40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  ATF eLog  Not logged in ELOG logo
Entry  Mon Oct 25 16:13:47 2021, Aidan, Update, Cryo vacuum chamber, Heater left on - chamber got warm Screenshot_from_2021-10-25_16-19-39.png
    Reply  Tue Oct 26 07:32:52 2021, Aidan, Update, Cryo vacuum chamber, Heater left on - chamber got warm 
       Reply  Tue Oct 26 08:43:38 2021, Aidan, Update, Cryo vacuum chamber, CTC100 temperature alarm and heater shutoff Screenshot_from_2021-10-26_08-42-57.pngIMG_5371.jpgIMG_5396.jpgIMG_5401.jpg
Message ID: 2690     Entry time: Tue Oct 26 08:43:38 2021     In reply to: 2689
Author: Aidan 
Type: Update 
Category: Cryo vacuum chamber 
Subject: CTC100 temperature alarm and heater shutoff 

Instructions on how to enable the alarm and heater shut off for the CTC100.

Status: This reports the status of the alarm. If LATCH is enabled, this must be manually set to OFF once it has been enabled.

Mode: Set to "Level"

Latch: Optional to set to "YES" if desired.

Output: Set to "Heater"

Max: Set to desired maximum temperature.

The attached photos show:

  • the menu where the settings are ALARM entered
  • the main display just before the alarm is enabled (at 300.350K with a 1s delay)
  • the main display just after the alarm is enabled - note that the Heater Output has been set to 0W.
Attachment 1: Screenshot_from_2021-10-26_08-42-57.png  74 kB  Uploaded Tue Oct 26 09:44:01 2021  | Hide | Hide all | Show all
Screenshot_from_2021-10-26_08-42-57.png
Attachment 2: IMG_5371.jpg  3.819 MB  Uploaded Tue Oct 26 10:13:57 2021  | Hide | Hide all | Show all
IMG_5371.jpg
Attachment 3: IMG_5396.jpg  2.739 MB  Uploaded Tue Oct 26 10:14:07 2021  | Show | Hide all | Show all
Attachment 4: IMG_5401.jpg  2.561 MB  Uploaded Tue Oct 26 10:17:11 2021  | Hide | Hide all | Show all
IMG_5401.jpg
ELOG V3.1.3-