40m QIL Cryo_Lab CTN SUS_Lab CAML OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  OMC elog, Page 13 of 13  Not logged in ELOG logo
ID Date Author Type Categoryup Subject
  606   Mon Aug 14 14:13:34 2023 Camille MakaremOpticsGeneralmode-mismatch in the cavity

For Set 1 of the data in https://nodus.ligo.caltech.edu:8081/OMC_Lab/604,

VREFL(unlocked) = 3.226V
VREFL(locked) = 0.13V
Pin = 20.56mW

Fraction of light that is reflected (mode-mismatched) = 0.135/3.226 = 4.03%
Pjunk = 0.0403*20.56mW = 0.83mW

From T1500060 Section 3.3, "The incident beam power to the cavity (Pin) can be split into the mode-matched (coupled) and mode-mismatched (junk) light power (Pcoupled and Pjunk, respectively)."
Pin­ = Pcoupled + Pjunk
20.56mW = Pcoupled + 0.83mW
Pcoupled = 19.73mW

This suggests that our cavity has nearly no loss, and the mode-matching efficiency is ~96%

 

However, this mode-matching efficiency is very different from the mode-matching efficiency determined from our transmitted PD signal on https://nodus.ligo.caltech.edu:8081/OMC_Lab/603.

From the PDtrans signal, the TEM00 signal is ~7.0V
There was only one higher-order mode observed with a signal of 0.070V.

0.070/7.0 = 1% mode-mismatch

  607   Tue Aug 15 16:15:13 2023 Camille MakaremOpticsGeneralnew power measurements (after cleaning)

[Camille, Thejas, Masayuki]

This afternoon we finished the realignment that we started after the FC cleaning in https://nodus.ligo.caltech.edu:8081/OMC_Lab/605.
We wanted to try to improve mode-matching before taking new power measurements. We used the signal from the transmission PD to characterize the mode-matching. We observed the TEM00 peak and one additional HOM peak:
TEM00 signal: 6.9V
HOM signal: 0.095V
--> mode-matching efficiency is ~1.4%

We observed the REFL CCD and include an attached picture. We recorded pictures of the beam spots using the CCD video camera (pictures attached).

We took one set of power budget measurements (measured values and outputs are shown in the attached screenshot).
The fraction of light that is reflected is
0.012V/3.35V = 3.6%
This is very similar to our previous data.
Similarly, our reflected power, incident power, and transmittd power are very similar to our previous values (Prefl=0.79mW, Pin=21.82mW, and POMCT=20.73mW)
This would seemingly indicate that we have very little loss in the cavity, however we still plan to further investigate the 3.5% loss observed by the REFL signal.

Attachment 1: REFL_CCD.PNG
REFL_CCD.PNG
Attachment 2: PowerBudgetAnalysis.PNG
PowerBudgetAnalysis.PNG
Attachment 3: ipad_notes.PNG
ipad_notes.PNG
Attachment 4: SpotPositions.PNG
SpotPositions.PNG
  608   Thu Aug 17 07:57:12 2023 Camille MakaremOpticsGeneralmode-matching in the cavity

[Camille, Thejas, Koji]
16 August 2023

We met in the lab to try to understand the mode-match discrepancy we see in our measurments. Adjusted the fiber coupler and the periscope to minimize the REFL PD signal. (REFL PD signal was 0.116 when locked.) The shape of the beam on the REFL CCD looked the same as in https://nodus.ligo.caltech.edu:8081/OMC_Lab/607.
We observed the transmission spectrum on the scope to identify higher order modes and side bands (need to attach plot). We closely examined the signal intensity of the weaker peaks in addition to the stong TEM00 peaks and exported the data from the scope. We also locked the cavity on the other modes to observe the shaped of these other modes (we see some pitch and yaw misalignment in the other modes).
The intensity signals of the other modes estimates ~1.8% mode-mismatch. (Still does not explain the 2% discrepancy between we mode-mismatch we calculate in our power budget analysis.)

We also varied amplitude of the phase modulation (from ~8-17dB) but this showed no improvement to the REFL PD signal.

Our plans moving forward:
-Center the beam path through the lenses to try to improve the mode-matching
-Further reduce REFL PD signal (~70mv?)
-Quick check: Attenuate the TRANS PD signal and compare ration between TEM00 signal and other modes.

  609   Thu Aug 17 09:06:51 2023 Camille MakaremOpticsGeneralrealignment of beam path through mode-matching lenses

[Camille, Thejas]
16 August 2023

We reconvened in the afternoon to begin realignment of the beam path through the mode-matching lenses. Before doing so, we placed two iris to mark our current beam path to the OMC. (one iris after the steering mirror, one iris right in front of the OMC (picture attached))
We made a few slights adjustments to the fiber coupler and the lenses: We used a level to adjust the height of the second lens so that it is at the same height as the first lens. We slightly adjusted the height of the fiber coupler mount so that the fiber height matches the height of the center of the lenses. We translated the fiber coupler slighly to adjust the centering while maintaining the same distance from the first lens.
After centering the path through the lenses, we repositioned the periscope mount and the steering mirror accordingly so that the beam path hits the centers of these mirrors.

