ID Date Author Typeup Category Subject Text Attachments
  442   Wed Aug 24 02:57:43 2022 KojiGeneralGeneralEP30-2 bonding setupEP30-2 bonding setup  IMG_1178.JPGIMG_1179.JPGIMG_1180.JPG 
  443   Wed Aug 24 03:20:59 2022 KojiGeneralGeneralOMC #002 Delamination repair Part2 (2)Bonding  11x 
  444   Wed Aug 24 03:26:43 2022 KojiGeneralGeneralOMC #002 Delamination repair Part2 (3)Inspection

 
 8x 
  445   Wed Aug 24 11:29:47 2022 KojiGeneralGeneralOMC #002 ready for shipment[Stephen Koji]

The OMC #002 is ready for
shipment.
 IMG_1207.JPGPXL_20220825_004259850.jpgPXL_20220825_004307204.jpgPXL_20220825_005423342.jpgPXL_20220825_005549985.jpg 
  446   Thu Aug 25 14:22:08 2022 KojiGeneralGeneralLLO OMC #001 Ballast Mass investigationInspected the past LLO add-on mass configuration.

There are unknown masses at the
DCPD side. It looks like a small SS mass
 DSCN0917.JPGDSCN0922.JPGP6108705.JPGP6108707.JPGP6108706.JPG 
  447   Thu Aug 25 20:05:00 2022 KojiGeneralGeneralLLO OMC #001 Ballast Mass investigationHere is the balance mass info for the LLO
OMC#001 analyzed from the photographs


Added masses are: 50+10g,
 balance_mass_config_LLO.pdf 
  448   Fri Aug 26 22:29:02 2022 KojiGeneralGeneralTool Shipping PrepShipping preparation for the LLO
trip

Started
July 15, 2022 and finished Aug 30. So it
 PXL_20220831_025623318.jpgPXL_20220831_024518032.jpgPXL_20220831_030234581.jpg 
  449   Tue Sep 20 08:54:33 2022 KojiGeneralGeneralPD cage arrangementUpon the LLO work, the current PD arrangement
in the cages are:
CAGE B
  
  450   Mon Sep 26 14:27:49 2022 KojiGeneralGeneralLLO OMC ICS workOMC
#001


ICS Link: https://ics.ligo-la.caltech.edu/JIRA/browse/ASSY-D1201439-1
Already
  
  461   Fri Nov 18 18:41:05 2022 Camille MakaremGeneralGeneral2nd deep cleaning of OMC #1The four cavity mirrors in OMC #1 were
each scrubbed using acetone and a cotton
swab.
 PXL_20221118_213955948.jpg 
  462   Mon Nov 21 19:13:35 2022 KojiGeneralGeneralTransmission measurement (2nd deep cleaning of OMC #1)OMC Transmission measurement after
the 2nd deep cleaning

The 2nd deep cleaning didn't
 PXL_20221122_030736513.jpgOMC_loss.pdf 
  463   Tue Nov 29 15:54:47 2022 KojiGeneralConfigurationWindows laptop for WincamD Beam'R2 recoveryAaron took the set to Cryo lab

 
  
  473   Wed Jan 25 23:51:04 2023 KojiGeneralGeneralThe items packed for DownsQty1 1/2 mounts
Qty2 prism mounts
Qty6 gluing fixures
 PXL_20230127_055920944.jpg 
  482   Wed Feb 1 01:44:14 2023 KojiGeneralGeneralOMC (004) plan2/1 2:30PM~ Bonding reinforcement (Last
EP30-2 gluing)

2/2 1:00PM~ Peripheral attachment
  
  487   Sat Feb 4 03:33:22 2023 KojiGeneralGeneralOMC Lab InventoryAttachment 1: North Cabinet 2nd from the
left

Attachment 2: North Cabinet 3rd
 PXL_20230204_072749311.jpgPXL_20230204_072744686.MP.jpgPXL_20230204_073603041.MP.jpg 
  525   Tue Apr 18 07:59:46 2023 Madeline LesovskyGeneralGeneralTest Cure Bake - Curved Mirror Subassembly BondingThis is a test bake conducted in Air Bake
Oven A (ABO-A) held in the 40m Clean and
Bake facility. The overall objective is to
 6x 
  529   Wed Apr 26 14:43:44 2023 StephenGeneralLoan / LendingBorrowed Items for PZT DC Response Shadow Sensor SetupBorrowed for PZT DC Response Shadow Sensor
Setup (see Attachment 1):


