Entry  Mon Jul 26 01:30:42 2021, Koji, Summary, Cryo vacuum chamber, Prep for the 2nd cooling of the suspension 
    Reply  Mon Jul 26 22:49:00 2021, Koji, Summary, Cryo vacuum chamber, Aquadag painting 10x
       Reply  Wed Jul 28 00:59:47 2021, Koji, Summary, Cryo vacuum chamber, The test mass successfully suspended P_20210727_154143.jpegP_20210727_190356.jpegP_20210727_190426.jpegP_20210727_190543.jpeg
          Reply  Thu Jul 29 00:42:38 2021, Koji, Summary, Cryo vacuum chamber, The test mass successfully suspended 
             Reply  Thu Jul 29 13:11:17 2021, Koji, Summary, Cryo vacuum chamber, Cooling progress: Update temp_log_cool_down_20210728_1830.pdfcooling_model1.pdfcooling_model2.pdfOSEM_cooling.pdf
                Reply  Fri Jul 30 12:22:56 2021, Koji, Summary, Cryo vacuum chamber, Cooling curve comparisons comparison_cold_head.pdfcomparison_inner_shield.pdfcomparison_test_mass.pdf
                   Reply  Sun Aug 1 22:22:00 2021, Koji, Summary, Cryo vacuum chamber, Cooling update temp_log_cool_down_20210728_1830.pdfcooling_meas.pdfOSEM_cooling.pdfcooldown_210728.zip
                      Reply  Mon Aug 2 13:25:32 2021, Stephen, DailyProgress, Cryo vacuum chamber, Warmup started 02 August 
                         Reply  Tue Aug 3 23:56:00 2021, Koji, DailyProgress, Cryo vacuum chamber, Warmup started 02 August P_20210804_000247.jpgP_20210803_235421.jpg
Message ID: 2625     Entry time: Fri Jul 30 12:22:56 2021     In reply to: 2622     Reply to this: 2629
Author: Koji 
Type: Summary 
Category: Cryo vacuum chamber 
Subject: Cooling curve comparisons 

In all aspects, the latest cooling shows the best performance thanks to better thermal connection, thermal isolation, and the black paint.

- The cold head cooling is faster and cooler

- The inner shield cooling is faster

- The test mass cooling is faster

Attachment 1: comparison_cold_head.pdf  22 kB  Uploaded Fri Jul 30 13:23:05 2021  | Hide | Hide all
comparison_cold_head.pdf
Attachment 2: comparison_inner_shield.pdf  37 kB  Uploaded Fri Jul 30 13:23:05 2021  | Hide | Hide all
comparison_inner_shield.pdf
Attachment 3: comparison_test_mass.pdf  30 kB  Uploaded Fri Jul 30 13:23:05 2021  | Hide | Hide all
comparison_test_mass.pdf
ELOG V3.1.3-