40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  ATF eLog  Not logged in ELOG logo
Entry  Fri Feb 19 20:15:03 2016, KojiN, Misc, PD QE, Measurement of the reflectivity, EQE, and IQE of the PD without glass window  Ref.pdfQE.pdfQE2.pdfRef_QE_wow.txtQE_cal3_NK.m
    Reply  Mon Feb 29 08:19:53 2016, KojiN, Misc, PD QE, Measurement of the reflectivity, EQE, and IQE of the PD without glass window  Reflectivity.pdfRef_wow_again.txt
       Reply  Wed Mar 2 14:38:31 2016, KojiN, Misc, PD QE, Measurement of the reflectivity, EQE, and IQE of the PD without glass window  QE3.pdfQE_ratio2.pdf
Message ID: 2024     Entry time: Fri Feb 19 20:15:03 2016     Reply to this: 2038
Author: KojiN 
Type: Misc 
Category: PD QE 
Subject: Measurement of the reflectivity, EQE, and IQE of the PD without glass window  

The glass window of the PD was removed and is placed in our shelf.

The reflectivity, EQE, and IQE of the PD were measured without the glass window as shown in following figures in the same setup as the measurements of the PD with glass window (see also elog2019, elog2020, and elog2022).

The laser power incident to the PD was 11.4 +/- 0.1 mW.

Fig. 1 Measured reflectivities of the PD without the galss window for P-pol and S-pol.
Fig. 2 Determined EQEs and IQEs of the PD without the galss window for P-pol and S-pol.
Fig. 3 Zoomed version of Fig. 2.

Ref_QE_wow.txt QE_cal3_NK.m

Attachment 1: Ref.pdf  5 kB  | Hide | Hide all
Ref.pdf
Attachment 2: QE.pdf  5 kB  | Hide | Hide all
QE.pdf
Attachment 3: QE2.pdf  5 kB  | Hide | Hide all
QE2.pdf
Attachment 4: Ref_QE_wow.txt  1010 Bytes  | Hide | Hide all
angle Ref(mW,P) error DC(V,P) error Ref(mW,S) error DC(V,S) error
-85 5.82 0.01 0.70 0.02 7.04 0.03 0.60 0.02
-80 4.05 0.02 1.34 0.02 4.72 0.01 1.22 0.02
-70 1.54 0.01 1.80 0.02 2.04 0.02 1.72 0.02
-60 0.532 0.002 1.98 0.02 .806 0.005 1.94 0.02
-50 0.197 0.002 2.06 0.02 0.408 0.002 2.02 0.02
-40 0.136 0.001 2.06 0.02 0.313 0.002 2.04 0.02
-30 0.217 0.001 2.06 0.02 0.363 0.001 2.06 0.02
-20 0.367 0.001 2.04 0.02 0.454 0.001 2.04 0.02
-10 0.480 0.001 2.02 0.02 0.504 0.001 2.02 0.02
... 11 more lines ...
Attachment 5: QE_cal3_NK.m  1 kB  | Hide | Hide all
% EQE calculation
%V=input('input voltage (V): ');
filename='Ref_QE_wow.txt';
delimiterIn = ' ';
headerlinesIn = 1;
A = importdata(filename,delimiterIn,headerlinesIn);

% filename1 = 'QE_Ref_P.txt';
% filename2 = 'QE_Ref_S.txt';
% filename3 = 'QE_enh.txt';
... 59 more lines ...