40m
QIL
Cryo_Lab
CTN
SUS_Lab
CAML
OMC_Lab
CRIME_Lab
FEA
ENG_Labs
OptContFac
Mariner
WBEEShop
|
QIL elog |
Not logged in |
 |
|
Mon Jun 6 09:45:58 2022, Radhika, DailyProgress, Emissivity estimation, Megastat cooldown - repainted inner shield aquadag, added aquadag foil to cold plate 6x
|
Wed Jun 8 12:15:06 2022, Radhika, DailyProgress, Emissivity estimation, Megastat cooldown - repainted inner shield aquadag, added aquadag foil to cold plate
|
Thu Jun 9 10:42:26 2022, Radhika, DailyProgress, Emissivity estimation, Megastat cooldown - repainted inner shield aquadag, added aquadag foil to cold plate
|
Fri Jun 10 09:47:07 2022, Radhika, DailyProgress, Emissivity estimation, Megastat cooldown - repainted inner shield aquadag, added aquadag foil to cold plate
|
Fri Jun 10 16:23:04 2022, rana, DailyProgress, Emissivity estimation, Megastat cooldown - repainted inner shield aquadag, added aquadag foil to cold plate
|
|
Message ID: 2784
Entry time: Fri Jun 10 09:47:07 2022
In reply to: 2783
Reply to this: 2785
|
Author: |
Radhika |
Type: |
DailyProgress |
Category: |
Emissivity estimation |
Subject: |
Megastat cooldown - repainted inner shield aquadag, added aquadag foil to cold plate |
|
|
I looked closer at the RTDs, and they are all PT100s with the same resistance-temperature curve [Attachment 1]. So we might need to re-calibrate them.