40m QIL Cryo_Lab CTN SUS_Lab CAML OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  OMC elog, Page 5 of 11  Not logged in ELOG logo
ID Date Author Typedown Category Subject
  472   Mon Jan 23 19:51:20 2023 KojiOpticsGeneralOMC #4: cavity mirror bonding

[Camille, Koji]

We continued to bond two CM mirrors and the other two steering mirrors for QPD2.

Before the bonding work, the FSR and TMSs were checked again.

FSR: 264.7925 MHz
TMS_V: 58.15125 MHz
TMS_H: 58.33375 MHz

Checked the transmission: The OMC loss was 4.3 +/- 0.2 %.

This does not make the HOMs coincidently resonant until the 18th-order (+9MHz). Looks good.


CM1/CM2/SM2/SM3 bonding

- Applied the bond to CM1 and the UV illuminated.
- Applied the bond to CM2 and the UV illuminated.
==> The cavity bonding is completed.

Removed the micrometer for CM2 to allow us to bond SM2/SM3
- Checked the spot at QPD2: The spot was a couple of mm too left. This was too much off compared to the QPD adjustment range. ==> Decided to shim the SM3 position with a piece of Al foil.
- Otherwise everything looked good. SM2/SM3 were bonded.


Invar block bonding

Prepared EP30-2
- There are three tubes of EP30-2 that expires on 2/22, 2023.
- A tube was almost empty. Used this tube to fill/purge the applicator. The 2nd tube was then attached to squeeze out 8g of glue mixture. 
- 0.4g of fused silica beads were added to the glue mixture.
- Mixed the bond and a test piece was baked by the oven. (200F=95C, 5min preheat, bakeing 15min).
- The glue test piece was "dry" and crisp. Looked good.
- Applied the glue on the invar blocks. Confirmed that the bonding surfaces were made completely "wet".
- 4-40 screws were inserted to the blocks so that the blocks were pushed toward the template. See Attachments 3 and 4.

 

Optic Inventory

Breadboard: #6
BS1: E6
FM1: A1
FM2: A3
CM1: PZT ASSY #8 (M7+PZT11+C11)
CM2: PZT ASSY #11 (M14+PZT13+C13)
SM1: E9
BS2: B8
SM2: E11
SM3: E14
BS3: B6

Attachment 1: HOM_plot_PZT0_0.pdf
HOM_plot_PZT0_0.pdf
Attachment 2: PXL_20230124_032358482.MP.jpg
PXL_20230124_032358482.MP.jpg
Attachment 3: PXL_20230124_032452805.MP.jpg
PXL_20230124_032452805.MP.jpg
Attachment 4: PXL_20230124_032425638.jpg
PXL_20230124_032425638.jpg
  474   Thu Jan 26 22:57:19 2023 KojiOpticsGeneralOMC #4: One Invar block bonded with tilt

[Camille, Koji]

The bottom side template was separated into two pieces and successfully removed from the breadboard. The template was assembled together again and bagged to store it in a cabinet.

We found that the invar block for DCPD(R) was bonded with some air gap (Attachment2 1/2).

 

The Allen key used as a weight was too small, which caused it to get under one of the screws used as hooks and lift the block.

 

We've investigated the impact of this tilt.

- Bonding strength: The bonding area is ~60% of the nominal. So this is weak, but we can reinforce the bonding with an aluminum bar.
- Misalignment of the DCPD housing: The tilt will laterally move the position of the DCPD. However, the displacement is small and it can be absorbed by the adjustment range of the DCPD housing.
- Removal: From the experience with the removal of the beam dump glass, this requires a long time of acetone soaking.

Conclusion:

- We don't need to remove the invar block.
- Action Item: Reinforcement of the bonding

Attachment 1: PXL_20230126_221425695.jpg
PXL_20230126_221425695.jpg
Attachment 2: PXL_20230126_221522416.jpg
PXL_20230126_221522416.jpg
  475   Thu Jan 26 23:07:14 2023 KojiOpticsGeneralOMC #4: Input beam dump bonding

[Camille, Koji]

During the second UV epoxy session, we did not bond the input beam dump. This is because this beam dump was not the one planned from the beginning and if it was bonded in place, it would have created difficulties when removing the template.

First, we aligned a couple of Allen wrenches to define the location of the beam dump. We've checked that the main transmission is not blocked at all while the stray beam from the OMC reflection is properly dumped.

After the confirmation, the UV epoxy + UV alight were applied.

The resulting position of the beam dump is shown in the attachment.

Attachment 1: PXL_20230126_221556977.jpg
PXL_20230126_221556977.jpg
  476   Thu Jan 26 23:16:35 2023 KojiOpticsConfigurationOMC #4: Top side bonding.

[Camille, Koji]

1. Flipping the OMC

It turned out that the transport fixture for this OMC could not be closed. The locks are too short, and the knobs could not be turned. We temporarily fastened the long 1/4-20 screws to secure the box and flipped it to make the top side face up.

2. Setting up the top-side template

The top side template was attached to the breadboard. We took care that the lock nuts on the positioning screws were not touched. The margins between the template and the glass edges were checked with a caliper. The long sides seemed very much parallel and symmetric, while the short sides were not symmetric. The lock nut on the short side was loosened, and the template was shifted to be symmetric w.r.t. the breadboard.

3. UV epoxy work

The cylindrical glass pieces were wiped, and the bonding surfaces were cleaned so that the visible fringes were <5 fringes. We confirmed the hooking side is properly facing up. The UV epoxy and UV curing were applied without any trouble. (Attachment 1)

4. EP30-2 bonding of the invar mounting blocks

Six invar blocks were bonded. This time the Allen key weights were properly arranged, so they didn't raise the blocks. The bond properly wetted the mating surfaces.

---

The final step of the bonding is to remove the template.
And replace the locks of the transport fixture.

 

Attachment 1: PXL_20230126_235309334.jpg
PXL_20230126_235309334.jpg
Attachment 2: PXL_20230127_053008237.jpg
PXL_20230127_053008237.jpg
Attachment 3: PXL_20230127_053020709_2.jpg
PXL_20230127_053020709_2.jpg
Attachment 4: PXL_20230127_053042809.MP_2.jpg
PXL_20230127_053042809.MP_2.jpg
Attachment 5: PXL_20230127_053058275_2.jpg
PXL_20230127_053058275_2.jpg
  477   Thu Jan 26 23:46:13 2023 KojiOpticsGeneralOMC #1 input beam dump bonding

A beam dump was stacked on the base of the previous beam dump. The angle was determined so that the main transmission goes through while the stray OMC reflection is blocked without clipping at the edge.

The resulting alignment of the beam dump is shown in Attachment 1.

The beam dump tended to slip on the base. To prevent that a couple of weights were placed around the bonding area. (Attachment 2)

 

Attachment 1: PXL_20230127_052955482.jpg
PXL_20230127_052955482.jpg
Attachment 2: PXL_20230127_052949869.jpg
PXL_20230127_052949869.jpg
  483   Thu Feb 2 03:09:41 2023 KojiOpticsGeneralOMC #4: One Invar block bonded with tilt

[Camille, Thejas, Koji]

We added a reinforcement bar at the back of the invar block which had the tilt issue.

The reinforcement bar was added to the backside rather than the side or front such that the DCPD housing does not interfere with the reinforcement bar.

Also, small amount of EP30-2 was added to the CM2 wire so that the repeated bend of the PZT wire cause the disconnection at the PZT.

Attachment 1: PXL_20230202_003156761.jpg
PXL_20230202_003156761.jpg
Attachment 2: PXL_20230202_003204058.jpg
PXL_20230202_003204058.jpg
  484   Sat Feb 4 03:16:10 2023 KojiOpticsGeneralOMC #4: One Invar block bonded with tilt

The attached photo shows the resulting bond spread.

