[Stephen, Koji]

The perpendicularity of some of the A and M prisms were tested.

Results

- The measurement results are listed as Attachment 1 and 2 together with the comparisons to the measurement in 2013 and the spec provided from the vendor.
- Here, the positive number means that the front side of the prism has larger angle than 90deg for the air side. (i.e. positive number = facing up)
- The RoC of the curved mirrors is 2.5m. Therefore, roughly speaking, 83arcsec corresponds to ~1mm beam spot shift. The requirement is 30 arcsec.
- The A prisms tend to have positive and small angle deviations while the M prisms to have negative and large (~50arcsec) angle deviations.
- The consistency: The measurements in 2013 and 2019 have some descrepancy but not too big. This variation tells us the reliability of the measurements, say +/-30arcsec.

Setup

- The photos of the setup is shown as Attachments 3/4/5. Basically this follows the procedure described in Sec 2.2.2 of T1500060.
- The autocollimator (AC) is held with the V holders + posts.
- The periscope post for the turning Al mirror was brought from Downs by Stephen.
- The turning mirror is a 2" Al mirror. The alignment of the turning mirror was initially aligned using the retroreflection to the AC. Then the pitching of the holder was rotated by 22.5deg so that the AC beam goes down to the prism.
- The prism is held on a Al mirror using the post taken from a prism mount.
- If the maximum illumination (8V) is used, the greenish light becomes visible and the alignment becomes easier.
- There are two reflections 1) The beam which hits the prism first, and then the bottom mirror second, 2) The beam which hits the bottom mirror first and then the prism second. Each beam gains 2 theta compared to the perfect retroreflection case. Therefore the two beams have 4 theta of their relative angle difference. The AC is calibrated to detect 2 theta and tells you theta (1div = 1 arcmin = 60 arcsec). So just read the angle defferencein the AC and divide the number by 2 (not 4).

Attachment 1: A_prism.png  42 kB  Uploaded Mon Jul 1 13:39:19 2019  | Show | Hide all | Show all
Attachment 2: M_prism.png  59 kB  Uploaded Mon Jul 1 13:39:19 2019  | Show | Hide all | Show all
Attachment 3: P_20190627_222658.jpg  320 kB  Uploaded Mon Jul 1 13:47:28 2019  | Show | Hide all | Show all
Attachment 4: setup2.JPG  3.578 MB  Uploaded Mon Jul 1 13:47:51 2019  | Hide | Hide all | Show all
setup2.JPG
Attachment 5: M01_1_id.JPG  3.117 MB  Uploaded Mon Jul 1 13:48:30 2019  | Hide | Hide all | Show all
M01_1_id.JPG
Attachment 6: A14_meas.JPG  2.620 MB  Uploaded Mon Jul 1 15:05:38 2019  | Show | Hide all | Show all
ELOG V3.1.3-