Entry  Tue Dec 4 19:34:22 2018, awade, DailyProgress, WOPO, Diabolo 532 nm power drop and retuning 18-12-04_DiaboloInside_annotate.jpegplot20181204_DiaboloSHGRaligned1064nm.pdfplot20181204_DiaboloSHGRErrSig.pdf20181204_RetuneDiaboloSHGAlignment.zip
    Reply  Thu Dec 6 20:04:50 2018, awade, DailyProgress, WOPO, Diabolo 532 nm power drop and retuning 
       Reply  Fri Dec 7 20:31:06 2018, awade, DailyProgress, WOPO, Diabolo: cleaning dust out of SHG unit, self heating issues 18-12-07_20-06-46_4774.jpg18-12-07_20-06-58_4776.jpg18-12-07_20-07-44_4780.jpgIMG_4785.JPGIMG_4766.MOV
          Reply  Thu Dec 13 14:01:13 2018, awade, DailyProgress, WOPO, Diabolo: contacted coherent 
             Reply  Thu Dec 13 16:24:06 2018, awade, DailyProgress, WOPO, Diabolo: baseline numbers for coherent plot20181218_PhaseMatchingCurveFastScanDiaboloSHG.pdf20181213_SHGScanWithLNTempTuning.zip
                Reply  Wed Dec 19 16:48:18 2018, awade, DailyProgress, WOPO, Diabolo: baseline numbers for coherent 
                   Reply  Thu Dec 20 19:20:50 2018, awade, DailyProgress, WOPO, Diabolo: trialing SHG oven set temperature points to see if there is a maxima when locked plot20181220_SHGLockedOutputVsOvenSetpointTemperature.pdf20181220_SHGLockedPowerVsSetPointTemp.zip
                      Reply  Fri Dec 21 15:33:26 2018, Koji, DailyProgress, WOPO, Diabolo: trialing SHG oven set temperature points to see if there is a maxima when locked 
                         Reply  Fri Dec 21 15:51:27 2018, awade, DailyProgress, WOPO, Diabolo: trialing SHG oven set temperature points to see if there is a maxima when locked 
                      Reply  Mon Dec 24 10:08:49 2018, awade, DailyProgress, WOPO, Diabolo: back to power with deliberate pre-misalignment 18-12-21_22-04-30_4812.jpg18-12-21_22-04-54_4813.jpg18-12-21_22-07-55_4814.jpg
                         Reply  Mon Dec 24 12:07:18 2018, Koji, DailyProgress, WOPO, Diabolo: back to power with deliberate pre-misalignment 
Message ID: 2286     Entry time: Fri Dec 21 15:33:26 2018     In reply to: 2284     Reply to this: 2287
Author: Koji 
Type: DailyProgress 
Category: WOPO 
Subject: Diabolo: trialing SHG oven set temperature points to see if there is a maxima when locked 

This means that you want to make the SHG crystal longer. Is that true? If so, can you change the temperature for the optimal phase matching by tuning the 1064 crystal temperrature? I suspect you need to cool the YAG crystal, but I am not sure what is the thero-optic constant of the SHG crystal, and how much you can gain from this.