ID |
Date |
Author |
Type |
Category |
Subject |
454
|
Mon Nov 14 08:34:45 2022 |
Camille | Optics | Characterization | transmission measurements through OMC #1 (before cleaning) |
[Camille, Koji]
Friday, Nov 11th, 2022
Setting up OMC #1 for transmission measurements:
The laser beam was aligned to the OMC cavity. The OMC cavity was locked and the transmission measurements were recorded. |
455
|
Mon Nov 14 09:27:13 2022 |
Koji | Optics | Characterization | transmission measurements through OMC #1 (before cleaning) |
The measured total optical loss of the OMC was
1st: 0.015 +/- 0.003
2nd: 0.085 +/- 0.005
3rd: 0.0585+/- 0.0008
4th: 0.047 +/- 0.002
In avegrage the estimated loss is
Loss = 0.055 +/- 0.014
This is unchanged from the measurement at LLO after the FC cleaning
Loss = 0.053 +/- 0.010 |
527
|
Mon Apr 24 15:29:48 2023 |
Camille Makarem | Optics | | summary of zygo setups |
Summary of Zygo setups
Initial Zygo Setup:
Our initial Zygo setup consisted of a flat transmission sphere with the 0.5" curved mirror mounted against a 1" flat mirror.
Mounting procedure:
The bottom part of the gluing fixture was attached to a mounting plate using two screws. The 1" reference flat was placed on the gluing fixture. The reference flat was inspected with a green flashlight to ensure that there was no dust on the mirror surface. Any dust was removed using top gun. If any dust remained after using top gun, it was removed with a swab.
The back surface of the curved mirror was inspected and cleaned using the same method (flashlight inspection, followed by top gun if necessary, followed by swab if necessary).
After ensuring that both surfaces are clean, the back surface of the curved mirror was placed on the front surface of the reference flat. The fiducial of the curved mirror was positioned at 12:00. (12:00 is defined as the top of assembly.) The two mirrors were held in place using a mounting plate with a 0.4" aperture. The mounting plate was fixed to the bottom part of the gluing fixture using two screws and a spring for each screw (see attached picture).
The mounting plate holding this assembly was then attached to a optical mount with tip/tilt adjustments (see attached picture).
This assembly was placed facing the Zygo transmission flat (see attached picture) and the mount was pitched/yawed until the fringes on the 1" reference flat were nulled. After nulling the fringes, the data was then recorded.
The mounting plate was then removed from the tip/tilt mount and dissassembled so that the curved mirror could be rotated so that the fiducial is in the 3:00 position. The procedure is then repeated and the data recorded.
This was repeated again with the fiducial in the 6:00, 9:00 and 12:00 (again) positions.
Review of this data shows that the positions of the curvature minimums was not reproducible with sufficient precision. A teflon mounting plate was added to clamp the 1" reference flat more securely to the gluing fixture (See attached pictures). Data was collected in the same manner (twice with the fiducial at 12:00 and once with fiducial at each of the positions 3:00, 6:00, and 9:00).
Additional data collected still failed to produce reproducible results and the removing/remounting process of the curved mirror was time-consuming, so we attempted a new setup for the Zygo measurments.
Final Zygo Setup:
The new setup used a fold mirror mounted at 45degrees to direct the Zygo beam downwards into the plane of the table. A 3" flat was used as our reference flat. The reference flat was placed on some lens tissue parallel to the plane of the table. The same inspection and cleaning method was used to ensure that there was no dust on the reference flat (flashlight inspection, followed by top gun if necessary, followed by swabbing if necessary).
The back of the curved optic was inspected and cleaned using the same method. The curved mirror was placed on the 3" reference flat with the fiducial at the 12:00 position. (12:00 here is defined as the direction ponting towards the Zygo instrument.) (See attached picture of this setup.)
The fold mirror was pitched/yawed so that the fringes on the 3" reference flat were nulled. (An additional advantage of this setup is that more surface of the reference flat was viewable.) After nulling the fringes, the curved mirror was picked up and replaced a few times to verify that the fringe pattern on the curved mirror appeared reproducible. The data was collected with the fiducial at the 12:00 position. This process was repeated with the fiducial at 3:00, 6:00, 9:00, and again at 12:00.
Results from this setup were reproducible and this setup was used to measure the surface profile of all the curved mirrors. |
598
|
Tue Aug 8 07:39:17 2023 |
Camille Makarem | Optics | General | steering the beam spot on the curved mirrors |
[Camille, Thejas]
7 August 2023
After optimizing the alignment, we used the CCD video cameras to observe the beam spots on CM1 and CM2 (pictures attached). For both mirrors the beam spot is off center in the direction away from FM1 and FM2.
In order to steer the beam axis towards FM1 and FM2, we rotated CM1 clockwise (as viewed from above) and we rotated CM2 counter-clockwise (as viewed from above).
We aimed to steer the beam axis by 1mm, so we displaced the micrometers on both curved mirrors by 10um.
After displacing the micrometers, we attempted to recover TEM00 using only the fiber coupler and the steering mirrors, but we were unable to achieve this. We made an additional adjustment to the CM2 micrometer so that the beam reflected from CM2 would better overlap the incoming beam on FM1. After doing this, we were able to observe TEM00 and proceeded to lock and optimize using the fiber coupler and steering mirrors. (REFL PD was 0.12V, REFL CCD looked similar to the one in https://nodus.ligo.caltech.edu:8081/OMC_Lab/597.
We observed the beam spots on CM1 and CM2 again and didn't observe any noticeable change (pictures attached).
We repeated this process. (We rotated CM1 clockwise and CM2 counter-clockwise each by the same amount.) We tried to recover TEM00 using only the steering mirrors and fiber coupler, but were unable to find TEM00. We made an additional small adjustment to the CM2 micrometer and found TEM00. After locking and optimizing, the beam spots on the curved mirrors still appear to be in the same locations. |
600
|
Tue Aug 8 10:42:11 2023 |
Thejas | Optics | General | steering the beam spot on the curved mirrors |
The spots look quite bright on the CMs, we need to move the spots towards the center of the mirorrs to avoid scatter points.
Quote: |
[Camille, Thejas]
7 August 2023
After optimizing the alignment, we used the CCD video cameras to observe the beam spots on CM1 and CM2 (pictures attached). For both mirrors the beam spot is off center in the direction away from FM1 and FM2.
In order to steer the beam axis towards FM1 and FM2, we rotated CM1 clockwise (as viewed from above) and we rotated CM2 counter-clockwise (as viewed from above).
We aimed to steer the beam axis by 1mm, so we displaced the micrometers on both curved mirrors by 10um.
