   Thu Mar 21 09:43:19 2019 AnjaliNoise Budget2micronLasersNoise budget for the frequency discriminator

Also, I have a doubt that whether the transfer function should be of low pass in nature. So, if the laser phase is fluctuating faster than the time it takes to propagate throught the delay fiber, we will not be able to discriminate .

If that is the case, the noise budget looks like as shown in attachment #1

Attachment 1: Noise_budget_with_low_pass_transfer_function.pdf
Noise_budget_with_low_pass_transfer_function.pdf
  2309   Thu Mar 21 00:49:35 2019 AnjaliNoise Budget2micronLasersNoise budget for the frequency discriminator

I didn't use the transfer funtion in the calculation of dark current noise thinking dark current noise is a characteristic of the detector alone

Regarding room thermal fluctuation, I was plotting the thermal noise of the detector which is given in A/rtHz as 

=\sqrt{\frac{4k_bT}{R_f}}

  2308   Wed Mar 20 19:17:50 2019 ranaNoise Budget2micronLasersNoise budget for the frequency discriminator

I think the dark noise should be very close to the shot noise and also have the same transfer function.

Also, the room temperature fluctuations are probably large at low frequencies. What PSD for room temperature did you use?

 

  2307   Wed Mar 20 15:57:55 2019 AnjaliNoise Budget2micronLasersNoise budget for the frequency discriminator

I was trying to prepare the noise budget for the frequency stabilisation setup for 2 micron laser. In this test, we use a fiber based Mach-Zehnder interferometer as a frequency discriminator to convert frequency noise to amplitude(voltage) fluctuations. The different noise sources to be considered in this analysis are the following

  1. Shot noise
  2. Dark current noise
  3. Room thermal noise
  4. Fiber thermal noise
  5. Fiber acoustic noise
  6. Laser intensity noise
  7. Photothermal noise
  • Attachment #1 shows the magnitude spectrum of the transfer fucntion for the frequency discriminator. I assumed the delay fiber length in one of the arms of the Mach-Zehnder interferometer is 10 m. This corresponds to a time delay of 50 ns. The corner frequency is inversely proportional to the time delay.
  • The transfer fucntion can be interpreted as, if the frequency variation is very slow- it can't be distiguished by the discriminator and we dont get any signal.
  • Attachment # 2 and # 3 shows the shot noise and fiber thermal noise respectively in rad/rtHz
  • The shot noise is calculated  considering thorlabs DET10D detector with a responsivity of 1.2 A/W at 2 micron. The laser output power from Eblana laser is assumed to be 2 mW and the power reaching the detector is estimated to be 1.23 mW.
  • The fiber thermal noise is calculated based on Wanser, K. H. (1992). Fundamental phase noise limit in optical fibres due to temperature fluctuations. Electronics letters28(1), 53-54.
  • The dark current noise and room thermal noise are calculted to be a constant value (3.97x10-12 rad/rtHz and 1.20x10-12 rad/rtHz respectively)- but this is not correct
  • The noise in rad/rtHz is converted to Hz/rtHz by mutiplying with the frequency (f) and it is shown in attachment #4
Attachment 1: High_pass_transfer_function.pdf
High_pass_transfer_function.pdf
Attachment 2: Shot_noise.pdf
Shot_noise.pdf
Attachment 3: Thermoconductive_noise_of_fiber_Glenn_parameter.pdf
Thermoconductive_noise_of_fiber_Glenn_parameter.pdf
Attachment 4: noise_budget_combined_log_log.png
noise_budget_combined_log_log.png
  2306   Wed Mar 13 17:49:14 2019 awadeSummaryWOPODamage threshold estimates for optical fibers

This is just a note about damage tolerances for fibers so we have a reference of the amount of power that can be used.  

