40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  ATF eLog  Not logged in ELOG logo
Entry  Thu Aug 5 11:53:26 2021, Radhika, DailyProgress, Cryo vacuum chamber, Cooldown model fitting for MS model_fit_v_data.pdf
    Reply  Wed Aug 11 14:58:47 2021, Radhika, DailyProgress, Cryo vacuum chamber, Cooldown model fitting for MS rad_model_fit_v_data.pdfrad_cond_model_fit_v_data.pdf
       Reply  Wed Aug 11 18:00:19 2021, Koji, DailyProgress, Cryo vacuum chamber, Cooldown model fitting for MS 
          Reply  Fri Aug 13 15:14:14 2021, Radhika, DailyProgress, Cryo vacuum chamber, Cooldown model fitting for MS rad_295_model_fit_v_data.pdf
             Reply  Fri Aug 13 21:01:42 2021, Radhika, DailyProgress, Cryo vacuum chamber, Cooldown model fitting for MS 
                Reply  Thu Aug 19 14:34:10 2021, Radhika, DailyProgress, Cryo vacuum chamber, Cooldown model fitting for MS model_fit_tm_painted.pdfmodels_painted.pdf
                   Reply  Fri Aug 20 13:44:58 2021, rana, DailyProgress, Cryo vacuum chamber, Cooldown model fitting for MS 
                      Reply  Fri Aug 20 14:05:45 2021, Radhika, DailyProgress, Cryo vacuum chamber, Cooldown model fitting for MS 
Message ID: 2648     Entry time: Fri Aug 20 13:44:58 2021     In reply to: 2647     Reply to this: 2649
Author: rana 
Type: DailyProgress 
Category: Cryo vacuum chamber 
Subject: Cooldown model fitting for MS 

I think the asymptotic temperature in the model is missing the data. i.e. the steady state temperature should match up, but the recorded data terminates too soon. Probably should figure out what the missing term is.

Can you post the covariance matrix of your fit so that we can see what the fractional errors are on the physical parameters? (i.e. construct the fit function so that physical parameters which are unknown are the fit parameters.)