Entry  Tue Nov 26 18:49:11 2019, Koji, Update, PD QE, Socket soldering test piece made 6x
    Reply  Tue Dec 3 13:52:04 2019, Koji, Update, PD QE, Socket soldering test piece made PC029784.jpegPC029788.jpegPC029786.jpegPC029787.jpeg
Message ID: 2465     Entry time: Tue Dec 3 13:52:04 2019     In reply to: 2462
Author: Koji 
Type: Update 
Category: PD QE 
Subject: Socket soldering test piece made 

[Raymond and Koji]

We dunked the PD socket test piece into LN2 and repeated heat cycle 8 times. No obvious change was observed. Then the wires were pulled to find any broken joint or etc.
None of the solder joints showed the sign of failure.

For cleanliness, we are going to use In-Ag solder (no flux) for the actual wiring.

Attachment 1: Frozen connector

Attachment 2-4: Inspection after thawing.