 

Attachment 1: PXL_20230202_213114009.jpg
PXL_20230202_213114009.jpg
  489   Wed Feb 8 16:10:52 2023 StephenOpticsCharacterizationA+ OMC, Parallelism of HR Prisms

[Camille, Thejas, Stephen]

We set up the white light autocollimator in the Downs B119 cleanroom. (Nippon Kogaku, from Mike Smith).

After some initial effort to refine the fixturing and alignment, we located the S1 crosshair reflection and aligned to the autocollimator reticle using the pitch and yaw adjustments in the prism mount.

We subsequently used the rotation stage adjustment to locate the S2 crosshair reflection and measure the vertical and horizontal wedges.

Faint horizontal crosshair from the S2 reflection can be seen in the image below.

This is aligned with the reticle using rotation mount on which the prism mount is clamped.

Initial readiing of the rotation mount screw: 9.2 

Final reading: 2.2

Here we see that the crosshair from S2 reflected light is offset in the vertical axis by approx. 2 div. From hte image below this should

correspond to 2 arcmin vertical wedge angle.The horizontal wedge angle is yet to be caluclated.

Attachment 3: IMG_3C6388ECD50E-1.jpeg
IMG_3C6388ECD50E-1.jpeg
  490   Thu Feb 9 15:54:41 2023 Thejas OpticsCharacterizationA+ OMC, Parallelism of HR Prisms

[Camille, Thejas, Stephen]

Continuing yesterday's efforts to measure the wedge angle of the back surface of the prisms. We completed measurement for all the 18 prisms.

The images below accompanying the readings represent the S2 crosshair image on top of the reticle, alighned for yaw.

But note that the vertical misalignement with the reticle does not give an accurate measurement for vertical wedge angle. This is because, as it's notecable in the images, 

the S1 reflected crosshair's horizontal axis goes out of coincidence from the horizontal axis of the reticle as the stage is rotated. Our thoughts: MAy be the horizontal 

plane of the mount is not the same as the horizontal plane of the autocollimator.

 

Each unit of the readings corresponds to 0.1 deg., the resolution of the rotational stage is 0.2 deg. The requirement is 0.5 deg of wedge angle. And this angle is related to the horizontal wedge angle by: 

Prism 02

Initial reading of the screw on the rotation (yaw) stage (ini): 7.6 

Final reading of the screw (fin): 0.2

 

Prism 04

ini: + 5.1

fin: - 8.0

Prism 05

ini: + 1.8

fin: - 5.5

Prism 06

ini: + 5.8

fin: - 8.5

Prism 07

ini: 8.2

fin: 1.0 

Prism 09

ini: +1.0

fin: - 4.2

Prism 10

ini: +9.1

final: +2.2

 

Prism 11

ini: 9.1

fin: 2.0 

Prism 12

ini: 9.0

fin: 2.2

 

Prism 13

ini: 9.0 

fin: 2.2

 

Prism 14

ini: 9.0 

fin: 2.1

Prism 15

ini: 9.0

fin: 2.0 

Prism 16

 

ini: 9.0 

fin: 2.2

Prism 17

ini: 9.0

fin: 2.0

Prism 22

ini: 9.0 

fin: 2.1

Prism 24

ini: 9.1

fin: 2.1

Prism 26

ini: 9.0 

fin: 2.3

This totals 18 prisms including yesterdays. 

 

 

 

 

 

 

 

 

Attachment 18: Raw_data.pdf
Raw_data.pdf Raw_data.pdf Raw_data.pdf Raw_data.pdf Raw_data.pdf Raw_data.pdf Raw_data.pdf Raw_data.pdf
  491   Tue Feb 14 10:45:00 2023 Thejas OpticsCharacterizationA+ OMC Prism perpendicularity

[Camille, Stephen, Thejas]

Following the wedge angle measurements of the prisms, perpendicularoty of their bottom surface with respect to their HR surface was measured usign the autocollimator. More info. about the procedure can be found in the OMC testing document. We want to set the requiremetns for perpendicularity to better than 30 arcsec (or 0.-0083 deg).

Images of the setup 

 

 

Prism 1: 

View through teh autocollimator (AC) while hte prism is unclamped:

Two horizontal crosshair lines can be seen, with a common vertical crosshair. These corresspond to the two separate reflections of the AC beam fom the retroflector (RR) surfaces formed by the prism and the flat Al mirror (see image below). When the RR formed is 90 deg the two horizontal lines overlap. The separation between the lines, when calibrated, represents 4 x the deviation of the prism from perpendicularity. Note that, since this prism is unclamped the crosshairs don't indicate a true reading. Note that since the autocollimator images are in far field, the splitting of the horizontal lines shouldn't depend on the pitch angle of the coupling mirror, this can also be checked by the adjusting the pitch screws. 

 

Clamped: 

Multiple images below to check reproducibility:

 

1 div. of the reticle in the above images corresponds to 1 arc min. By measuring the separation of the horizontal shifting gives angle of deviation from perpendicularity. 

From the above images it can be inferred that the surfaces form a 90 deg RR. 

 

 

Prism 2

As it can be seen in the top images there's a splitting of hte horizontal lines indicating deviation from perpendicularity. The direction of the deviation can be inferred by softly tocuhing/pressing on the front orn the back en of the flat Al mirror surface as shown in the images below. 

 

Prism 4

Prism 5

 

Prism 6

Prism 7

 

Prism 9

 

Prism 10

Prism 11

 

Prism 12

 

Prism 13

Prism 14

 

Attachment 10: IMG_379CF9F79CCB-1.jpeg
IMG_379CF9F79CCB-1.jpeg
Attachment 12: IMG_146D1BDD8AC5-1.jpeg
IMG_146D1BDD8AC5-1.jpeg
Attachment 17: IMG_5783285B694E-1.jpeg
IMG_5783285B694E-1.jpeg
Attachment 23: IMG_FC0EC9B1CA92-1.jpeg
IMG_FC0EC9B1CA92-1.jpeg
Attachment 25: OMC_2_(dragged).pdf
OMC_2_(dragged).pdf OMC_2_(dragged).pdf OMC_2_(dragged).pdf OMC_2_(dragged).pdf OMC_2_(dragged).pdf OMC_2_(dragged).pdf OMC_2_(dragged).pdf OMC_2_(dragged).pdf
Attachment 26: OMC_2_(dragged)_(dragged).pdf
OMC_2_(dragged)_(dragged).pdf
  492   Tue Feb 14 22:52:35 2023 Thejas OpticsCharacterizationA+ OMC Prism perpendicularity of HR Prisms and BS

[Camille, Stephen, Thejas]

Continuing with the efforts to measure the perpndicularity.

Prism 15

Prism 16

Prism 17

 

Prism 22

 

Prism 24

Prism 26

 

Perpendicularity measurement for Beam Splitters

BS 25

 

BS 29

 

 

BS 28

 

 

BS 36

 

BS 33

 

BS 34

 

 

BS 35

 

BS 37

 

 

BS 38

 

BS 39

 

 

 

Attachment 17: OMC.pdf
OMC.pdf OMC.pdf OMC.pdf OMC.pdf OMC.pdf OMC.pdf OMC.pdf OMC.pdf
Attachment 18: OMC_annex.pdf
OMC_annex.pdf
  494   Wed Feb 15 17:40:21 2023 Thejas OpticsCharacterizationA+ OMC perpendicularity of hole prisms

[Camille, Stephen, Thejas]

Contnuing the efforts to measure and check perpendicularity: tombstone prisms with holes/ hole prisms (HP).