Tomorrow, we will lock the cavity and repeat power measurements to determine if there is any improvement to the mode-matching.

Attachment 1: irises.PNG
irises.PNG
  610   Mon Aug 21 07:59:40 2023 Camille MakaremOpticsGeneralcavity realignment and new power measurements

[Camille, Thejas]
18 August 2023

We continued the work started on https://nodus.ligo.caltech.edu:8081/OMC_Lab/609. The beam is well-centered through the mode-matching lenses. We used the periscope to optimize cavity alignment while locked on the TEM00 mode.

We checked the steering on the REFL PD and the TRANS PD to make sure both are aligned.
The REFL PD signal was 3.0V (unlocked) and 46mV (locked). (This is the lowest REFL power we have had with this cavity.) A picture of the REFL CCD is attached.

We also checked the intensity of the HOMs on the mode spectrum (pictures attached). The TEM00 signal was ~7.2V while the observed HOMs had a signal of 23mV and 2 mV.

We proceeded to take 2 sets of power budget measurements (measurements and screenshot attached). After running the measurements in the power analysis script, we have and OMC throughput of ~99% and mode-matching efficiency of 98%. This seems to agree better with our mode-spectrum. (The excel spreadsheet with the analysis is attached as Attachment 5.)

Attachment 1: IMG_0353.jpg
IMG_0353.jpg
Attachment 2: IMG_0355.jpg
IMG_0355.jpg
Attachment 3: set1.PNG
set1.PNG
Attachment 4: set2.PNG
set2.PNG
Attachment 5: OMC_Powerbudget.xlsx
Attachment 6: powerbudget.PNG
powerbudget.PNG
  611   Mon Aug 21 15:02:32 2023 KojiOpticsGeneralcavity realignment and new power measurements

This is WOW!

Excellent mode matching work.

The measurement is still consistent with the low loss even with the different mode-matching level.

99.5%? The IFO commissioners will cry.


Edit: Wait a sec. The incident * mode matching = 20.14mW. This is the cavity-coupled power.
And you have the transmission of 9.78*2=19.56mW.
The ratio of these is ~97% and not 99.5%. Did I miss something?

=> Ah, understood. You have the incident power measurement with a significantly different reference voltage from the one at the transmission measurement. (4.21V vs 4.11V)
This is because the laser output power depends on the laser PZT feedback.
The quick hack to reduce this is that check the laser PZT feedback voltage (on the Thorlabs driver) right before the unlock, and bring the "output offset" close to that value after unlocking.
This brings the laser frequency back to the one during lock. At the same time, the laser freq is now close to the cavity resonance. So reading the unlocked REFL voltage, you need a bit of care.

  615   Tue Aug 22 15:51:15 2023 KojiGeneralGeneralInventory: C30655 PDs

During the OMC(004) assembly, the stock situation for Excelitas C30655 PDs was checked.

PD Cage H
Slot 1: Laser Components 3mm ->
Slot 2: Laser Components 3mm -> one of them is broken
Slot 3: C30655
Slot 4: Empty

PD Cage I
Slot 1: C30655 PD window unopened - Sourced from 40m
Slot 2: C30655 PD window unopened - Sourced from 40m
Slot 3: C30655 PD window unopened - Sourced from 40m
Slot 4: C30655 PD window opened - collected from photon-recycling experiment / state unknown

OMC DCPD bag
C30665 #4/#5/#6/#8/#13

- Some of the PDs were still in OMC (004) -> will be used for 40m BHR
- More PDs will be used for 40m BHR

  618   Thu Aug 24 16:09:46 2023 Camille MakaremOpticsGeneralPZT length-to-angle coupling

[Camille, Thejas]

We checked the length-to-angle coupling of each PZT by monitoring the position of the transmitted beam on the CCD camera. The CCD camera was placed behind the steering mirror that guides the transmitted beam to the PD. We used a ThorLabs piezo controller to actuate the PZT.

We first tested PZT2. We increased the voltage to PZT2 in 50V increments from 0V to 150V. We did not observe any change in the position of the transmitted beam. We monitored the signal of the TRANS PD on the scope and did not see any change. (The signal was between 191-195V.) We monitored the REFL CCD and did see changes in the beamshape, which was expected (see pictures). The REFL PD signal also increased slightly with PZT actuation (see attachment).

We repeated this process for PZT1, which showed similar results (see attachment). We did not observe any movement in the position of the transmitted beam. Increasing PZT voltage shows increasing pitch misalignment in the REFL CCD and increasing REFL PD signal.

Attachment 1: PZT2.jpg
PZT2.jpg
Attachment 2: PZT1.jpg
PZT1.jpg
  619   Fri Aug 25 22:01:29 2023 KojiFacilityGeneralLED Strip installation plan

Here are the dimensions of the LED strips and their gaps.

