40m
QIL
Cryo_Lab
CTN
SUS_Lab
TCS_Lab
OMC_Lab
CRIME_Lab
FEA
ENG_Labs
OptContFac
Mariner
WBEEShop
|
ATF eLog |
Not logged in |
 |
|
Thu Jul 16 04:59:38 2009, Dmass, Laser, PSL, Razor Blades! 
|
Thu Jul 16 15:01:48 2009, Koji & Connor , Laser, PSL, Razor Blades!
|
Thu Jul 16 17:43:59 2009, Dmass, Laser, PSL, Razor Blades!
|
Mon Jul 20 23:19:49 2009, rana, Laser, PSL, Razor Blades!
|
Tue Jul 21 20:41:56 2009, Dmass, Laser, PSL, Razor Blades!
|
Fri Jul 17 15:39:31 2009, Dmass, Laser, PSL, Razor Blades!
|
|
Message ID: 177
Entry time: Thu Jul 16 04:59:38 2009
Reply to this: 178
182
|
Author: |
Dmass |
Type: |
Laser |
Category: |
PSL |
Subject: |
Razor Blades! |
|
|
Since the beamscan is in a questionable state of scanny goodness, Koji advised that I do some razor blade occlusion power measurements of the beam and then fit an erf to it to find the waist. I took data with the tinker toys pictured below.
I will compare these results with some beamscans results to verify (hopefully) that the beamscan is outputting useful results, not lies. |
|
|
|
|