Here is an update to the prioritized materials testing queue for UHV compatibility, which was last reported in elog:3
- On-going test - can be terminated now: E1900220-v1: Optical Contamination Test results: SR3 ROC Actuator, Ceramic Heater Assy
- On-going test - some value in continuing test: E1900232-v1: Optical Contamination Test results: Low-Reflection, Matte Black Aluminum Foil
- Accu-Glass PEEK UHV Heat-Shrink Tubing (useful for in-chamber cabling)
- Krytox LVP (not officially approved for LIGO, but permitted extremely small amounts on beam diverters & within New Focus picomotor 830X-UHV)
- Torr Seal epoxy (proposed for ZnSe window seal for in-vacuum, non-pumped, wavefront sensor by Australians, possibly for A+?)
- Castrol Braycote 601EF (used in a proposed vacuum rated heat switch motor assembly for Voyager)
Note that the following test has been terminated/completed "recently":
|