Entry  Thu Aug 5 09:20:44 2021, Aidan, Laser, 2um Photodiodes, Optical design for 2um PD in new cryo chamber Cryo_Chamber_optical_layout.pdf
    Reply  Thu Aug 5 12:39:56 2021, Aidan, Laser, 2um Photodiodes, Optical design for 2um PD in new cryo chamber Cryo_in-air_setup.pdf
       Reply  Thu Aug 5 14:29:42 2021, Stephen, Laser, 2um Photodiodes, Optical design for 2um PD in new cryo chamber IMG_9685.JPGIMG_9687.JPGrouting_markup_of_IMG_9688.pngIMG_9682.JPG
          Reply  Tue Aug 24 15:01:25 2021, Stephen, Laser, 2um Photodiodes, 2um PD in new cryo chamber 
             Reply  Thu Aug 26 14:31:46 2021, Aidan, Stephen, Laser, 2um Photodiodes, Incident power calibration of 2um PD in new cryo chamber IMG_4420.jpgREF_PD_POW_CAL_cryo.pdf
                Reply  Fri Aug 27 17:57:30 2021, Stephen, DailyProgress, 2um Photodiodes, Start of PD pumpdown  
                   Reply  Sat Aug 28 07:01:21 2021, Aidan, DailyProgress, 2um Photodiodes, Start of PD pumpdown  20210828_-_IMG_4445.jpg20210828_-_IMG_4443.jpg
                      Reply  Sun Aug 29 11:17:43 2021, Aidan, DailyProgress, 2um Photodiodes, Start of PD pumpdown  Screen_Shot_2021-08-29_at_11.16.55_AM.pngScreen_Shot_2021-08-29_at_11.22.03_AM.png
                         Reply  Mon Aug 30 08:04:52 2021, Aidan, DailyProgress, 2um Photodiodes, Monday morning observation of JPL A1 PD cool down  Screen_Shot_2021-08-30_at_8.04.15_AM.png
Message ID: 2654     Entry time: Fri Aug 27 17:57:30 2021     In reply to: 2651     Reply to this: 2655
Author: Stephen 
Type: DailyProgress 
Category: 2um Photodiodes 
Subject: Start of PD pumpdown  

[Aidan, Stephen]

After Aidan validated that model inputs were creating physical parameter changes, we proceeded with some last few checks before closing up. Notes:

 - Aidan set up a helpful script turning laser power on and off, and strip tool to follow PD and monitor PD signals. A strip tool chart was used to confirm that there was no loss of functionality or alignment during pump down.

 - Checked Heater function and Workpiece RTD response - all good.

 - Confirmed alignment by steering at periscope output mirror and watching PD voltage.

At this point Aidan gave a green light for pumpdown prep, and to start pumpdown and start cooldown. Notes:

 - Stephen disconnected the PD connector accidentally while trying to add strain relief (the irony is palpable). Reattachment of connector didn't seem to affect any signal levels.

 - During installation of radiation shield lids, shields became misaligned and PD signal fell (presumably due to clipping). Recovered previous signal levels by realligning outer shield.

 - Double checked that everything seemed good to go, no issues!

 - Timeline of pumpdown and cooldown:

     17:32 - Pumpdown started.

     17:47 - Turbo spin up started (pump station delay parameter).

     17:55 - Pressure dropped below 1 mTorr, so I started cryocooler.

     18:05 - Healthy so far - Pressure had come down half a decade more, and Coldhead RTD was reading 235 K.