Note: Veritcal crosshair splitting can be seen in the some of the image. This is probably because the horizontal of the Al flat mirror is not parallel to that of the coupling mirror. This was confirmed by touching the so that the setup roll a bit so as to reduce the vertical splitting. In some cases the position of the prism on the flat mirror was changes to reduce this effect, in some other cases this was not very helpful and measurement was done anyway. We expect that teh vertical splitting and horizontal splitting don't couple into each other. We think the clamping mechanism for this kind of measurement can be improved to avoid these artefacts. 

HP40

HP41

HP42

 

HP43

 

HP44

HP45

HP46

HP47

HP48

HP49

 

HP50

HP51

 

HP 52

HP 53

HP 54

HP 55

 

HP 56

 

HP 57

 

 

 

 

Attachment 19: OMC_2_(dragged)_2.pdf
OMC_2_(dragged)_2.pdf OMC_2_(dragged)_2.pdf OMC_2_(dragged)_2.pdf OMC_2_(dragged)_2.pdf OMC_2_(dragged)_2.pdf OMC_2_(dragged)_2.pdf OMC_2_(dragged)_2.pdf OMC_2_(dragged)_2.pdf
Attachment 20: OMC_2_(dragged)_3.pdf
OMC_2_(dragged)_3.pdf OMC_2_(dragged)_3.pdf OMC_2_(dragged)_3.pdf OMC_2_(dragged)_3.pdf
  495   Fri Feb 17 17:11:28 2023 Thejas OpticsCharacterizationA+ OMC beam-splitter prisms wedge angle measurement

[Camille, Stephen, Thejas]

 

Yesterday we measured wedge angle of the beamsplitter (BS) prisms. I reckon these measurements are not as important as the BSs will be used outside the cavity and the angle of incidence is significant. 

Measurement procedure and setup used are the same as that for the prism mirrors wedge angle measurements.

BS25

 

initial division reading: 9.0 

finbal division reading: 2.5 

BS28

 

ini: 9.0 

fin: 2.0 

 

BS29

ini: 9.0 

fin: 1.9 

BS33

ini: 9.0 

fin: 2.0 

 

BS34

ini: 9.0 

fin: 1.7

 

BS35

ini: 9.0 

fin: 2.0

BS36

ini: 9.0

fin: 2.3

 

 

BS37

ini: 9.0 

fin: 2.3

BS38

ini: 9.0

fin: 2.2

 

BS39

 

ini: 9.0

fin: 2.4

 

 

 

 

 

 

 

 

Attachment 11: OMC_5_(dragged)_2.pdf
OMC_5_(dragged)_2.pdf OMC_5_(dragged)_2.pdf OMC_5_(dragged)_2.pdf OMC_5_(dragged)_2.pdf OMC_5_(dragged)_2.pdf OMC_5_(dragged)_2.pdf OMC_5_(dragged)_2.pdf
  496   Fri Feb 17 17:25:39 2023 Thejas OpticsCharacterizationA+ OMC Breadboard measuerements

Yesterday we also measured weight and dimensions of breadboard. Error for the following measurements is same as the least count of the instruments used. 

26

6149 g 

450.56 mm x 41.45 mm x 150.39 mm 

23

6127 g

450.37 mm x 41.25 mm x 150.17mm

 

25

6155 g

450.83 mm x 41.44 mm x 150.15 mm

 

24

6158 g

450.30 mm x 150.42 mm x 41.42 mm

 

20

6147 g

450.06 mm x 150.18 mm x 41.42 mm

 

22:

6149 g

450.01 mm x 150.57 mm x 41.43 mm

21:

 

6143 g 

450.01 mm x 150.06 mm x 41.44 mm

 

 

 

Attachment 1: IMG_3753BB8D72D5-1.jpeg
IMG_3753BB8D72D5-1.jpeg
Attachment 2: IMG_62A5AD50E8D1-1.jpeg
IMG_62A5AD50E8D1-1.jpeg
Attachment 3: OMC_5_(dragged).pdf
OMC_5_(dragged).pdf OMC_5_(dragged).pdf OMC_5_(dragged).pdf OMC_5_(dragged).pdf
  498   Mon Feb 27 17:40:27 2023 Thejas OpticsCharacterizationA+ OMC curved mirror characterization

[Camille, Stephen, Thejas]

Today, before the ZYGO lab was cleaned and prepared for the cureved mirrors' radius of curvature (ROC) characterization, Mirror no. 6 was mounted into one of the half inch mirror holders. The cleanliness of the envoronment and handling was not satisfactory. Tomorrow efforts will be made to start doing the ROC measurements with class B cleanroom garbing.

  499   Wed Mar 1 10:23:10 2023 Thejas OpticsCharacterizationA+ OMC curved mirror characterization

[Camille, Thejas, Stephen]

Yesterday, efforts were made to measure ROC of curved mirrors (#6) in the ZYGO lab using a Fizeau Interferometer. Peculiar observation: Stray fringes were seen that dominated the fringes that conformed with the expectation. The origin of these fringes is still not accounted for (see attached screenshot). moreover, once the right fringe pattern is achieved by moving the end mirror of the interferometer using a translation stage, the cavity length is measured using a metre stick. This makes the measurement limited by the accuracy using ruler stick for cavity length measurement, which is not expected to be any better than usign a beam profiler to find the focal point from the curved mirror. Today we will, move ahead to corved mirror surface profile characterization.

 

 

  501   Mon Mar 6 16:02:56 2023 Thejas OpticsGeneralOMC test setup locking

On Feb 16, Camille and I attampted at locking the OMC cavity. It was quick to re-align the beam to the cavity (by using only the fine adjustment of the output fibre couple). This was done by looking to minimize the power reflected from the cavity and observing the mode shapes on the CCD. After we achieved locking we placed the lid of the OMC back and turned off the laser. 

  502   Tue Mar 7 10:20:55 2023 Thejas OpticsCharacterizationA+ OMC Curved mirror radius of curvature

[Camille, Stephen, Thejas]

Stephen returned the curved mirror #6 to Liyuan for point transmission measurement. We are now using #5 for to setup/align the ZYGO Fizeau interferometer setup to characterize the curvature center of the mirrors. It was setup such that the focal point of the input reference sphere was coincident with the radius of curvature of the test mirror. 

The curved mirror was mounted on a flat reference mirror, with the help of the sub-assembly bonding fixture:

The fringe pattern seen was:

 

 

Efforts were made today to improve the contrast of the fringe pattern and take some measurements.

 

 

 

 

 

 

 

Attachment 1: image.jpeg
image.jpeg
Attachment 3: image.jpeg
image.jpeg
  503   Tue Mar 7 23:00:16 2023 Thejas OpticsCharacterizationOMC Lab clean up

Today, Koji and I cleaned up the the lab space and made some space on the optical table for radius of curvature measurement of the A+ OMC curved mirrors. 

 

 

  504   Wed Mar 8 17:27:51 2023 Thejas OpticsCharacterizationA+ OMC Curvature minimum of curved optics

[Stephen, Thejas]

Today, a more rigorous effort was made to re-measure the position of the optics forming the Fizeau cavity and re-position the curved optic to get more contrasting fringes. Distance measurements were made using a Fluke laser displacement sensor. We obtained a contrasting fringe pattern but the phase profile measured was assymmeteric and un-satisfactory. Tomorrow an attempt will be made to place an iris infront of the curved optic to define the edge of the beam and limit it only to the curved optic surface. 

Attachment 1: image.jpeg
image.jpeg
  505   Fri Mar 10 10:23:08 2023 ThejasOpticsCharacterizationA+ OMC curved mirror radius of curvature

OMC test set-up

Yesterday, laser beam output from the fibre follwoing teh mode-matching lenses was picked off and beam profile was characterized using beam profiler Thorlabs BP209-VIS. 