After displacing the micrometers, we attempted to recover TEM00 using only the fiber coupler and the steering mirrors, but we were unable to achieve this. We made an additional adjustment to the CM2 micrometer so that the beam reflected from CM2 would better overlap the incoming beam on FM1. After doing this, we were able to observe TEM00 and proceeded to lock and optimize using the fiber coupler and steering mirrors. (REFL PD was 0.12V, REFL CCD looked similar to the one in https://nodus.ligo.caltech.edu:8081/OMC_Lab/597.
We observed the beam spots on CM1 and CM2 again and didn't observe any noticeable change (pictures attached).
We repeated this process. (We rotated CM1 clockwise and CM2 counter-clockwise each by the same amount.) We tried to recover TEM00 using only the steering mirrors and fiber coupler, but were unable to find TEM00. We made an additional small adjustment to the CM2 micrometer and found TEM00. After locking and optimizing, the beam spots on the curved mirrors still appear to be in the same locations.
|
|
601
|
Wed Aug 9 07:47:21 2023 |
Camille, Thejas | Optics | General | steering the beam spot on the curved mirrors |
[Camille, Thejas, Masayuki]
8 August 2023
We continued our efforts to steer the beam spots on CM1 and CM2 towards FM1 and FM2. We rotated CM1 clockwise by displacing the micrometer 20um (2 small divisions). We rotated CM2 counter-clockwise by the same amount by displacing the micrometer 20um.
We then walked the beam with the periscope until we recovered the TEM00 mode. Once recovered, we locked the cavity and continued to optimize using the periscope and the fiber coupler output. We checked the beam spot positions with an IR card and viewer and recorded images with the CCD video camera (images attached).
We also monitored the transmission PD on the scope (green trace in attached picture, pink trace is the laser freq sweep) to compare the voltage signal from TEM00 (4.3V) and the voltage signal from the higher-order mode (76mV). (1.7% from mode-mismatch = 76 mV *100 / 4.3 V + 0.76 mV) This means we should be able to resonate the cavity at 0.017*3.4 V ~ 70 mV of refl power with the current level of mode-mathcing
Our refl power (refl.jpeg) shows significant relfection of about 400 mV, suggests the cavity is lossy. (See scatter on CM1). A separate elog will look at the scatter plot.
For this afternoon, we will plan to clean the optic surfaces with First Contact and see if this reduces the cavity loss. |
583
|
Tue Aug 1 11:04:13 2023 |
Camille Makarem | General | General | reoptimizing cavity alignment before locking |
[Camille, Thejas]
We rechecked the alignment this morning and re-optimized the resonance a bit. There was some horizontal drift in the alignment from yesterday.
Plan for this afternoon:
Camille, Thejas, and Masayuki will meet with Koji in the OMC lab to go over electronics setup. (Need to set up EOM driver? Laser is not sweeping.) |
609
|
Thu Aug 17 09:06:51 2023 |
Camille Makarem | Optics | General | realignment of beam path through mode-matching lenses |
[Camille, Thejas]
16 August 2023
We reconvened in the afternoon to begin realignment of the beam path through the mode-matching lenses. Before doing so, we placed two iris to mark our current beam path to the OMC. (one iris after the steering mirror, one iris right in front of the OMC (picture attached))
We made a few slights adjustments to the fiber coupler and the lenses: We used a level to adjust the height of the second lens so that it is at the same height as the first lens. We slightly adjusted the height of the fiber coupler mount so that the fiber height matches the height of the center of the lenses. We translated the fiber coupler slighly to adjust the centering while maintaining the same distance from the first lens.
After centering the path through the lenses, we repositioned the periscope mount and the steering mirror accordingly so that the beam path hits the centers of these mirrors.
Tomorrow, we will lock the cavity and repeat power measurements to determine if there is any improvement to the mode-matching. |
604
|
Fri Aug 11 10:42:05 2023 |
Camille Makarem | Optics | Characterization | power budget measurements |
[Camille, Thejas]
10 August 2023
We want to evaluate the loss in the cavity, so we recorded the power measurements needed to enter into the power budget analysis (see attached picture with recorded values). We collected three sets of measurements to be averaged.
The screenshot below shows the output from the python analysis. (OMC throughput values are surprisingly high: 100.0%, 103.3%, and 99.8%) We plan to try to improve the mode-matching and re-evaluate the power throughput. |
582
|
Mon Jul 31 13:19:14 2023 |
Camille Makarem | General | General | plans for this week 31 July 2023 |
Plans for this week:
1) Verify electronics setup for cavity locking.
Current state: Output from function generator is going directly to laser driver (not currently using the Newport Servo Module). We will need to set up the servo module.
2) Finish setting up the table optics so we can monitor the reflected beam (need to set up photodiode and CCD camera to monitor this beam).
3) Once the cavity is locked, we can use the steering mirrors to optimize the cavity while monitoring the CCD cameras for the transmitted and reflected beams. We need to walk the beam so that more of the TEM00 mode is transmitted. (Currently still seeing a lot of higher order modes transmitted, as shown in attachment). |
585
|
Thu Aug 3 07:22:19 2023 |
Camille Makarem | General | General | particle count and cavity optimization/alignment |
Yesterday, we measured the particle count in the enclosure to ensure that the lower setting on one of the HEPAs is still acceptable. The particle count is still zero for all measured particle sizes (0.3um, 0.5um, 1.0um, 2.0um, and 5.0um).
We also relocked the cavity and repeated the optimization efforts in https://nodus.ligo.caltech.edu:8081/OMC_Lab/584. The reflected signal was around 110mV (compared to 80mV on Tuesday).
We used a CCD camera to view the beam spots on the curved mirrors (pictures attached). We will compare the spot positions to the scatter plots for these mirrors and try to steer the beam spots accordingly.
(*Note: We did see some stray light scattering on the edge of FM2. We will further examine this and see where the stray light is coming from.)
|
591
|
Mon Aug 7 07:36:09 2023 |
Camille Makarem | General | General | particle count |
[Camille, Thejas]
We lowered the setting to MEDIUM on the HEPA furthest from the entrance. (The two HEPA settings are now LOW and MEDIUM.) We measured the particle count for all particle sizes at different locations in the enclosure (next to the OMC fixture and at the edge of the table closest to the entrance). All counts read zero. |
603
|
Thu Aug 10 12:33:42 2023 |
Camille Makarem | Optics | General | optimizing the cavity post-cleaning |
[Camille, Thejas]
This morning, we continued the efforts from http://nodus.ligo.caltech.edu:8080/OMC_Lab/602.
