OK...but how will the amplitude stabilization be done? How about a diagram showing the feedback loop and electronics?

  1. what is the feedback circuit?
  2. Can it drive the low impedance of an AOM driver?
  3. What is the photodiode?
  4. It has to be an AC coupled feedback (DC coupled is always terrible).
  5. What is the requirement on the intensity noise as a function of frequency?