Wed Dec 11 15:53:48 2019, Aidan, Chris, Raymond, Howto, Cryo vacuum chamber, Procedure to record photodiode output vs temperature
|
Wed Dec 11 19:50:25 2019, rana, Howto, Cryo vacuum chamber, Procedure to record photodiode output vs temperature
|
|
Message ID: 2477
Entry time: Wed Dec 11 19:50:25 2019
In reply to: 2476
|
Author: |
rana |
Type: |
Howto |
Category: |
Cryo vacuum chamber |
Subject: |
Procedure to record photodiode output vs temperature |
|
|
don't we also want to record the dark noise spectrum as a function of T and V_Bias ? I would guess that the dark noise doesn't always scale with dark current at low frequencies since its probably more like a random walk than shot noise. |