While monitoring the REFL PD, we walked/optimized the alignment after cleaning the optics with FC. After optimization, our PD signals were:
REFL PD: 0.11V (locked)
REFL PD: 3.3V (unlocked)
We also have a picture of the REFL CCD (attached). The REFL CCD does not appear to show so much TEM00 as it did previously.
We also checked the mode scan on the transmission PD and recorded the signals of the TEM00 mode and another higher-order mode that was still observed.
TEM00: 7.0V
HOM: 0.070V
--> ~1% mode mismatch.
From this we estimate that we should be able to reduce the REFL PD signal to ~30mV.
We used the CCD video cameras to record pictures of the beam spots on the cavity optics (attached). The scatter looks much better post-cleaning and the beam spots are closer to center. However, we can still steer the beam closer to center (need to move away from FM1 and FM2). |
592
|
Mon Aug 7 07:40:33 2023 |
Camille Makarem | Optics | General | optimizing reflected signal |
[Camille, Thejas] 4 August 2023
-REFL PD signal (unlocked): 3.5V
We optimized the alignment of the steering mirror on the REFL PD.
-REFL PD signal (locked): 1.5V (initially) --> 0.15V after we optimized alignment into the cavity.
-We finished setting up the table optics so we monitor the transmitted beam with a photodiode. (Previously, we were only monitoring the transmitted beam with a CCD camera.) A mirror is used to steer the beam to the PD.
-We checked the beam spot positions on the mode-matching lenses and adjusted the fiber mount. (It is centered on the first lens. It is somewhat off-center on the second lens [attach picture].)
-We optimized the alignment to reduce the REFL PD signal when locked. (We slightly pitched the periscope mirror, adjusted the fiber mount, and slightly walked the beam with the periscope. Lowest achievable REFL PD signal was 0.15V
Our mode-matching ratio is ~95.8%
-We unlocked the laser and used the frequency offset to sweep and observe shape of the transmitted modes on the CCD camera. TEM00 modes are the most prominent. The other observed modes appear to be 1st/2nd/3rd order rectangular modes. [attach pictures of modes]
Plans for this week:
-If mode-matching ratio is acceptable, we will proceed to move the cavity axis between CM1 and CM2. (The axis needs to move towards FM1 and FM2.)
-If mode-matching ratio needs to be improved, we will re-characterize the beam waist position and re-verify the location of our beam waist. (We will temporarily place a mirror or beamsplitter after the mode-matching breadboard, so we can steer the beam away from the OMC optics and use the beam profiler without disrupting the OMC optics.) |
593
|
Mon Aug 7 08:37:59 2023 |
Koji | Optics | General | optimizing reflected signal |
With the current level of the mode matching, the spot positions can be optimized. I'd work on it first as the mirror RoC and loss depends on the spot positions on the curved mirrors.
----
I'd also look at the reflection CCD image:
With the previous OMC, we saw the reflection to be 60~70mV. We saw ~85mV with the previous alignment when I went down to the lab.
Today it is ~150mV.
- If this increase is coming from the worse reflection of the cavity (lossy cavity somehow), the reflection CCD image should definitely show TEM00 mode.
-> The cavity needs to be cleaned
- If this is coming from the mode matching, the image seen with the room lights off should show (somewhat symmetric) higher-order modes.
-> I'd try moving the lenses.
|
594
|
Mon Aug 7 08:51:56 2023 |
Camille Makarem | Optics | General | optimizing reflected signal |
We will plan to look at the reflection CCD image this afternoon.
Quote: |
With the current level of the mode matching, the spot positions can be optimized. I'd work on it first as the mirror RoC and loss depends on the spot positions on the curved mirrors.
----
I'd also look at the reflection CCD image:
With the previous OMC, we saw the reflection to be 60~70mV. We saw ~85mV with the previous alignment when I went down to the lab.
Today it is ~150mV.
- If this increase is coming from the worse reflection of the cavity (lossy cavity somehow), the reflection CCD image should definitely show TEM00 mode.
-> The cavity needs to be cleaned
- If this is coming from the mode matching, the image seen with the room lights off should show (somewhat symmetric) higher-order modes.
-> I'd try moving the lenses.
|
|
607
|
Tue Aug 15 16:15:13 2023 |
Camille Makarem | Optics | General | new power measurements (after cleaning) |
[Camille, Thejas, Masayuki]
This afternoon we finished the realignment that we started after the FC cleaning in https://nodus.ligo.caltech.edu:8081/OMC_Lab/605.
We wanted to try to improve mode-matching before taking new power measurements. We used the signal from the transmission PD to characterize the mode-matching. We observed the TEM00 peak and one additional HOM peak:
TEM00 signal: 6.9V
HOM signal: 0.095V
--> mode-matching efficiency is ~1.4%
We observed the REFL CCD and include an attached picture. We recorded pictures of the beam spots using the CCD video camera (pictures attached).
We took one set of power budget measurements (measured values and outputs are shown in the attached screenshot).
The fraction of light that is reflected is
0.012V/3.35V = 3.6%
This is very similar to our previous data.
Similarly, our reflected power, incident power, and transmittd power are very similar to our previous values (Prefl=0.79mW, Pin=21.82mW, and POMCT=20.73mW)
This would seemingly indicate that we have very little loss in the cavity, however we still plan to further investigate the 3.5% loss observed by the REFL signal. |
606
|
Mon Aug 14 14:13:34 2023 |
Camille Makarem | Optics | General | mode-mismatch in the cavity |
For Set 1 of the data in https://nodus.ligo.caltech.edu:8081/OMC_Lab/604,
VREFL(unlocked) = 3.226V
VREFL(locked) = 0.13V
Pin = 20.56mW
Fraction of light that is reflected (mode-mismatched) = 0.135/3.226 = 4.03%
Pjunk = 0.0403*20.56mW = 0.83mW
From T1500060 Section 3.3, "The incident beam power to the cavity (Pin) can be split into the mode-matched (coupled) and mode-mismatched (junk) light power (Pcoupled and Pjunk, respectively)."
Pin = Pcoupled + Pjunk
20.56mW = Pcoupled + 0.83mW
Pcoupled = 19.73mW
This suggests that our cavity has nearly no loss, and the mode-matching efficiency is ~96%
However, this mode-matching efficiency is very different from the mode-matching efficiency determined from our transmitted PD signal on https://nodus.ligo.caltech.edu:8081/OMC_Lab/603.
From the PDtrans signal, the TEM00 signal is ~7.0V
There was only one higher-order mode observed with a signal of 0.070V.
