Message ID: 2674     Entry time: Sat Oct 2 23:57:50 2021
Author: Stephen 
Type: DailyProgress 
Category: Cryo vacuum chamber 
Subject: Chamber pumping down, carbon paint flakes cleaned up 

Pump down sequence executed tonight; Aidan plans to automate data collection during cooldown and warmup both, and the script will be activated early in the coming week.

- Carbon paint flakes (mentioned by Aidan in QIL/2667) were either picked up or scrubbed by IPA wipe, except the biggest, which were nudged near the closest 1/4-20 hole and picked up with tweezers for removal. There are still some small flakes of paint as there was less benefit to cleaning flakes closer to the PD or lens, and I opted not to risk any bumps. Aidan was correct, some areas of inner shield ID wall have flaked, but it seems the main location of delaminated paint is on the wrinkly foil excess covering the rim of the shield, an accidental paint location.

- While adding the lids, I monitored the PD outputs using Aidan's strip tool kindly left running. I never noticed any clipping in the trends - should I have been more skeptical?

- While adding the lids, I forgot to monitor the RTD outputs to the CTC100 controller, and the outer shield ended up shorting and ceasing to read any temperature. Didn't notice until I had turned on the roughing pump! Had to reopen and fix the short. Good reminder.

- Turbo came on at 12:07 am on 03 October at a pressure of 30 Torr - the setting is actually a timer and not a gauge reading.