Message ID: 1375     Entry time: Fri Apr 1 15:54:28 2011     In reply to: 1370     Reply to this: 1376
Author: rana 
Type: Electronics 
Category: GYRO 
Subject: Trans PD measured 

There is no case in which you can run the MAX4107 or LMH with less gain than the minimum gain of the datasheet. If you look at the time series of the PD with low gain its probably oscillating like crazy.

From the datasheet's Bode plots, you can see how the phase margin is for low gains: bad news.