0.070/7.0 = 1% mode-mismatch |
608
|
Thu Aug 17 07:57:12 2023 |
Camille Makarem | Optics | General | mode-matching in the cavity |
[Camille, Thejas, Koji]
16 August 2023
We met in the lab to try to understand the mode-match discrepancy we see in our measurments. Adjusted the fiber coupler and the periscope to minimize the REFL PD signal. (REFL PD signal was 0.116 when locked.) The shape of the beam on the REFL CCD looked the same as in https://nodus.ligo.caltech.edu:8081/OMC_Lab/607.
We observed the transmission spectrum on the scope to identify higher order modes and side bands (need to attach plot). We closely examined the signal intensity of the weaker peaks in addition to the stong TEM00 peaks and exported the data from the scope. We also locked the cavity on the other modes to observe the shaped of these other modes (we see some pitch and yaw misalignment in the other modes).
The intensity signals of the other modes estimates ~1.8% mode-mismatch. (Still does not explain the 2% discrepancy between we mode-mismatch we calculate in our power budget analysis.)
We also varied amplitude of the phase modulation (from ~8-17dB) but this showed no improvement to the REFL PD signal.
Our plans moving forward:
-Center the beam path through the lenses to try to improve the mode-matching
-Further reduce REFL PD signal (~70mv?)
-Quick check: Attenuate the TRANS PD signal and compare ration between TEM00 signal and other modes. |
362
|
Thu May 16 12:41:28 2019 |
Chub | General | General | fire pillow found on optics table |
That is an expanding fire pillow, also known as firebrick. It is used to create a fire block where holes in fire-rated walls are made and prevents lab fires from spreading rapidly to adjacent labs. I had to pull cable from B254 to our labs on either side during a rather narrow window of time. Some of the cable holes are partially blocked, making it difficult to reach the cable to them. The cable is then just guided to the hole from a distance. With no help, it's not possible to see this material getting shoved out of the hole. I can assure you that I took great pains not to allow the CYMAC coax to fall into any equipment, or drag against any other cables. |
107
|
Wed Apr 10 00:40:30 2013 |
Zach | Optics | Configuration | fauxMC locked |
[Koji, Zach]
Tonight, we locked the "fauxMC". We obtained a visibility of >99%.
Koji had aligned it roughly last night, but we wanted to have a couple steering mirrors in the path for this practice cavity (the periscope mirrors will serve this function in the real setup), so we marked the alignment with irises and installed two extra mirrors.
After obtaining flashes with the WinCam placed at the output coupler, we removed the WinCam and put a CCD camera at one of the curved mirror transmissions and used this to get a strong TEM00 flash. Then, we installed the REFL PD/CCD, swept the laser PZT and optimized the alignment by minimizing the REFL dips. Finally, we connected the RF electronics and locked the cavity with the LB box. We used whatever cables we had around to trim the RF phase, and then Koji made some nice SMA cables at the 40m.
One thing we noticed was that we don't have enough actuation range to keep the cavity locked for very long---even with the HV amp (100V). We are going to offload to the NPRO temperature using an SR560 or pomona box circuit. We may also make an enclosure for the cavity to protect it from the HEPA blasting.
Tomorrow, after we do the above things, we will practice measuring the transmission, length (FSR) and mode spectrum of the cavity before moving on to the real McCoy.

|
67
|
Tue Mar 5 19:37:00 2013 |
Zach | Optics | Characterization | eLIGO OMC visibility vs. power measurement details |
EDIT (ZK): Koji points out that (1 - Ti) should really be the non-resonant reflectivity of the aligned cavity, which is much closer to 1. However, it should *actually* be the non-resonant reflectivity of the entire OMC assembly, including the steering mirror (see bottom of post). The steering mirror has T ~ 0.3%, so the true results are somewhere between my numbers and those with (1 - Ti) -> 1. In practice, though, these effects are swamped by the other errors.
More information about the power-dependent visibility measurement:
As a blanket statement, this measurement was done by exact analogy to those made by Sam and Sheon during S6 (c.f. LHO iLog 11/7/2011 and technical note T1100562), since it was supposed to be a verification that this effect still remains. There are absolutely better ways to do (i.e., ways that should give lower measurement error), and these should be investigated for our characterization. Obviously, I volunteer.
All measurements were made by reading the output voltages produced by photodetectors at the REFL and TRANS ports. The REFL PD is a BBPD (DC output), and the TRANS is a PDA255. Both these PDs were calibrated using a Thorlabs power meter (Controller: PM100D; Head: S12XC series photodiode-based---not sure if X = 0,2... Si or Ge) at the lowest and highest power settings, and these results agreed to the few-percent level. This can be a major source of error.
The power was adjusted using the HWP/PBS combination towards the beginning of the experiment. For reference, an early layout of the test setup can be seen in LLO:5978 (though, as mentioned above, the REFL and TRANS PDs have been replaced since then---see LLO:5994). This may or may not be a "clean" way to change the power, but the analysis should take the effect of junk light into account.

Below is an explanation of the three traces in the plot. First:
- TRANS: TRANS signal calibrated to W
- REFL_UL: REFL signal while cavity is unlocked, calibrated to W
- REFL_L: REFL signal while cavity is locked, calibrated to W
- Psb: Sideband power (relative to carrier)
- Ti: Input mirror transmission (in power)
Now, the traces
- Raw transmission: This measurement is simple. It is just the raw throughput of the cavity, corrected for the power in the sidebands which should not get through. I had the "AM_REF" PD, which could serve as an input power monitor, but I thought it was better to just use REFL_UL as the input power monitor and not introduce the error of another PD. This means I must also correct for the reduction in the apparent input power as measured at the REFL PD due to the finite transmission of the input coupler. This was not reported by Sam and Sheon, but can be directly inferred from their data.
- trans_raw = TRANS ./ ( REFL_UL * (1 - Psb) * (1 - Ti) )
- Equivalently, trans_raw = (transmitted power) ./ (input power in carrier mode)
- Coupling: This is how much of the power incident on the cavity gets coupled into the cavity (whether it ends up in transmission or at a loss port). Sheon plots something like (1 - coupling) in his reply to the above-linked iLog post on 11/8/2011.
- coupling = ( REFL_UL * (1 - Ti) - REFL_L ) ./ ( REFL_UL * (1 - Psb) * (1 - Ti) )
- Equivalently, coupling = [ (total input power) - (total reflected power on resonance) ] ./ (input power in carrier mode)
- Visibility: How much of the light that is coupled into the cavity is emerging from the transmitted port? This is what Sam and Sheon call "throughput" or "transmission" and is what is reported in the majority of their plots.