Thorlabs gives the maximum theoretical CW power as 1 MW/cm² and a 'practical' safe power level of 250 kW/cm² (or a quarter of the max).  They don't seem to provide information about wavelength dependence and assume its the same for all wavelengths.  We don't expect to be affected by any of the exotic ultra high power effects like bend loss induced damage and photodarkening.  The air/glass interface is where the damage will occur.  This can either be because of heating of the ferrule/connector (causing epoxies etc to break down and damage the interface by depositing on the optical surface) or regular mechanism that are the same as bulk optics (dielectric break down and thermal effects).  

The intensity profile of light confined in the fiber is defined by the Mode Field Diameter (MFD) -- the cross-sectional diameter of the light that includes the core of the fiber and a region just beyond the cladding the mode occupies. MFD of 1064 PM fiber (PM980-XP) is 6.6 ± 0.5 μm @ 980 nm and for the 532 nm PM fiber (PM460-HP) is 3.3 ± 0.5 µm @ 515 nm.  

Fiber effective area is 

A = \pi (\textrm{MDF}/2)^2

which is 8.5x10^-8 cm^2 for PM460-HP and 3.4x10^-7 cm^2 for PM980-XP.  Taking the conservative 'practical' damage threshold this indicates a maximum power of 21.4 mW into the 532 nm fiber and 85.5 mW into the 1064 nm PM fiber. The absolute maximum is just a factor of four more than this: 85.6 mW into 532 nm PM fiber and 342 mW into 1064 nm fiber. If the fiber ends are kept clean then we should be fine if the power level is kept below 85 mW.

 

Fiber beam splitter

Of particular concern is the power handling capability of the fiber beam splitter (PN1064R5A2).  There will be some waste 532 nm light coming out of the WOPO and I don't want these potentially multi-mode components to exit the cladding at the point of the coupler and damage the surrounding material. The 1064 nm maximum power rating of this 50:50 PM beam splitter is listed as 1 W (for the connectorized fiber), so we should be well clear of that threshold for the LO light.  For 532 nm its less clear. The equivalent 532 nm PM 50:50 beam splitter (PN530R5A2) has a rated power of 100 mW @ 530 nm for bare or connectorized fibers. As the MFD of the 1064 nm version of this PM beam splitter has a much larger MFD and the exiting 532 nm light will already be expanded in the cable patching the WOPO to the BS, we should be well clear of this damage threshold point.  

So bottom line is that we need to keep power below 85 mW going into the WOPO device and keep all the end connectors super clean and it will be fine.

 

 

  2305   Wed Mar 13 12:44:41 2019 awade, anchalDailyProgressWOPOPol launch into PM fiber 1064 nm

[awade, anchal ]

After a bit of reading I've realized that the standard use of these PM fibers is to launch along the slow axis (see for example Thorlabs and OzOptics info on fiber beam splitters).  It should be much of the sameness for patch cables, but polarization sensitive elements like beam splitters are mostly tested and specified for slow axis launch unless they are custom made to order. 

We are switching the polarization alignment to slow axis in the 1064 nm and 532 nm fiber coupling.  Anchal is re-optimizing​ the 1064 nm launch to get the PM fiber extinction ratio back to a good place.  We've also changed input launch to use a laser line PBS mounted in a rotation mount for clean linear polarization.  With the optimized setup the for the 1064 nm fiber path the output polarization signal goes from 3700 mV to 39.3 mV which is an extinction​ ratio of -19.7 dB.

Here the max theoretical extinction​ ratio is 

ER = -10 \log_{10}[\tan^2(\theta)]

which would place our goodness of alignment to with 0.61 deg.

Updated 1064 nm launch. Uses rotation mounted PBS for guaranteed linear
polarization, half wave plate is to maximize power.

 

 

Quote:

I've replaced the SM fiber in the 1064 nm launch with a PM fiber (P3-1064PM-FC-5). I also moved the fiber collimator (F240APC-1064) back 2.54 cm back to give more space for a PBS cube (to check linearly of the light).  