Thorlabs PDA100A Photodiode
 PXL_20230426_185227710.jpg 
  535   Thu May 4 15:12:14 2023 KojiGeneralCharacterizationOMC (004): Final Testing
FSR measurement (dip) -
done May 3, 2023
FSR
measurement (RFAM injection) - done
  
  537   Thu May 11 10:46:17 2023 Thejas GeneralCharacterizationaLIGO OMC[Thejas, Camille, Koji]

We aligned the laser beam to the
cavity and drove the OMC cavity PZTs (0 to
  
  538   Fri May 12 10:00:06 2023 Camille MakaremGeneralLoan / LendingBorrowed Items for PZT DC Response Shadow Sensor SetupThe ThorLabs MDT694B piezo driver was returned
to the OMC lab.

 PXL_20230511_201545235.jpg 
  541   Tue May 16 10:26:12 2023 KojiGeneralCharacterizationOMC (004): Final Testing
FSR measurement (dip) -
done May 3, 2023
FSR
measurement (RFAM injection) - done
  
  543   Fri May 19 16:09:31 2023 StephenGeneralGeneralA+ OMC Build Progress, Week of 2023.05.15A+ OMC Build efforts ongoing or completed
this week:


 PZT lead onboard
  
  546   Fri Jun 2 17:15:44 2023 StephenGeneralGeneralA+ OMC Build Progress, Weeks of 2023.05.22 and 2023.05.29A+ OMC Build efforts ongoing or completed
in the last two weeks:


 PZT lead onboard
  
  549   Sun Jun 25 21:04:35 2023 KojiGeneralCharacterizationOptical assembly/testing of OMC (004) completedOptical assembly/testing of OMC
(004) completed
- OMC (004) with the transport fixture
 PXL_20230624_072237980.jpgPXL_20230624_072254264.jpgPXL_20230624_072304754.jpgPXL_20230624_072320422.jpg 
  550   Sun Jun 25 21:08:37 2023 KojiGeneralCharacterizationOMC (004) transported to Downs for cable furnishingOMC (004) + the transport fixture was transported
to Downs 227 with the boxes for OMC cabling
hardware and DCPD/QPD connector parts.
 PXL_20230624_073552303.jpgPXL_20230624_073531044.jpg 
  553   Thu Jul 6 10:19:23 2023 Thejas GeneralCharacterizationOMC Lab optic table [Camille, Thejas]

The optics table in OMC lab was
cleaned up, cameras and photodiodes used
  
  557   Mon Jul 10 20:29:05 2023 Koji GeneralCharacterizationOMC(004) PZT characterizationBelated entry: OMC(004) PZT Slow
scan / freq response measurements (May 10/11/17,
2023)
 PZT_Scan.pdfPZT_response.pdfOMC_OL_CLTF.pdf 
  559   Wed Jul 12 17:27:44 2023 KojiGeneralGeneralMitsutoyo Digital Caliper (100mm)We had missing digital caliper from the
clean room. As it is an essential component,
I have ordered a replacement.
  
  582   Mon Jul 31 13:19:14 2023 Camille MakaremGeneralGeneralplans for this week 31 July 2023Plans for this week:
1) Verify electronics setup for cavity
locking.
 4ae4ed99-007e-4246-85b6-fadccc5ba19d.mp4 
  583   Tue Aug 1 11:04:13 2023 Camille MakaremGeneralGeneralreoptimizing cavity alignment before locking[Camille, Thejas]
We rechecked the alignment this morning
and re-optimized the resonance a bit. There
  
  585   Thu Aug 3 07:22:19 2023 Camille MakaremGeneralGeneralparticle count and cavity optimization/alignmentYesterday, we measured the particle count
in the enclosure to ensure that the lower
setting on one of the HEPAs is still acceptable.
  