- visibility = TRANS ./ ( REFL_UL * (1 - Ti) - REFL_L )
- Equivalently, visibility = (transmitted power) ./ [ (total input power) - (total reflected power on resonance) ]
- Also equivalently, visibility = trans_raw ./ coupling
The error bars in the measurement were dominated, roughly equally, by 1) systematic error from calibration of the PDs with the power meter, and 2) error from noise in the REFL_L measurement (since the absolute AC noise level in TRANS and REFL_L is the same, and TRANS >> REFL_L, the SNR of the latter is worse).
(1) can be helped by making ALL measurements with a single device. I recommend using something precise and portable like the power meter to make measurements at all the necessary ports. For REFL_L/UL, we can place a beam splitter before the REFL PD, and---after calibrating for the T of this splitter very well using the same power meter---both states can be measured at this port.
(2) can probably be helped by taking longer averaging, though at some point we run into the stability of the power setting itself. Something like 30-60s should be enough to remove the effects of the REFL_L noise, which is concentrated in the few-Hz region in the LLO setup.
One more thing I forgot was the finite transmission of the steering mirror at the OMC input (the transmission of this mirror goes to the QPDs). This will add a fixed error of 0.3%, and I will take it into account in the future. |
68
|
Wed Mar 6 23:24:58 2013 |
Zach | Optics | Characterization | eLIGO OMC visibility vs. power measurement details |
I found that, in fact, I had lowered the modulation depth since when I measured it to be 0.45 rads --> Psb = 0.1.
Here is the sweep measurement:

This is Psb = 0.06 --> gamma = 0.35 rads.
This changes the "raw transmission" and "coupling", but not the inferred visibility:

I also measured the cavity AMTF at three powers today: 0.5 mW, 10 mW, and 45 mW input.

They look about the same. If anything, the cavity pole seems slightly lower with the higher power, which is counterintuitive. The expected shift is very small (~10%), since the decay rate is still totally dominated by the mirror transmissions even for the supposed high-loss state (Sam and Sheon estimated the roundtrip loss at high power to be ~1400 ppm, while the combined coupling mirrors' T is 1.6%). I have not been able to fit the cavity poles consistently to within this kind of error. |
602
|
Wed Aug 9 17:00:52 2023 |
Camille Makarem | Optics | General | cleaning the cavity optics |
[Camille, Thejas]
We inspected surfaces of the cavity optics with a halogen lamp and noticed a particulates on optic HR surfaces. We cleaned all 4 optics with top gun, this did not seem to reduce the cavity loss upon locking. The cavity loss seemed to be higher with refl power now at 600 mV.
So we proceeded to apply First Contact on all the four optics.
The bottom surfaces of the prisms and tombstones were swabbed with IPA. We also cleaned the breadboard positions for these four optics with an IPA swab. (The CM1 mirror looked better after cleaning (no scatter observed).)
After cleaning the optics were returned to their positions on the breadboard. The spot positions on the curved mirrors still look roughly centered, but the steering mirrors need to be optimized.
When viewed through the IR viewer, the spots on the curved optics are not as bright as they were before FC. (The brightness is more or less equal to the brightness on the FMs.)
We observed the transmission PD which was mainly dominated with TEM00 modes, and also some higher-order modes:
TEM00: 6.85V
HOMs: 0.174V and 0.054V
--> 3.2% mode-mismatch
REFL PD: 3.6V (unlocked)
REFL PD: 0.340V (locked) (We estimate we should be able to reach 0.115V for the same level of mode-matching.) (The refl mode shape looks similar to last time on the CCD (some TEM00 mode).)
This implies there is still ~200 mV of loss from the cavity. |
456
|
Tue Nov 15 07:46:58 2022 |
Camille Makarem | Optics | General | cleaning OMC #1 |
Monday, November 14, 2022
Camille and Koji did a "deep cleaning" of OMC#1:
1) Applied First Contact to the mirror surfaces. Removed first contact after ~10 minutes.
2) Acetone scrub of the mirror surfaces with a cotton swab.
3) Applied First Contact again. Removed after ~10 minutes. We left the FC paint on for the work on Thu.
The foggy spot on the input mirror was unchanged after the first round of First Contact. But the foggy spot came off during the acetone scrub. |
610
|
Mon Aug 21 07:59:40 2023 |
Camille Makarem | Optics | General | cavity realignment and new power measurements |
[Camille, Thejas]
18 August 2023
We continued the work started on https://nodus.ligo.caltech.edu:8081/OMC_Lab/609. The beam is well-centered through the mode-matching lenses. We used the periscope to optimize cavity alignment while locked on the TEM00 mode.
We checked the steering on the REFL PD and the TRANS PD to make sure both are aligned.
The REFL PD signal was 3.0V (unlocked) and 46mV (locked). (This is the lowest REFL power we have had with this cavity.) A picture of the REFL CCD is attached.
We also checked the intensity of the HOMs on the mode spectrum (pictures attached). The TEM00 signal was ~7.2V while the observed HOMs had a signal of 23mV and 2 mV.
We proceeded to take 2 sets of power budget measurements (measurements and screenshot attached). After running the measurements in the power analysis script, we have and OMC throughput of ~99% and mode-matching efficiency of 98%. This seems to agree better with our mode-spectrum. (The excel spreadsheet with the analysis is attached as Attachment 5.) |
611
|
Mon Aug 21 15:02:32 2023 |
Koji | Optics | General | cavity realignment and new power measurements |
This is WOW!
Excellent mode matching work.
The measurement is still consistent with the low loss even with the different mode-matching level.
99.5%? The IFO commissioners will cry.
Edit: Wait a sec. The incident * mode matching = 20.14mW. This is the cavity-coupled power.
And you have the transmission of 9.78*2=19.56mW.
The ratio of these is ~97% and not 99.5%. Did I miss something?
=> Ah, understood. You have the incident power measurement with a significantly different reference voltage from the one at the transmission measurement. (4.21V vs 4.11V)
This is because the laser output power depends on the laser PZT feedback.
The quick hack to reduce this is that check the laser PZT feedback voltage (on the Thorlabs driver) right before the unlock, and bring the "output offset" close to that value after unlocking.
This brings the laser frequency back to the one during lock. At the same time, the laser freq is now close to the cavity resonance. So reading the unlocked REFL voltage, you need a bit of care. |
612
|
Mon Aug 21 16:05:07 2023 |
Camille Makarem | Optics | Characterization | cavity realignment and new power measurements |
Thanks for clarifying that. We will repeat some power measurements and check the output offset voltage to the laser.
Quote: |
This is WOW!
Excellent mode matching work.
The measurement is still consistent with the low loss even with the different mode-matching level.