 

 

The gaussian fit beam diameter was measured to be about wx = 939 um wy = 996 um at the location of a distance of 0.4 m from the high reflector. The mode content of this beam is about 98% TEM00. We want to use this beam within the Rayleigh range (near field) to measure radium of curvature of the curved optics. 

The Rayleigh range is about 0.74 m. 

 

Attachment 4: OMC_8_(dragged).pdf
OMC_8_(dragged).pdf OMC_8_(dragged).pdf OMC_8_(dragged).pdf
  506   Fri Mar 10 11:12:57 2023 Thejas OpticsCharacterizationA+ OMC Curvature minimum of curved optics

[GariLynn, Stephen, Thejas]

Yesterday, we placed an iris (borrowed from OMC Lab) infront of the spherical transmission sphere to limit the spot size, on the other end of the cavity, to only the curved optic. This produced a crisp boundary for the interference pattern. We obtained some data at different imaging focal planes. The transmission optic here is a spherical mirror. This was replaced with a plane reference and the curved optic was moved closer to this optic. Intereference fringes were nuled for the plane mirror upon which the curved optic sits. This ensures that the curved mirror is head on to the laser beam. The spherical fringes were obscured by some diffraction artifacts. Today, we will be makign an attempt to eliminate that and try to see fringes from the whole curved optic. 

 

 

  507   Tue Mar 14 10:41:06 2023 Thejas OpticsCharacterizationA+ OMC curved mirror characterization

[Camille, GarriLynn, Stephen, Thejas]

Folllowing the replacement of the spherical transmission / reference mirror with a flat mirror, on Friday we were able to observe fringes that facilitated characterization of the curvature minimum. 

 

\

By rotating the curved optic by 90 deg we couodn't reproduce consistent data. 

This is probably due to insufficient attention given to the orientation/centering of the curved mirror under the clamp. 

 

 

 

Attachment 1: image.jpeg
image.jpeg
  508   Tue Mar 14 12:12:41 2023 Koji OpticsCharacterizationA+ OMC curved mirror characterization

RoC: 2.65m ! Interesting. I'll wait for the follow-up analysis/measurements. The RoC may be dependent on the area (diameter) for the fitting. You might want to run the fitting of your own. If so, let me know. I have some Matlab code that is compatible with the CSV file exported from MetroPro data.

  509   Tue Mar 14 18:24:03 2023 Thejas OpticsCharacterizationA+ OMC Curved mirror radius of curvature

Today, I tried to measure the radius of curvature of the curved mirror using the input beam for the OMC test set-up. It was noticed that the half inch curved optic (ROC=2.5 m), when placed within the Rayleigh range of the beam waist, did not focus the beam. This is probably becasue the beam diameter is small for this optic's radius of curvature to produce any focussing. This can be illustrated even further using the JAMMT software by replacing a concave sperical mirror with a ocnvex lens of focal length of 1.25 m. 

Substrate: 1/2 inch optic with f= 0.25 m 

Substrate: 1/2 inch optic with f= 1.25 m

 

 

Substrate: 1/2 inch optic with f= 1.25 m

The only wasy to resolve this is by incresing the beam diameter to > 2 mm

 

  510   Tue Mar 14 20:06:03 2023 Koji OpticsCharacterizationA+ OMC Curved mirror radius of curvature

If the mirror has the RoC, it works as a lens. And you should be able to see the effect in the beam profile.

Just what you need to do is to compare the beam profile without the mirror (or with a flat mirror) and then with the curved mirror.

 

  511   Wed Mar 15 15:28:24 2023 Thejas OpticsCharacterizationA+ OMC Curved mirror radius of curvature

Thanks for teh comment Koji. Yes, I did see this effect by comparing the beam sizes with and without the curved mirror. But the observation did not conform with the expectation that the beam should focus at a distance of 1.25 m from the curved mirror (as seen in the software images). So, I plan to use some lenses to increase the beam waist and perform the measurement.

Quote:

If the mirror has the RoC, it works as a lens. And you should be able to see the effect in the beam profile.

Just what you need to do is to compare the beam profile without the mirror (or with a flat mirror) and then with the curved mirror.

 

 

  512   Wed Mar 15 17:07:35 2023 Koji OpticsCharacterizationA+ OMC Curved mirror radius of curvature

I hope you can find useful lenses from the lens kit in the cabinet. If you need more lenses and mounts, talk to our students in WB and the 40m.

  513   Fri Mar 17 15:01:21 2023 Koji OpticsCharacterizationA+ OMC Curved mirror radius of curvature

Thanks Koji, the lenses available in the cabinet in the lab actually sufficed. 

Quote:

I hope you can find useful lenses from the lens kit in the cabinet. If you need more lenses and mounts, talk to our students in WB and the 40m.

 

  514   Fri Mar 24 07:38:54 2023 Camille MakaremOpticsCharacterizationROC measurements of the curved mirrors

[Thejas, Camille]
21 March 2023

We made slight adjustments to the beam expander lenses in the ROC setup. The position of the second lens was moved slightly (a few mm) to improve the collimation of the beam.  The beam profiler was used to measure the beam size at various distances (measurements attached). This will be used to characterize the beam divergence.
This beam was reflected off the curved mirror and the beam profiler was used to measure the beam size at various positions near the focal point. This process was repeated for various curved mirrors (measurements attached). These values will be used to determine the ROC of each mirror. ROC=2*FL

  515   Fri Mar 24 07:47:37 2023 Camille MakaremOpticsCharacterizationROC measurements of the curved mirrors

[Camille]
22 March 2023
Beam profile measurements were continued for more of the curved mirrors.
Mirror sn07 was repeated to verify that Camille and Thejas get the same focal length measurement (plot attached).

Attachment 1: CurvedMirror_FL_measurments.PNG
CurvedMirror_FL_measurments.PNG
Attachment 2: sn07.PNG
sn07.PNG
  516   Tue Mar 28 11:21:27 2023 Camille MakaremOpticsCharacterizationSagitta measurements of curved mirrors

[Camille, Stephen, Thejas]

Curved mirror sn02 was used to test the method for collecting Zygo measurements on the curved mirrors. The curved mirror was mounted with its back surface against a reference flat. The reference flat was pitched/yawed until its fringes were nulled. Then a measurement of the surface profile of the curved mirror + flat mirror together was taken.
The curved mirror was rotated in 90deg increments and the measurements were repeated. (5 measurements in total were taken, with the curved mirror's fiducial in the 12:00, 3:00, 6:00, 9:00 and 12:00 again positions.) The curvature minumum was seen to clock as expected with the rotation of the mirror.

The attached figures show the surface profile of the central 8.5 mm diameter of the mirror (central with respect to the coating edge). Also attached is a plot of the surface profile across the line drawn in the figure.

 

Attachment 1: CurvedMirrorSN02_SurfaceProfiles.PNG
CurvedMirrorSN02_SurfaceProfiles.PNG
  517   Mon Apr 3 11:06:47 2023 Thejas OpticsCharacterizationFlat mirror flatness

Data can be found in DCC document T2300050. 

 

On Friday. Camille and I measured the flatness of the flat mirror.  Tilt values (without subtracting tilt) were less than 100 nm and PV across the surface was about 50 nm. 

This checks that the flat mirror surface distortions are not contributing to the systematic deviations in our measurement of curvature minimum with varying the fiducial clocking angle. The deviations in the data show a far more disagreement between Y-Tilt of different clocking angles than the X-Tilt. 

 

 

  518   Tue Apr 4 16:14:06 2023 KojiOpticsLoan / LendingC7 mirror to Downs

I handed Camille the C7 mirror for the cross-calibration of the ROC characterization techniques.

  520   Tue Apr 11 16:18:25 2023 ThejasOpticsCharacterizationA+ OMC Curved mirror radius of curvature

Curved mirror radius of curvature raw data can be found in the DCC document: T2300050

The input beam falling on the curved optic was characterized. The beam waist and it's position was found by curve fitting gaussian beam propagation formula in near field:

Fitting gives the following values for the initial beam waist (w_0) and waist position (z_0) (see pdf attached below).

