It looks like the laser frequency noise will be slightly too high to measure the vibration of the mirrors in the cavity using the control signal for the locking, especially if our cavity is quieter than the MZ at the 40m (which we are hoping it will be).  It would still be nice to get a measurement of the mechanical noise of the mirrors somehow.

 

Quote:

 Here is a plot of laser frequency noise from a Lightwave NPRO 126 laser converted into m*Hz-1/2 for comparison with mechanical noise taken from the MZ at the 40m.

 