40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  ATF eLog  Not logged in ELOG logo
Entry  Thu Aug 5 11:53:26 2021, Radhika, DailyProgress, Cryo vacuum chamber, Cooldown model fitting for MS model_fit_v_data.pdf
    Reply  Wed Aug 11 14:58:47 2021, Radhika, DailyProgress, Cryo vacuum chamber, Cooldown model fitting for MS rad_model_fit_v_data.pdfrad_cond_model_fit_v_data.pdf
       Reply  Wed Aug 11 18:00:19 2021, Koji, DailyProgress, Cryo vacuum chamber, Cooldown model fitting for MS 
          Reply  Fri Aug 13 15:14:14 2021, Radhika, DailyProgress, Cryo vacuum chamber, Cooldown model fitting for MS rad_295_model_fit_v_data.pdf
             Reply  Fri Aug 13 21:01:42 2021, Radhika, DailyProgress, Cryo vacuum chamber, Cooldown model fitting for MS 
                Reply  Thu Aug 19 14:34:10 2021, Radhika, DailyProgress, Cryo vacuum chamber, Cooldown model fitting for MS model_fit_tm_painted.pdfmodels_painted.pdf
                   Reply  Fri Aug 20 13:44:58 2021, rana, DailyProgress, Cryo vacuum chamber, Cooldown model fitting for MS 
                      Reply  Fri Aug 20 14:05:45 2021, Radhika, DailyProgress, Cryo vacuum chamber, Cooldown model fitting for MS 
Message ID: 2644     Entry time: Fri Aug 13 21:01:42 2021     In reply to: 2643     Reply to this: 2647
Author: Radhika 
Type: DailyProgress 
Category: Cryo vacuum chamber 
Subject: Cooldown model fitting for MS 

Wow, nice fit!

How is this c compared to the 300K exposure from the open aperture for the oplev beam?
We tried to cover all the ports but one. A naive assumption is that we have this opening with this aperture size, but the heat leakage could be more than that.
e.g. the aperture allows angled infrared to come into the chamber and the inner shield reflects the hot radiation inside (and a part of it reaches the test mass).