Using these fitted parameters in JAMMT (beam propagation software) gives the following results for a 1.25 m focal length optic:

P-plane

w_0x: 1.429 mm +/- 0.006 mm
z_0x: 0.421 m +/- 0.131 m

Beam waist @ 3.063 +/- 0.005 m

f = 1.25 m optic @ 1.807 m

Thus beam focuses at 1.256 +/- 0.005 m for the p-plane.

S-plane

w_0y: 1.526 mm +/- 0.023 mm
z_0y: 0.352 m +/- 0.597 m

Beam waist @ 3.064 +/- 0.02 m 

f = 1.25 m optic @ 1.807 m 

Thus beam focuses at 1.257 +/- 0.02 m for the s-plane. 

Hence we can use the distance measured from the optic to the beam profiler as a suitable figure for focal length (hence radius of curvature). Also astigmatism in the input beam is calculated to have negligible influence in causing astigmatism in following the curved optic. Hence, any astigmatism measured at the focus following the curved optic is due to the curved optic (?). Also because the incident beam on the curved optic is at an arbitrary angle of incidence, this introduces further astigmatism in the reflected beam given by equation 12 in this paper: https://opg.optica.org/ao/fulltext.cfm?uri=ao-8-5-975&id=15813

 

 

 

 

Attachment 3: ini_beam.pdf
ini_beam.pdf
  521   Thu Apr 13 07:47:28 2023 Camille MakaremOpticsCharacterizationZygo setup for curved mirror measurements

[Camille, Thejas, Stephen]

We modified the Zygo setup for measuring the sagitta of the curved mirrors. A mirror at 45deg was used to reflect the interferometer beam down towards the surface of the table (see picture). A fused silica flat was placed horizontally with the surface of the table and was used as our reference flat. The back surface of the curved mirror and the top surface of the reference flat were cleaned using top gun and/or swabs. Once it was verified that the surfaces were clean, the curved mirror could be easily placed on the surface of the reference flat.

Once the curved mirror was placed on the reference flat, the fringes of the reference flat were nulled using the 45deg mirror. After nulling the flat's fringes, the data was recorded. The curved mirror was then rotated 90deg clockwise. The measurment was repeated with the curved mirror's fiducial located at 12:00, 3:00, 6:00 and 9:00. The 12:00 position was measured twice to ensure repeatability. (A drop of first contact had been placed at the edge of the optic to indicate where the fiducial arrow is. This helped with clocking alignment.)

The already-characterized aLIGO C7 mirror was measured to verify the setup. After verifying agreement with the results in T1500060, this process was repeated with all the remaining curved mirrors.

The data was analyzed using Thejas's python script (separation distance between mirror center and curvature minimum, angular position of curvature minimum.) Those mirrors with a large spread in the measurements will be remeasured.

 

Attachment 1: dist_angle_spread.PNG
dist_angle_spread.PNG
  522   Fri Apr 14 15:52:53 2023 Camille MakaremOpticsCharacterizationDetermining the curvature bottom of the curved mirrors

[Camille, Thejas]

We repeated Zygo measurements (using the same setup and method as below) for curved mirrors sn07, sn11, sn12, sn18, sn19, sn25, sn26, and sn30.

sn11 and sn25 still show a large spread in angular measurements (see attached.) This is attributed to the low decentering values for these two mirrors (0.072mm and 0.158mm, respectively).

Attachment 1: dist_angle_spread_v2.PNG
dist_angle_spread_v2.PNG
  523   Fri Apr 14 16:36:29 2023 ThejasOpticsCharacterizationDetermining the curvature bottom of the curved mirrors

Attached here with are relevant plots.

Quote:

[Camille, Thejas]

We repeated Zygo measurements (using the same setup and method as below) for curved mirrors sn07, sn11, sn12, sn18, sn19, sn25, sn26, and sn30.

sn11 and sn25 still show a large spread in angular measurements (see attached.) This is attributed to the low decentering values for these two mirrors (0.072mm and 0.158mm, respectively).

 

Attachment 1: sn129-02.pdf
sn129-02.pdf
Attachment 2: sn129-04.pdf
sn129-04.pdf
Attachment 3: sn129-05.pdf
sn129-05.pdf
Attachment 4: sn129-06.pdf
sn129-06.pdf
Attachment 5: sn129-07.pdf
sn129-07.pdf
Attachment 6: sn129-11.pdf
sn129-11.pdf
Attachment 7: sn129-12.pdf
sn129-12.pdf
Attachment 8: sn129-14.pdf
sn129-14.pdf
Attachment 9: sn129-16.pdf
sn129-16.pdf
Attachment 10: sn129-17.pdf
sn129-17.pdf
Attachment 11: sn129-18.pdf
sn129-18.pdf
Attachment 12: sn129-19.pdf
sn129-19.pdf
Attachment 13: sn129-20.pdf
sn129-20.pdf
Attachment 14: sn129-21.pdf
sn129-21.pdf
Attachment 15: sn129-22.pdf
sn129-22.pdf
Attachment 16: sn129-23.pdf
sn129-23.pdf
Attachment 17: sn129-25.pdf
sn129-25.pdf
Attachment 18: sn129-26.pdf
sn129-26.pdf
Attachment 19: sn129-29.pdf
sn129-29.pdf
Attachment 20: sn129-30.pdf
sn129-30.pdf
Attachment 21: Screenshot_2023-04-14_at_16.40.20.png
Screenshot_2023-04-14_at_16.40.20.png
  524   Mon Apr 17 17:56:25 2023 Thejas OpticsCharacterizationA+ OMC Piezos wedge angle

PZT dimension analyzed and characterized. The blue dot in the images represents the position of the cathode. The length of the arrows indicates the amount of wedging. 

 

Attachment 2: PZT30.pdf
PZT30.pdf
Attachment 3: PZT31.pdf
PZT31.pdf
Attachment 4: PZT32.pdf
PZT32.pdf
Attachment 5: PZT33.pdf
PZT33.pdf
Attachment 6: PZT35.pdf
PZT35.pdf
Attachment 7: PZT34.pdf
PZT34.pdf
Attachment 8: PZT36.pdf
PZT36.pdf
Attachment 9: PZT37.pdf
PZT37.pdf
Attachment 10: PZT38.pdf
PZT38.pdf
Attachment 11: PZT39.pdf
PZT39.pdf
Attachment 12: PZT40.pdf
PZT40.pdf
Attachment 13: PZT42.pdf
PZT42.pdf
Attachment 14: PZT41.pdf
PZT41.pdf
Attachment 15: PZT43.pdf
PZT43.pdf
Attachment 16: PZT44.pdf
PZT44.pdf
Attachment 17: PZT45.pdf
PZT45.pdf
Attachment 18: PZT47.pdf
PZT47.pdf
Attachment 19: PZT46.pdf
PZT46.pdf
  526   Wed Apr 19 14:28:08 2023 ThejasOpticsCharacterizationA+ OMC Curved mirror radius of curvature

Restimation of the parameters

Camille and I went back to the lab to re-measure the beam profile follwoing the beam expanding lenses. I was observed after turning on the laser that the beam spot on  the turning mirror had displaced off to the mirror edge. We had to re-align the beam.

