Subject: +/-18V power supply to AI/AA chassis was actually +18V, -14.5V 

[Aidan]

I was working in the QIL on Friday and I heard a clicking sound coming from the rack where the DAQ is installed. It turned out to be the DC power supply for the AI/AA chassis. One of the voltage was floating around from ~14.2V to ~14.8V and the unit was clicking as it did this. Since the AA/AI chassis expect +/-18V which is regulated down to +/-15V, this was, to use the scientific term, bad.

I set the low voltage channel back to 18V. We have noticed previous drifts DAC channels - it's possible this was the cause.

 