For the 1064 nm launch it seemed to be a lot harder to find the initial alignment of the collimator using the alignment of the back propagated 650 nm fiber laser source. Here I aligned a pair of irises in the forward propagating direction and then back propagated through the PM fiber using 650 nm to get the initial​ pointing of collimator. I don't know why this is so much harder than the 532 nm case.  I suspect one of the steering mirrors is not really reflecting off the front dielectric surface.  In the end I did a bunch of systematic walking of the fiber launch mount and eventually fount the alignment.  

From 4.44 mW of input light I get 2.74 mW of light out the other end of the fiber.  This is an efficiency of 62 % which is more than enough for my needs.  I expect the HD will only need 1 mW (2 mW max), so this is fine. Getting this in coupling higher will require a bit of lens walking, not really worth it at this stage.

I had already carefully aligned the collimator orientation to put the fast axis on aligned to p-pol (wrt the table), by eye.  It seems like the launch pretty much hit the correct launch polarization on the first go.  I see little variation in the polarization when I pulse the heat on the fiber.  This is now good to go for optimizing the homodyne visible and polarization overlap output from the SQZ.

 

  2304   Tue Mar 12 16:58:50 2019 awadeMiscWOPOTodo list W11

These are things to get do this week on WOPO experiment:

☑️ Reinstall 50:50 fiber splitter into homodyne setup and go back to fiber launch of both ends of the HD directly onto photodetectors (rather than free space)

☐ Check visibility of HD by launching 1064 nm into both arms of HD using splitter and extra length on one arm, ramp laser frequency to get fringes and look at Vpp on each diode seperatly

☑️ Optimize subtraction of HD for max Common Mode Rejection (CMR) of LO amplitude noise.  Inject 3.21 kHz line into laser BNC port and minimize this peak on the subtracted output of the HD.

☐ Check 1064 nm -> 532 nm conversion in WOPO device to establish polarization basis for correct pol alignment into fiber (change 532 nm launch polarization if necessary), should be along the fast axis but for some reason this isn't in the datasheet explictly

☐ Double check how much power change there is from 2 pi modulation depth from PZT mounted 1064 nm mirror.  We don't want to over actuate on this element as it slightly misaignes into the fiber launch and causes some change in power, this is ok over a small range as the CMR of the homodyne will reject this.  We just want to be sure that this isn't the dominant effect that we are seeing at the output once we thing we should be seeing SQZ.

 

☑️Tick mark when done.

  2303   Tue Mar 12 16:35:43 2019 awadeDailyProgressWOPOPol launch into PM fiber 1064 nm

I've replaced the SM fiber in the 1064 nm launch with a PM fiber (P3-1064PM-FC-5). I also moved the fiber collimator (F240APC-1064) back 2.54 cm back to give more space for a PBS cube (to check linearly of the light).  

For the 1064 nm launch it seemed to be a lot harder to find the initial alignment of the collimator using the alignment of the back propagated 650 nm fiber laser source. Here I aligned a pair of irises in the forward propagating direction and then back propagated through the PM fiber using 650 nm to get the initial​ pointing of collimator. I don't know why this is so much harder than the 532 nm case.  I suspect one of the steering mirrors is not really reflecting off the front dielectric surface.  In the end I did a bunch of systematic walking of the fiber launch mount and eventually fount the alignment.  

From 4.44 mW of input light I get 2.74 mW of light out the other end of the fiber.  This is an efficiency of 62 % which is more than enough for my needs.  I expect the HD will only need 1 mW (2 mW max), so this is fine. Getting this in coupling higher will require a bit of lens walking, not really worth it at this stage.

I had already carefully aligned the collimator orientation to put the fast axis on aligned to p-pol (wrt the table), by eye.  It seems like the launch pretty much hit the correct launch polarization on the first go.  I see little variation in the polarization when I pulse the heat on the fiber.  This is now good to go for optimizing the homodyne visible and polarization overlap output from the SQZ.

ELOG V3.1.3-