We have the following parameters from the fitting now, see attached.

w_0x: 1.44 mm +/- 0.0016 mm 
z_0x: 0.575 m +/- 0.046 m
w_0y: 1.50 mm +/- 0.0014 mm
z_0y: 0.004 m +/- 0.029 m

For the p - plane:

Beam waist occurs at 1.249 m +/- 0.002 m from the curved optic of f = 1.25 m 

For the s plane:

Beam waist occurs at 1.269 m +/- 0.001 m from the curved optic of f = 1.25 m 

And the angle of incidence on th ecurved optic was astimated to be 2.66 deg. This imparts a very negligible astigmatism in the reflected beam. But after collecting more data points of the beam profile we see that there is significant astigmatism in the input beam which translates to a decent amount of astigmatism in the reflected beam. 

 

Quote:

Curved mirror radius of curvature raw data can be found in the DCC document: T2300050

The input beam falling on the curved optic was characterized. The beam waist and it's position was found by curve fitting gaussian beam propagation formula in near field:

Fitting gives the following values for the initial beam waist (w_0) and waist position (z_0) (see pdf attached below).

Using these fitted parameters in JAMMT (beam propagation software) gives the following results for a 1.25 m focal length optic:

P-plane

w_0x: 1.429 mm +/- 0.006 mm
z_0x: 0.421 m +/- 0.131 m

Beam waist @ 3.063 +/- 0.005 m

f = 1.25 m optic @ 1.807 m

Thus beam focuses at 1.256 +/- 0.005 m for the p-plane.

S-plane

w_0y: 1.526 mm +/- 0.023 mm
z_0y: 0.352 m +/- 0.597 m

Beam waist @ 3.064 +/- 0.02 m 

f = 1.25 m optic @ 1.807 m 

Thus beam focuses at 1.257 +/- 0.02 m for the s-plane. 

Hence we can use the distance measured from the optic to the beam profiler as a suitable figure for focal length (hence radius of curvature). Also astigmatism in the input beam is calculated to have negligible influence in causing astigmatism in following the curved optic. Hence, any astigmatism measured at the focus following the curved optic is due to the curved optic (?). Also because the incident beam on the curved optic is at an arbitrary angle of incidence, this introduces further astigmatism in the reflected beam given by equation 12 in this paper: https://opg.optica.org/ao/fulltext.cfm?uri=ao-8-5-975&id=15813

 

 

 

 

 

Attachment 1: ini_beam.pdf
ini_beam.pdf
  527   Mon Apr 24 15:29:48 2023 Camille MakaremOptics summary of zygo setups

Summary of Zygo setups

Initial Zygo Setup:
Our initial Zygo setup consisted of a flat transmission sphere with the 0.5" curved mirror mounted against a 1" flat mirror.
Mounting procedure:
The bottom part of the gluing fixture was attached to a mounting plate using two screws. The 1" reference flat was placed on the gluing fixture. The reference flat was inspected with a green flashlight to ensure that there was no dust on the mirror surface. Any dust was removed using top gun. If any dust remained after using top gun, it was removed with a swab.
The back surface of the curved mirror was inspected and cleaned using the same method (flashlight inspection, followed by top gun if necessary, followed by swab if necessary).

After ensuring that both surfaces are clean, the back surface of the curved mirror was placed on the front surface of the reference flat. The fiducial of the curved mirror was positioned at 12:00. (12:00 is defined as the top of assembly.) The two mirrors were held in place using a mounting plate with a 0.4" aperture. The mounting plate was fixed to the bottom part of the gluing fixture using two screws and a spring for each screw (see attached picture).

The mounting plate holding this assembly was then attached to a optical mount with tip/tilt adjustments (see attached picture).
This assembly was placed facing the Zygo transmission flat (see attached picture) and the mount was pitched/yawed until the fringes on the 1" reference flat were nulled. After nulling the fringes, the data was then recorded.

The mounting plate was then removed from the tip/tilt mount and dissassembled so that the curved mirror could be rotated so that the fiducial is in the 3:00 position. The procedure is then repeated and the data recorded.
This was repeated again with the fiducial in the 6:00, 9:00 and 12:00 (again) positions.

Review of this data shows that the positions of the curvature minimums was not reproducible with sufficient precision. A teflon mounting plate was added to clamp the 1" reference flat more securely to the gluing fixture (See attached pictures). Data was collected in the same manner (twice with the fiducial at 12:00 and once with fiducial at each of the positions 3:00, 6:00, and 9:00).
Additional data collected still failed to produce reproducible results and the removing/remounting process of the curved mirror was time-consuming, so we attempted a new setup for the Zygo measurments.

Final Zygo Setup:
The new setup used a fold mirror mounted at 45degrees to direct the Zygo beam downwards into the plane of the table. A 3" flat was used as our reference flat. The reference flat was placed on some lens tissue parallel to the plane of the table. The same inspection and cleaning method was used to ensure that there was no dust on the reference flat (flashlight inspection, followed by top gun if necessary, followed by swabbing if necessary).
The back of the curved optic was inspected and cleaned using the same method. The curved mirror was placed on the 3" reference flat with the fiducial at the 12:00 position. (12:00 here is defined as the direction ponting towards the Zygo instrument.) (See attached picture of this setup.)
The fold mirror was pitched/yawed so that the fringes on the 3" reference flat were nulled. (An additional advantage of this setup is that more surface of the reference flat was viewable.) After nulling the fringes, the curved mirror was picked up and replaced a few times to verify that the fringe pattern on the curved mirror appeared reproducible. The data was collected with the fiducial at the 12:00 position. This process was repeated with the fiducial at 3:00, 6:00, 9:00, and again at 12:00.
Results from this setup were reproducible and this setup was used to measure the surface profile of all the curved mirrors.

Attachment 1: IMG_0153.jpg
IMG_0153.jpg
Attachment 2: IMG_2551.jpg
IMG_2551.jpg
Attachment 3: IMG_0183.jpg
IMG_0183.jpg
Attachment 4: IMG_0157.jpg
IMG_0157.jpg
Attachment 5: IMG_3665.jpg
IMG_3665.jpg
Attachment 6: IMG_0203.jpg
IMG_0203.jpg
  528   Wed Apr 26 13:23:00 2023 StephenOpticsCharacterizationPZT DC Response with Shadow Sensor - QPD Calibration

[Stephen, Camille, Thejas, with support from Marie, Dean, Luis]

We setup a shadow sensor! (Attachment 1)

  • QPD (Thorlabs PDQ80A) and HeNe laser (< 4 mW @ 633 nm, Thorlabs HNLS008R, 12V / .7A) borrowed from Marie.
    • QPD supplied by 5V from DC power supply.
  • Razor blade flag used to eliminate effect of pitch misalignment, atop a 62.5 gram mass.
  • Lab jack used to provide height adjustment.
  • Dial indicator (Mitutoyo Absolute +/- .001 mm, p/n S112TXB) mounted to height gauge, used to monitor stage height.

The calibration measurement proceeded as follows:

  • Manually lowered the stage until the flag was not occluding any light on the QPD
  • Raised the stage in 0.5 mm increments and recorded raw data (using the oscilloscope "Sampling" mode), until the flag occluded all of the light previously on the QPD.
  • Identified the central 0.5 mm increment, and stepped through that range in .05 mm increments

Data collected in table at T2300050 Optical Component Testing Measurements, sheet "PiezoDCResponse"

In the linear range from stage position [3.450, 3.550] the least squares linear fit is:

QPD_Sum = m * Stage_Position + b

  • QPD_Sum is the dependent variable of QPD signal voltage, derived from the Channel 1 (sum) average, unit V.
  • m = -9.685 V / mm is the response of the QPD to change in stage position.
  • Stage_Position is the independent variable of stage height, observed via dial indicator readout, unit mm.
  • b = 35.6111 V is the best fit y-intercept - not a physical quantity.

So, does this setup allow us to measure the DC response of the PZT?