99.5%? The IFO commissioners will cry.
Edit: Wait a sec. The incident * mode matching = 20.14mW. This is the cavity-coupled power.
And you have the transmission of 9.78*2=19.56mW.
The ratio of these is ~97% and not 99.5%. Did I miss something?
=> Ah, understood. You have the incident power measurement with a significantly different reference voltage from the one at the transmission measurement. (4.21V vs 4.11V)
This is because the laser output power depends on the laser PZT feedback.
The quick hack to reduce this is that check the laser PZT feedback voltage (on the Thorlabs driver) right before the unlock, and bring the "output offset" close to that value after unlocking.
This brings the laser frequency back to the one during lock. At the same time, the laser freq is now close to the cavity resonance. So reading the unlocked REFL voltage, you need a bit of care.
|
|
613
|
Mon Aug 21 16:15:09 2023 |
Koji | Optics | Characterization | cavity realignment and new power measurements |
You should proceed to the next steps. You'll need to repeat the power measurements before the bonding. Next time, the measurement procedure will be improved. |
614
|
Tue Aug 22 10:25:10 2023 |
Thejas | Optics | Characterization | cavity realignment and new power measurements |
Thanks Koji, we did observe this discrepency and thought that the code normalizes the power/voltage readings to the reference power value. We will then proceed towards FSR and TMS measurements today.
Quote: |
You should proceed to the next steps. You'll need to repeat the power measurements before the bonding. Next time, the measurement procedure will be improved.
|
|
588
|
Thu Aug 3 15:04:26 2023 |
Stephen | General | General | aLIGO Unit 4 Build, Cable Routing, Preparation |
[Dean, Stephen]
Unit #4 aLIGO OMC unit was transported to the 40m Bake Lab.
We unwrapped the Transport Fixture and identified the orientation which would place the Light Side up and accessible. We removed the 4 long SHCS (Dark Side up), then rolled the Transport Fixture to the Light Side Up orientation. The PEEK pads were kept in the locked state, with set screws pushed forward. We loosened the set screws, retracted the PEEK pads, lifted the transport fixture lid, and saw no issues with the breadboard. While rotating the Transport Fixture, Stephen had allowed the Transport Fixture to tilt down a little bit harder than intended.
We noticed the oxidation of the PZT lead which had happened during acetone debonding (OMC_Lab/378).
We opened and inspected the Class A bulkhead 4-pin male Mighty Mouse connectors (Glenair p/n 803-003-07M6-4SN-598A) from Bake-7594. There are quantity 13 in the bag, with approximately the right amount of pins. We also opened and inspected the PEEK cable ties, aluminum Cable Pegs D1300057-v2, and additional button head screws for cable pegs (Koji had considerately installed some, while waiting for the Cable Pegs).
Dean added, to the PZT leads, the pins for Mighty Mouse connectors using a wiped-down crimper. He then bundled the PZT leads and routed the PZT wire bundles to Dark Side of the breadboard.
Koji supplied us with the on-board DCPD and QPD cables, and we anticipate that he will bring the DCPDs when we install the on-board cables. We are ready to install the on-board cables and complete Dark Side cable routing, and we will coordinate a work session with Koji.
Photos are posted to this Google Photos album: aLIGO Unit 4 Build, Cable Routing - https://photos.app.goo.gl/LAJ1Q6Fw9X8TrUBu6 |
530
|
Fri Apr 28 13:25:40 2023 |
Thejas | Optics | General | aLIGO OMC Unit-4 |
[Camille, Koji, Thejas]
Yesterday, we cleaned the cavity optics with first contact, aligned the input laser beam to the cavity and measured the power at different terminals on the cavity breadboard.

The measured OMC losses were:
SET1 0.042 +/- 0.003
SET2 0.035 +/- 0.002
SET3 0.030 +/- 0.0014
-> 0.033 +/- 0.001
The measured OMC mode-matching efficiencies were:
SET1 0.9795 +/- 0.00016
SET2 0.9797 +/- 0.00005
SET3 0.9794 +/- 0.00035
Attached herewith is the scrrenshot of the notes of with input power parameters. |
537
|
Thu May 11 10:46:17 2023 |
Thejas | General | Characterization | aLIGO OMC |
[Thejas, Camille, Koji]
We aligned the laser beam to the cavity and drove the OMC cavity PZTs (0 to 5 V from signal generator with 15x amp from the piezo driver) with a ramp signal and logged the transmission mode spectrum. The drive PZT voltage changes from 3.4 V to 7 V for one fringe shift or half wavelength change in cavity length. The voltage gain of the PZT driver is 15 V/V so that's a difference of 54 V for half weavelngth of driving or 532 nm/54V or 9.85 nm/V. |
291
|
Thu Feb 22 20:21:02 2018 |
Koji | Optics | Characterization | aLIGO EOM test |
POSTED to 40m ELOG |
292
|
Mon Apr 2 17:27:04 2018 |
Koji | Optics | Characterization | aLIGO EOM test |
2nd optical test http://nodus.ligo.caltech.edu:8080/40m/13725 |
153
|
Fri Aug 16 17:21:38 2013 |
Koji | General | General | [OMC002] Build started |
[Jeff Koji]
The breadboard (SN2) was loaded on the transportation fixture.
The laser side template was installed and the cavity mirrors were placed.
The laser beam will be resonated in the cavity next week. |
74
|
Wed Mar 20 09:38:02 2013 |
Zach | Optics | Characterization | [LLO] OMC test bench modified |
For various reasons, I had to switch NPROs (from the LightWave 126 to the Innolight Prometheus).
I installed the laser, realigned the polarization and modulation optics, and then began launching the beam into the fiber, though I have not coupled any light yet.
A diagram is below. Since I do not yet have the AOM, I have shown that future path with a dotted line. Since we will not need to make AMTFs and have a subcarrier at the same time, I have chosen to overload the function of the PBS using the HWP after the AEOM. We will operate in one of two modes:
- AMTF mode: The AOM path is used as a beam dump for the amplitude modulation setup. A razor dump should physically be placed somewhere in the AOM path.
- Subcarrier mode: The AEOM is turned off and the HWP after it is used to adjust the carrier/subcarrier power ratio. I chose a 70T / 30R beamsplitter for the recombining, since we want to be able to provide ~100 mW with the carrier for transmission testing, and we don't need a particularly strong subcarrier beam for probing.

One thing that concerns me slightly: the Prometheus is a dual-output (1064nm/532nm) laser, with separate ports for each. I have blocked and locked out the green path physically, but there is some residual green light visible in the IR output. Since we are planning to do the OMC transmission testing with a Si-based Thorlabs power meter---which is more sensitive to green than IR---I am somewhat worried about the ensuing systematics. I *think* we can minimize the effect by detuning the doubling crystal temperature, but this remains to be verified.