Attachment 1: LED_Strip.pdf
LED_Strip.pdf
  620   Tue Aug 29 16:54:46 2023 KojiFacilityGeneralLED Strip installed

[JC, Koji]

We've installed the LED strips on the HEPA frame. We tried not to touch the OMC there. But please check if everything is still ok.

Attachment 1: Installed LED light. Notice the room light is off. At the max brightness, it's still sufficient to work with the room light off.

Attachment 2: The strips are connected at the south side of the HEPA booth. LEDs are attached to the frame with the default double-sided tape. We can improve how the wire is fixed on the frame by more tapes.

Attachment 3: The switch is close to the TOPGUN unit. The single click does turn on/off, and the long touch makes the brightness go up and down. At the max and min brightness, it blinks.

Attachment 1: PXL_20230829_232310007.jpg
PXL_20230829_232310007.jpg
Attachment 2: PXL_20230829_232345702.MP.jpg
PXL_20230829_232345702.MP.jpg
Attachment 3: PXL_20230829_232317162.jpg
PXL_20230829_232317162.jpg
  621   Wed Aug 30 12:38:52 2023 Thejas OpticsGeneralA+ OMC #1: HOM spectrum

Examination into bonding template confimred the limitation of space available to change cavity length by at least 10 mm to improve the cavity HOM spectrum. Here's an anlysis of HOM spectrum for various possible ROCs and corresponding required cavity length change for optimum HOM spectrum. 

Assume: Astigmatism: Rx-Ry = 8mm

Current cavity length: 1.132 m

Ry = measured, Desired Cavity length: <1.12476 m

Ry = 2.5 m, Desired cav length: >1.13876 m

Ry = 2.55 m, “: No change 

Ry = 2.6 m, “:<1.12

So travel range of ~ (1.138 - 1.12) / 4 = 5 mm on each CM is required. WIth a safety factor for alignemnt say we require 30 mm/4 ~ 7 mm on each curved mirror.

  623   Mon Sep 18 15:08:26 2023 Thejas OpticsGeneralOMC HEPA enclosure filter fan speed

I went in lab today and turned the HEPA filter (clsoe to the entrance) to high since we are not doing any measurements at the moment. 

 

 

  248   Fri Dec 18 15:33:24 2015 KojiGeneralLoan / LendingLoan from Rich

Loan Record: I borrowed a PD can opener from Rich => Antonio Returned Sep 9, 2016

Tungsten Carbide Engraver (permanently given to the OMC lab)

KEITHLEY SOURCE METER + Laptop

  250   Thu Feb 18 21:08:32 2016 KojiGeneralLoan / Lending(all returned) Antonio loan

Antonio borrowed: Rich's PD cutter (returned), Ohir power meter(returned), Thorlabs power meter head, Chopper

  258   Tue Apr 5 18:14:55 2016 KojiGeneralLoan / LendingQPD Lending Crackle

Xiaoyue

QPD head
X-Z stage
Mounting brackets
DB15 cable
QPD matrix circuit
+/-18V power supply cable

  293   Thu May 3 21:45:58 2018 awadeGeneralLoan / LendingBorrowed toaster oven

I’ve borrowed the black and decker toaster oven to dry some sonicated parts. It is temporarly located in the QIL lab. 

Attachment 1: 9CE80545-7A58-4236-B7E3-1EE6C4042DAA.jpeg
9CE80545-7A58-4236-B7E3-1EE6C4042DAA.jpeg
  385   Tue Oct 22 15:54:59 2019 KojiElectronicsLoan / LendingBorrowed LB1005 from Cryo Cav

From Cryo Cav setup

Borrowed LB1005 Servo box -> OMC

 

  391   Mon Aug 10 15:34:04 2020 KojiFacilityLoan / LendingGlue bake oven

Black and Decker Glue Baking Oven came back to the OMC lab on Aug 10, 2020, Georgia had lent the unit for the SAMS assembly/testing.

  518   Tue Apr 4 16:14:06 2023 KojiOpticsLoan / LendingC7 mirror to Downs

I handed Camille the C7 mirror for the cross-calibration of the ROC characterization techniques.

  529   Wed Apr 26 14:43:44 2023 StephenGeneralLoan / LendingBorrowed Items for PZT DC Response Shadow Sensor Setup

Borrowed for PZT DC Response Shadow Sensor Setup (see Attachment 1):

  • Thorlabs PDA100A Photodiode (and power supply)
  • Thorlabs MDT694B Piezo Driver

Current Location: Downs 227

Attachment 1: PXL_20230426_185227710.jpg
PXL_20230426_185227710.jpg
  538   Fri May 12 10:00:06 2023 Camille MakaremGeneralLoan / LendingBorrowed Items for PZT DC Response Shadow Sensor Setup

The ThorLabs MDT694B piezo driver was returned to the OMC lab.

Quote:

Borrowed for PZT DC Response Shadow Sensor Setup (see Attachment 1):

  • Thorlabs PDA100A Photodiode (and power supply)
  • Thorlabs MDT694B Piezo Driver

Current Location: Downs 227

 

Attachment 1: PXL_20230511_201545235.jpg
PXL_20230511_201545235.jpg
ELOG V3.1.3-