  • Over the linear range, the "m" sensitivity parameter would correspond with a signal of .009685 V / micron
  • We expect a stroke of 3 or 4 microns, yielding a signal of .03 to .04 V.
    • (Noliac NAC2124 nominal free stroke is 3.3 microns +/- 15% for a maximum operating voltage of 200 V)
  • The typical standard deviation of each measurement is .009 to .011 V.
    • This would be a large error bar in comparison to the signal level.
  • We will try again using a different photodiode.

 

Attachment 1: qpd_dc_response_setup_image_20230425.jpg
qpd_dc_response_setup_image_20230425.jpg
Attachment 2: Shadow_Sensor_Stage_Calibration_plot_20230426.png
Shadow_Sensor_Stage_Calibration_plot_20230426.png
  530   Fri Apr 28 13:25:40 2023 Thejas OpticsGeneralaLIGO OMC Unit-4

[Camille, Koji, Thejas]

Yesterday, we cleaned the cavity optics with first contact, aligned the input laser beam to the cavity and measured the power at different terminals on the cavity breadboard. 

The measured OMC losses were:
SET1 0.042 +/- 0.003
SET2 0.035 +/- 0.002
SET3 0.030 +/- 0.0014
-> 0.033 +/- 0.001

The measured OMC mode-matching efficiencies were:
SET1 0.9795 +/- 0.00016
SET2 0.9797 +/- 0.00005
SET3 0.9794 +/- 0.00035

 

Attached herewith is the scrrenshot of the notes of with input power parameters.

Attachment 1: image.jpeg
image.jpeg
Attachment 2: PXL_20230428_010507088.png
PXL_20230428_010507088.png
  531   Fri Apr 28 14:24:03 2023 Thejas OpticsCharacterizationA+ OMC Matching

[Thejas, Camille, Stephen] 

Here are some notes on how I plan to approach matching of the PZTs, mounting prisms and curved optics. 

Step 1: Match the prisms and the PZTs such that resulitng 18 combiations will have minimum vertical wedging.

- I will be usign scipy.optimize.minimize to implement this.

Step 2: Arrange the curved mirror wedge angles ascending order. This prioritizes matching of low wedge angled mirorrs first. The high wedge angled ones have a much larger range of vertical component of wedge angle due to freedom of rotation of the mirrors. Attention should also be given to error in the wedge angle due to phase spread of the various clocking data. The more the wedge angle, the more it is sensitive to this error.

- This will be implemented using standard loops. 

 

  532   Mon May 1 13:30:40 2023 StephenOpticsCharacterizationPZT DC Response with Shadow Sensor - PD Calibration

[Stephen, Camille, Thejas, with support from Luis]

Calibration reattempted with the PD borrowed from Koji, equivalent to the last PZT measurment (OMC_Lab/336). There were a couple of differences in contrast to the last measurement:

  • We navigated to the central 1 mm of the range (the interval we found to be sensitive in the QPD calibration, dictated by the beam size) and we stepped through in .05 mm intervals.
  • We had the oscilloscope on 10x Voltage probe mode. See setup in Attachment 1.

Data collected in table at T2300050 Optical Component Testing Measurements, sheet "PiezoDCResponse" (Day 2 section).

In the linear range from stage position [3.450, 3.550] the least squares linear fit is:

QPD_Sum = m * Stage_Position + b

  • QPD_Sum is the dependent variable of QPD signal voltage, derived from the Channel 1 (sum) average, unit V.
  • m = -136.45 V / mm is the response of the QPD to change in stage position.
  • Stage_Position is the independent variable of stage height, observed via dial indicator readout, unit mm.
  • b = 501.91 V is the best fit y-intercept - not a physical quantity.

So, does this setup allow us to measure the DC response of the PZT?

  • Over the linear range, the "m" sensitivity parameter would correspond with a signal of .13645 V / micron
  • We expect a stroke of 3 or 4 microns, yielding a signal of .40 V to .55 V.
    • (Noliac NAC2124 nominal free stroke is 3.3 microns +/- 15% for a maximum operating voltage of 200 V)
  • The standard deviation was not consistent for each measurement; at this linear range, the standard deviation was between .10 V and .15 V.
    • This would be a large error bar in comparison to the signal level

So, is this DCPD setup better than the QPD?

  • Comparing sensitivity against standard deviation, there is not much difference.
    • Standard deviation is the same for each measurement in the QPD case, while the PD has some variation (as the signal increases, the standard deviation also increases, but not with uniform scaling; at lower signal levels, the standard deviation is smaller for the PD than the QPD).
    • The ratio between Channel 1 Average and Standard Deviation is similar for the two setups, so neither setup reduces the error significantly.

We will probably just keep the PD in place, since there is not a great motivation to revert to the QPD, and the QPD could then be used for the OpLev independently. We will look at using the oscilloscope "Averaging" mode to reduce the noise in our measurement.

Attachment 1: pd_pzt_dc_response_setup_image_20230426.jpg
pd_pzt_dc_response_setup_image_20230426.jpg
Attachment 2: Shadow_Sensor_Stage_Calibration_plot_PD_20230427.png
Shadow_Sensor_Stage_Calibration_plot_PD_20230427.png
  533   Tue May 2 12:49:43 2023 StephenOpticsCharacterizationPZT DC Response with Shadow Sensor - First measurement attempt

[Stephen, Camille, Thejas]

A running log of our efforts from Monday 1 May. Data continues to be placed in T2300050 at sheet "PiezoDCResponse":

  • Continuing with the DCPD and 3-axis piezo driver for our "final" setup
    • PD Thorlabs
    • Driver Thorlabs MDT694B
    • Function Generator in line
  • "Averaging" mode of the oscilloscope only reported 3 significant figures, so there was no benefit from switching away from "Sampling" mode
  • Acoustic buzzing test of PZT 36 and PZT 31 yielded consistent results
    • Audible buzzing tone with 1 kHz, 10 Vpp input
    • Reverse polarity also buzzes
  • We attempted on/off testing, and also AC drive testing, for PZTs 31, 32, and 36
    • PZT 36 was used in the initial setup effort, and since there was inadequate insulation of the PZT from the setup, there was charge buildup and static discharge during that early effort. This PZT was chosen because it has the greatest wedge. We should be skeptical of it now that we have used it in these setup efforts.
    • PZT 31 was used next, also during setup but only after adequate insulation of the PZT had been implemented. It probably has been treated well! Except during today's efforts, one of the solder joints debonded. (Need to follow up with Dean to confirm that LIGO soldering practice follows recommendations of vendor - pads should be clean, solder should include Ag; see D1102070 and other sources)
    • PZT 32 was used afterwards, and was measured.
  • We were not really satisfied with the "feel" of the measurement
    • The oscilloscope output didn't seem to change on time scales that we were expecting, when we manipulated the frequency of the drive
    • We attempted to directly measure the PZT extension using the dial indicator, but did not succeed
  • We tried to make a list of items that we didn't understand fully, and came up with these:
    • We aren't really that familiar with the driver/amplifier and how it interacts with the input from the function generator:
      • Thorlabs documentation was reviewed - driver manual
        • https://www.thorlabs.com/drawings/d90d63f542e805de-80E05D38-00C3-80BB-4CF17F9FDBF42AB6/MDT693B-Manual.pdf
      • V_out = V_manual + 15 * V_external
        • We applied a 2 V peak-to-peak external voltage, so we were not driving through the full range of the output voltage
        • We may benefit from a low frequency, triangular wave drive so that we can better monitor the output voltage
    • Maybe the time constant of the capacitive charging is too long?
      • Thorlabs documentation was reviewed - Piezo Bandwidth section of Piezo Tutorial
        • https://www.thorlabs.com/newgrouppage9.cfm?objectgroup_id=5030
      • Slew rate (rate of capacitive charging, V/s) is more complicated for a sinusoidal wave than a triangular wave, but the maximum is characterized by the Maximum Driver Current / System Capacitance.
        • Max Driver Current I_max = .06 A, System Capacitance C_sys = C_PZT + C_driver = 5.1 E-7 F + 1 E-9 F
        • Slew rate = 1.18 E5 V/s, or a charging time of 1.28 E-3 s to reach 150 V (note that this is not a true value as the input waveform was sinusoidal)
      • More helpfully, the system bandwidth for a sinusoidal wave is f_max = I_max / (pi * V_pp * C_sys); a triangular wave with a minimum of 0 V is f_max = I_max / (V_pp * C_sys)
        • Our system, if driven sinusoidally from 0 to 150 V, would have a bandwidth of 249 Hz
        • Our system, if driven triangularly from 0 to 150 V, would have a bandwidth of 783 Hz
  • We should try to get a low frequency triangular wave drive output from 0 to 150 V with a frequency of 100 Hz, and see if that generates any meaningful signal.