  588   Thu Aug 3 15:04:26 2023 StephenGeneralGeneralaLIGO Unit 4 Build, Cable Routing, Preparation[Dean, Stephen]

Unit #4 aLIGO OMC unit
was transported to the 40m Bake Lab.
 cable_routing_20230803_IMG_3527.jpg 
  591   Mon Aug 7 07:36:09 2023 Camille MakaremGeneralGeneralparticle count[Camille, Thejas]

We lowered the setting to MEDIUM on
  
  595   Mon Aug 7 16:00:30 2023 KojiGeneralGeneralOMC (004): Cable installation The writing of this elog is still on going

[Dean, Stephen, Koji]

4th
 IMG_3546.jpgIMG_3547.jpgIMG_3564.jpgIMG_3571.jpg 
  596   Mon Aug 7 16:01:05 2023 KojiGeneralGeneralOMC (004): Production completedOMC (004) Production completed. The OMC
together with the transport fixture is still
placed at the 40m clean room.
 PXL_20230807_210716110.jpgPXL_20230807_210724685.jpg 
  615   Tue Aug 22 15:51:15 2023 KojiGeneralGeneralInventory: C30655 PDsDuring the OMC(004) assembly, the stock
situation for Excelitas C30655 PDs was checked.

PD Cage H
  
  626   Mon Nov 27 21:05:51 2023 KojiGeneralGeneralInventory: C30655 PDsThe Cage H&I, as well as OMC DCPD bag,
were brought to the 40m for the 40m BHD.

 
  
  52   Sun Jan 6 23:22:21 2013 KojiMechanicsGeneralSolidWorks model of the OMC suspension    D0900295_AdvLIGO_SUS_Output_Mode_Cleaner_Overall_Assembly.pngD0900295_AdvLIGO_SUS_Output_Mode_Cleaner_Overall_Assembly.easm 
  58   Tue Jan 22 17:56:32 2013 KojiMechanicsGeneralRotary stage selectionNewport UTR80

    Sensitivity: 4 arcsec
  
  69   Thu Mar 7 15:53:47 2013 KojiMechanicsGeneralOMC Transportation fixture, OMC PD/QPD mounts   P3073218.JPGP3073219.JPGP3073223.JPGP3073225.JPGP3073227.JPG 
  70   Thu Mar 14 17:06:21 2013 KojiMechanicsGeneralOMC SUS work @LLOEDIT (ZK): All photos on
Picasa. Also, I discovered that
 2013-03-14_16.03.40.jpg2013-03-14_16.04.07.jpg 
  90   Mon Apr 1 10:28:03 2013 KojiMechanicsGeneralAdditional UV blast for the top surface[Koji, Lisa, Jeff, Zach]

Jeffs concern after talking with the
 P3283483.JPGP3283459.jpgP3283473.JPG 
  92   Wed Apr 3 17:39:38 2013 KojiMechanicsCharacterizationCalibration of the test PZTs before the glue testWe want to make sure the responses of the
PZT actuator does not change after the EP30-2
gluing.
 P4033491.JPGshadow_sensor_calib.pdf 
  98   Fri Apr 5 14:39:26 2013 KojiMechanicsCharacterizationCalibration of the test PZTs after the heat cureWe attached fused silica windows on the
test PZTs. http://nodus.ligo.caltech.edu:8080/OMC_Lab/93
The glued assemblies were brought
 shadow_sensor_calib_after_bake.pdfPZTresponse.pdf 
  102   Mon Apr 8 11:49:18 2013 KojiMechanicsCharacterizationPZT actuator tested at LLOTest result of the PZTs by Valera and Ryan
PZT  Length Angle
 #
  [nm/V] [urad/um]
 11  14.5   17.6
  
  124   Mon May 13 14:49:35 2013 KojiMechanicsCharacterizationMounting Glass Bracket still broke with tightenin stress[Koji / Jeff]
This is the elog about the work on
May 9th.
  
  125   Mon May 13 14:59:16 2013 KojiMechanicsGeneralInvar shim gluingThe invar reinforcement shims were glued
on the glass brackets on the breadboard.
  
  130   Thu May 23 23:41:48 2013 KojiMechanicsGeneralDCPD/QPD MountDCPD mounts and QPD mounts were attached
on the breadboard. They are not aligned yet
and loosely fastened.
  
  148   Sat Jul 6 17:10:07 2013 KojiMechanicsCharacterizationPZT Response analysisAnalysis of the PZT scan / TF data taken
on May 31st and Jun 1st.
[DC scan]
 PZT_Scan.pdfL1OMC_PZT_Response.pdf 
  158   Tue Aug 27 17:02:31 2013 KojiMechanicsCharacterizationSpot position measurement on the diode mountsAfter the PZT test, the curved mirrors
were aligned to the cavity again.
In order to check the height of the
 DCPD1.png 