EDIT (ZK): Valera says there should be a dichroic beam splitter in the lab that I can borrow. This should be enough to selectively suppress the green. |
146
|
Tue Jun 18 15:45:38 2013 |
Koji | General | General | [LLO] OMC installation in HAM6 |
OMC installed in HAM6!
https://alog.ligo-la.caltech.edu/aLOG/index.php?callRep=7486
|
140
|
Fri Jun 7 16:04:44 2013 |
Koji | General | General | [LLO] OMC arrived |
[Koji Zach Suresh]
The OMC arrived at LLO without any destruction!
- We found that one shock sensor on the box turned red, the other stayed white.
- We brought the Perican case to the changing room and the wrapping was opened in the optics lab.
- The OMC was discovered without any obvious damage. Successful shipment!
- The inspection with a halogen light indicated some amount of particules on the breadboard.
The both sides of the breadboard were wiped with the cleanroom cloth.
- The First Contact layers on the optics were removed while the ionized nitrogen gas was brew.
|
143
|
Thu Jun 13 12:12:20 2013 |
Zach | General | General | [LLO] OMC and OMCS in LVEA |
https://alog.ligo-la.caltech.edu/aLOG/index.php?callRep=7395
|
141
|
Mon Jun 10 10:04:37 2013 |
Koji | General | General | [LLO] OMC Test |
[Koji Zach]
We worked on the OMC test over the weekend.
- At the beginning, the measured OMC transmission was ~85% even after subtracting the junk light and sidebands from the calculation.
- A pretty visible (by eye) dust were on CM1. Also a small residue of First Contact was found on the same mirror.
- We applied FC only on CM1 to remove these.
- The measued transmission went up to the level of 96%.
- We swept the incident power from 0.3mW to 30mW in order to see the dependence of the transmission against the incident power.
- The variation of the transmission ~10% was observed (attached figure 1, Red). This was compared with the similar dependence measured at Caltech (Magenta)
- So, the reduction of the transmission was observed as in eLIGO, although the measurements at Caltech and LLO are not consistent.
- Can this be attributed to the dependence of the PD efficiency? We measured the incident power on the PDs together with the preamp DC output. (Figure.2)
This gives us how the responsivity changes with the incident power.
- Nevertheless, the dependence remains. We'll make more accurate measurement today. |
142
|
Thu Jun 13 12:04:57 2013 |
Koji | General | General | [LLO] OMC Optical Test - completed |
https://alog.ligo-la.caltech.edu/aLOG/index.php?callRep=7373 |
144
|
Fri Jun 14 06:35:21 2013 |
Jeff | General | General | [LLO] L1 OMC status |
https://alog.ligo-la.caltech.edu/aLOG/index.php?callRep=7410
Weights:
Suspension cage and transportation box: 250.8lb
Suspension cage and transportation box: 150.2lb ==> 100.6lb ==> 45,630 g
Metal Breadboard: 7261 g
Glass Breadboard and transportation fixture: 16382 g
Transportation fixture only: 9432 g ==> 6950 g
Added mass (up to now): 300 g ==> 7250 g
Preamp arrangement
|
521
|
Thu Apr 13 07:47:28 2023 |
Camille Makarem | Optics | Characterization | Zygo setup for curved mirror measurements |
[Camille, Thejas, Stephen]
We modified the Zygo setup for measuring the sagitta of the curved mirrors. A mirror at 45deg was used to reflect the interferometer beam down towards the surface of the table (see picture). A fused silica flat was placed horizontally with the surface of the table and was used as our reference flat. The back surface of the curved mirror and the top surface of the reference flat were cleaned using top gun and/or swabs. Once it was verified that the surfaces were clean, the curved mirror could be easily placed on the surface of the reference flat.
Once the curved mirror was placed on the reference flat, the fringes of the reference flat were nulled using the 45deg mirror. After nulling the flat's fringes, the data was recorded. The curved mirror was then rotated 90deg clockwise. The measurment was repeated with the curved mirror's fiducial located at 12:00, 3:00, 6:00 and 9:00. The 12:00 position was measured twice to ensure repeatability. (A drop of first contact had been placed at the edge of the optic to indicate where the fiducial arrow is. This helped with clocking alignment.)
The already-characterized aLIGO C7 mirror was measured to verify the setup. After verifying agreement with the results in T1500060, this process was repeated with all the remaining curved mirrors.
The data was analyzed using Thejas's python script (separation distance between mirror center and curvature minimum, angular position of curvature minimum.) Those mirrors with a large spread in the measurements will be remeasured.
|
19
|
Wed Aug 22 20:16:43 2012 |
Koji | Facility | General | Workbenches have been installed / Clean room stools |
Last Friday, new workbenches were installed. Vladimir got a new table and a cleanroom stool.

The other two workbenches were also nicely set.

|
300
|
Mon Jul 2 15:27:31 2018 |
Rich Abbott | Electronics | General | Work on EOM (3rd IFO unit) |
Koji, Rich
We took apart the unit removed from the 3rd IFO (Unit serial number aLIGO #3, XTAL 10252004) to see what makes it tick. Koji has done a fine job of adding the plots of the impedance data to this log book. Attached are some details of the physical construction showing the capacitor values used in shunt before the coils. |
33
|
Wed Nov 7 20:21:42 2012 |
Koji | General | General | Work completed in October [!] |
Completed work of the previous months: [Jul] [Aug] [Sep] [Oct] [Nov] [Dec]
- Work done
- Particle Level measured / HEPA activated [ELOG]
- Particle counter peripherals arrived ~Oct 12.