The measurement "results" for PZT 32 (note that these results seem to be dominated by a slow drift in the measurement, and this measurement was not reproducible):

  • PD signal with PZT 32 on (150 V) : 3.242934 V, standard deviation 0.013894 V
  • PD signal with PZT 32 off (0 V): 3.315395 V, standard deviation 0.014658 V
  • sensitivity: -13.65102 V per mm PZT displacement
    • (from PD Calibration, removing factor of 10 from oscilloscope during that measurement)
  • displacement of PZT = 5.308 microns
    • the standard deviation of the On measurement was 1.018 microns, and the standard deviation of the Off measurement was 1.074 microns.
  • Conclusion: this measurement has a rather large error bar, and was not very repeatable / could not be directly observed by another means (see above comments in log) we were not really satisfied with the "feel" of the measurement
    • The oscilloscope output didn't seem to change on time scales associated with low frequency AC drive; Since high frequency AC drive does cause motion (witnessed by buzzing) maybe we need to find other ways to measure the motion that are more sensitive?
  534   Thu May 4 11:39:19 2023 StephenOpticsCharacterizationPZT DC Response with Shadow Sensor - First measurement attempt

Efforts from Tuesday through Thursday Wednesday, 02 through 04 May, are described below.

The main outcomes were:

  • We are able to see some response in the frequency domain (which confirms the response that we hear at acoustic drive frequencies) on a few PZTs Wednesday and Thursday.
  • We then improved our drive input and monitoring, so we now see the full waveform delivered to the PZT.
  • Now, we are able to reliably see some length response, though typically the reponse seems to be somewhat less than the expected 3.3 micron / 200 V from the spec sheet.
  • We are also still subject to low SNR, though at least now the DC response is visible to the eye.
    • We also recalibrated the PD using the stage in the current setup.
  • We met this morning to talk about data collection, and we realized it would be best to try to improve the setup.
    • Koji provided some good suggestions, some of which reflected feedback collected from Gabriele, Dean, and Luis.
    • Our setup was sensitive enough to observe the response (yay) but not sensitive enough to make a good measurement.

The next log will reflect the updated setup.

I hope to come back to edit this log to fill out more detail, but the detail reflects intermediate steps with the old setup.

  536   Fri May 5 18:37:31 2023 StephenOpticsCharacterizationPZT DC Response with Shadow Sensor - First measurement attempt

[Thejas, Camille, Stephen]

In Downs 227 on 05 May 2023

We took the following actions:

  • Reconfigure our data acquisition with a SR785 instead of the oscilloscope.
    • The scope had been limited to ~10 mV precision, and the SR785 has better than mV precision.
  • Contract the beam using some lenses borrowed from Gabriele.
    • Increases sensitivity, following some feedback from Koji which Gabriele had also mentioned
    • We measured the spot size on our flag to be .44 mm (had been about 1 mm).
  • Moved all electronics, especially AC function generator, off of optical table.
    • Feedback from Koji.

We recalibrated the photodiode using the razorblade flag. We continued to be plagued by slow variations but we had much better data quality (less fuzziness and fewer spikes in the time series). It seemed like the razorblade was not stiff enough and was sensitive to airflow.

We may readjust and recalibrate to see if our data is better, but in the meantime, below are images of the calibration and setup. Data is at T2300050 Optical Component Testing Measurements, sheet "PiezoDCResponse".

Revisiting the measurement requirements and applying the sensitivity of the new setup:

  • The NAC2124 PZT nominally extends 3.3 microns with a 200 V drive, +/- 15%.
    • Our 150 V drive should extend the PZT by 2.5 microns nominally.
    • Our shadow sensor would ideally be able to observe with accuracy better by order of magnitude (.25 microns).
  • Our stage-driven, dial-indicator-observed PD calibration reveals a sensitivity of -6.704 V PD output per mm of displacement.
    • This corresponds with 149 microns of displacement per V PD output.
    • Across these 11 stage calibration measurements, the error terms were:
      • Standard Deviation: max. 0.00182 V, min. 0.00008 V, avg. 0.00069 V (no, I'm not going to report the standard deviation of the standard deviations.)
      • Range, V (max. value - min. value): max. 0.00563V, min. {C} 0. 00042 V, avg. 0. 00232 V
  • If we want to measure with accuracy of .25 microns, that corresponds with a PD output of .0017 V
    • Looks like some of our measurements had low enough error for this accuracy, but there were some measurements which did not.
    • Hopefully we can improve by reducing the impact of vibration and airflow.

 

Attachment 1: Shadow_Sensor_Stage_Calibration_plot_PD_smaller_beam_20230505.png
Shadow_Sensor_Stage_Calibration_plot_PD_smaller_beam_20230505.png
Attachment 2: updated_setup_smaller_beam_spectrum_analyzer_20230505.jpg
updated_setup_smaller_beam_spectrum_analyzer_20230505.jpg
  539   Mon May 15 14:20:03 2023 Thejas OpticsCharacterizationA+ OMC ROC

Herewith attached are the results of curved mirror radius of curvature characterization. 

 

 

Attachment 1: ROCsn02.pdf
ROCsn02.pdf
Attachment 2: ROCsn04.pdf
ROCsn04.pdf
Attachment 3: ROCsn05.pdf
ROCsn05.pdf
Attachment 4: ROCsn06.pdf
ROCsn06.pdf
Attachment 5: ROCsn07.pdf
ROCsn07.pdf
Attachment 6: ROCsn11.pdf
ROCsn11.pdf
Attachment 7: ROCsn12.pdf
ROCsn12.pdf
Attachment 8: ROCsn16.pdf
ROCsn16.pdf
Attachment 9: ROCsn17.pdf
ROCsn17.pdf
Attachment 10: ROCsn18.pdf
ROCsn18.pdf
Attachment 11: ROCsn19.pdf
ROCsn19.pdf
Attachment 12: ROCsn20.pdf
ROCsn20.pdf
Attachment 13: ROCsn21.pdf
ROCsn21.pdf
Attachment 14: ROCsn22.pdf
ROCsn22.pdf
Attachment 15: ROCsn23.pdf
ROCsn23.pdf
Attachment 16: ROCsn25.pdf
ROCsn25.pdf
Attachment 17: ROCsn26.pdf
ROCsn26.pdf
Attachment 18: ROCsn29.pdf
ROCsn29.pdf
Attachment 19: ROCsn30.pdf
ROCsn30.pdf
Attachment 20: ROCsn14.pdf
ROCsn14.pdf
Attachment 21: table.png
table.png
  540   Mon May 15 14:24:41 2023 Thejas OpticsCharacterizationA+ OMC ROC

Koji's mirror measurement result attached herewith for comparison. 

Quote:

Herewith attached are the results of curved mirror radius of curvature characterization. 

 

 

 

Attachment 1: Screenshot_2023-05-15_at_14.24.02.png
Screenshot_2023-05-15_at_14.24.02.png
ELOG V3.1.3-