Fri Jun 11 20:33:17 2010, Dmass, Computing, DAQ, Attempt at oven transfer functions
|
Wed Jun 16 15:54:20 2010, Dmass, Computing, DAQ, Attempt at oven transfer functions
|
Wed Jun 16 23:10:28 2010, Dmass, Computing, DAQ, Attempt at oven transfer functions
|
Thu Jun 17 15:14:58 2010, Dmass, Computing, DAQ, Attempt at oven transfer functions
|
Fri Jun 18 18:04:23 2010, Dmass, Computing, DAQ, Attempt at oven transfer functions
|
Fri Jun 18 18:07:53 2010, Dmass, Computing, DAQ, Attempt at oven transfer functions
|
|
Message ID: 818
Entry time: Thu Jun 17 15:14:58 2010
In reply to: 816
Reply to this: 820
|
Author: |
Dmass |
Type: |
Computing |
Category: |
DAQ |
Subject: |
Attempt at oven transfer functions |
|
|
And a measurement of the oven's step response
- I used a lab power supply as a current source for the TEC on the oven
- I tweaked the current nob to make a current step
- I looked at the step response
- I used a phone video of the readout on the front of the Newport 3040 to get time stamps b/c it was so easy
- (I will still be hooking up the RS232)
Results
- I get a time constant of 60 seconds
- This is an oven pole frequency of 16.7 mHz
|
|
|