- Making the OMC optical test setup [ELOG] [ELOG] [ELOG] [ELOG] [ELOG]
- OMC Bread board dimensions / weights measurement by Jeff and Jam [ELOG]
- UV epoxy has arrived - stored in a freezer in the office
- Laser sign installed during my trip by Peter/Eric
- OMC design downselect [DCC Link]
|
47
|
Mon Dec 31 01:45:04 2012 |
Koji | General | General | Work completed in Nov and Dec [!] |
Completed work of the previous months: [Jul] [Aug] [Sep] [Oct] [Nov] [Dec]
- Things delivered
- The ionized gun used in the clean room at Downs: made by Terra Universal.com (Jeff's room)
http://www.terrauniversal.com/static-control/ionizing-blow-off-guns.php
- Flow path: N2 cylinder - Filter - Gun (Jeff's room)
- Power strips Tripp Lite PS3612 (Ordered Nov. 8, Delivered Nov. 12)
- Kapton tapes (1in x 6, 1/2in x 12 Delivered Nov. 15)
- Sticky Mats (VWR 18888-216 Delivered Nov. 12 and 21992-042)
- Duck tape (PK3) (Delivered Nov. 12)
- Wipers 12"x12" 2ply x 119 pairs x case15 (Delivered Nov. 12)
- Syringes (1mL&2mL) & Needles (20G x dozen)
- Stainless trays with cover (Steve Delivered Nov. 12)
- Gold Plated allen keys (Steve Delivered Nov. 12)
- Forceps (Delivered Nov. 12) / Tweezers / Scissors (Delivered Nov. 12)
- OMC testing optics / opto-mechanics
- SolidWorks raytracing model
- Mode design for HAM6 layout [Zach]
- Black Glass / Black Glass holder / AR ==> Some at the 40m, some from LLO
- Ionized air blow
- N2 or Air cylinder: 4N - UHP or 5N - Research Grade. (... steal from Downs)
|
12
|
Tue Jul 31 21:29:43 2012 |
Koji | General | General | Work completed in July [!] |
- Completed Work: Facility/Supplies
- Plug a big hole on the wall [ELOG]
- Purchasing work benches
- Wooden work benches removed(arranging the work with Louisa)
- Al foils (All foils inc, should get a certificate everytime to ensure UHV compatibility)
- Laser / UV safety glass/face mask (Ordered with Gina, UV face shield ordered through Techmart)
- Sticky mat
- VWR MAT ADHESIVE 30L 18X36 BLU, 21924-110
- Shoe cover
- VWR SHOECVR NSKID AP XL 150PR, 414004-650
- VWR SHOECVR NSKID AP 2XL 150PR, 414004-651
- Lab coat
- VWR Lab coat L 82007-618 / XL 82007-620
- Hat
- Mask
- Gloves
- VWR GLOVE ACCTCH NR-LTX SZ7.5 PK25 79999-306 x4
- VWR GLOVE ACCTCH NR-LTX SZ8 PK25 79999-308 x4
- Plastic boxes for storage
http://www.drillspot.com/products/422140/Rubbermaid_2282-00-CLR_18GAL_Clear_Snap_Case
(We have 12 for now. More stored at the 40m)
- Completed Work: Optics
- UV Lamp arrived (shipped from LLO)
- Fiber light guide for UV lamp (Quote obtained / Ordered via techmart)
- Optical test planning by Zach [ELOG]
- How to handle First Contact by Margot [ELOG]
- Useful links / OMC scanning [ELOG]
|
28
|
Tue Oct 16 15:50:09 2012 |
Koji | General | General | Work completed in August/September [!] |
- Work done
- Things ordered
- Office Depot
- [delivered] Office Depot(R) Brand Stretch Wrap Film, 20 x 1000 Roll, Clear / 445013
- [delivered] Eveready(R) Gold AA Alkaline Batteries, Pack Of 24 / 158448
- [delivered] Rubbermaid(R) Roller Sponge Mop / 921841
- [delivered] Rubbermaid(R) Roller Sponge Mop Replacement / 921858
- [delivered] Rubbermaid(R) Sanitizing Caddy, 10 Quarts, Yellow / 674125
- [delivered] Glad(R) Tall Kitchen Trash Bags, 13 Gallon, White, Box Of 28 / 269268
- Global Industrial Equipment
- [delivered] Extended Surface Pleated Cartridge Filter Serva-Cell Mp4 Slmp295 12X24X2 Gl WBB431699
- Global Industrial Equipment
- [delivered] Nexel Poly-Z-Brite Wire Shelving 30"W x 21"D x 63"H Nexel Poly-Z-Brite™ Wire Shelving Starter Unit WB189209
- [delivered] Stem Casters Set of (4) 5" Polyurethane Wheel, 2 With Brakes 1200 lb. Capacity WB500592
- Rack Solutions
- [delivered] Open Frame Server Racks
1 x 20" Depth Kit (Ideal for Audio/Video or Networking Racks) P/N: 111-1779
1 x 36U, Rack-111 Post Kit P/N: 111-1728
1 x Caster Kit for Open Frame RACK-111 P/N: 111-1731
- [delivered] 36U Side Panel Kit $199.99 P/N: 102-1775
- Rack shelf
- [delivered] 1 RMS 19 X 15 SINGLE SIDED NON-VENTED SHELF 70121637
- Work bench, Stools
- [not yet] 72"L X 30"W Production Bench - Phenolic Resin Square Edge-Blue Form attached WB237381LBL
- [not yet] 72"W Lower Shelf For Bench - 15"D- Blue Form attached WB606951
- [not yet] ESD-Safe Vinyl Clean Room Stool with Nylon Base with Drag Chain Blue Form attached WBB560852
- P Touch
- [delivered] Brother PT-2030 Desktop Office Labeler Punch-out product 672828
- [delivered] Brother(R) TZe-241 Black-On-White Tape, 0.75 x 26.2 Punch-out product 239384
- [delivered] Brother(R) TZe-231 Black-On-White Tape, 0.5 x 26.2 Punch-out product 239400
- UV light guide
- [delivered] Fiber Optic Single Light Guide 5mm OD X 3mm ID X 1M L Note: This light guide can be used with MKIII UV Cure unit. OLB1081
- Gloves (7.5, 8.0)
- [delivered] GLOVE ACCTCH NR-LTX SZ7.5 PK25 Punch-out product 79999-306
- [delivered] GLOVE ACCTCH NR-LTX SZ8 PK25 Punch-out product 79999-308
- Lab coat (L,XL), Sticky Mat, Shoe Covers (L, XL), Cap, Mask
- [delivered] LAB XP WH EL WR.COLL. NP L30EA Punch-out product 82007-618
- [delivered] LAB XPWH EL WR.COLL. NP XL30EA Punch-out product 82007-620
- [delivered] VWR MAT ADHESIVE 30L 18X36 BLU Punch-out product 21924-110 (This was too small)
- [delivered] VWR SHOECVR NSKID AP 2XL 150PR Punch-out product 414004-651
- [delivered] VWR SHOECVR NSKID AP XL 150PR Punch-out product 414004-650
- [delivered] CAP BOUFFANT 24IN RAYON CS500 Punch-out product 10843-053
- [delivered] MASK VLTC TIES N/STRL PK50 Punch-out product 10869-020
- VWR
- [delivered] FACE SHIELD UVC-803 Supplier: UVP 33007-151
- [Delivered] Laser safety glasses
|