Tue Sep 21 17:11:12 2010, Alastair, Electronics, GYRO, pdh box re-modifying 6x
|
Wed Sep 22 15:16:07 2010, Alastair, Electronics, GYRO, pdh box re-modifying
|
Wed Sep 22 17:03:24 2010, Alastair, Electronics, GYRO, pdh box re-modifying 
|
|
Message ID: 1078
Entry time: Wed Sep 22 17:03:24 2010
In reply to: 1077
|
Author: |
Alastair |
Type: |
Electronics |
Category: |
GYRO |
Subject: |
pdh box re-modifying |
|
|
I've replaced the 3.3uF capacitor as C28 in PDH box 1437. The measured open loop transfer function of the box now has a weird resonance up around 70kHz though which I really don't like the look of and it loses a huge amount of phase above this. Below that frequency it looks very much like the liso model. |
|
|
|
|