Entry  Sun Jul 13 17:46:28 2014, Koji, Optics, Characterization, OMC backscatter measurement OMC_backscatter.pdf
    Reply  Tue Jul 15 03:00:42 2014, Koji, Optics, Characterization, OMC backscatter measurement 
       Reply  Tue Jul 15 03:34:16 2014, Koji, Optics, Characterization, OMC backscatter measurement OMC_backscatter.pdfOMC_backscatter.png
Message ID: 209     Entry time: Tue Jul 15 03:34:16 2014     In reply to: 208
Author: Koji 
Type: Optics 
Category: Characterization 
Subject: OMC backscatter measurement 

Backscatter measurement ~ 2nd round


Summary

- The backscatter reflectivity of the 3rd OMC is 0.71 ppm

- From the spacial power distribution, it is likely that this is not the upper limit but the actual specular spot from the OMC,
propagating back through the input path.


Improvement

- The power meter was heavily baffled with anodized Al plates and Al foils. This reduced many spourious contributions from the REFL path and the input beam path.
  Basically, the power meter should not see any high power path.

- The beam dump for the forward going beam, the beamsplitter, and the mirrors on the periscope were cleaned.

- The power meter is now farther back from the BS to reduce the exposed solid angle to the diffused light

- The REFL path was rebuilt so that the solid angle of the PD was reduced.

OMC_backscatter.png


Backscattering measurement

- Pin = 12.3 +/- 0.001 [mW]

- RBS = 0.549 +/- 0.005

- Pback = 4.8 +/- 0.05 [nW] (OMC locked)       ==> ROMC(LOCKED) = 0.71 +/- 0.01 [ppm]

- Pback = 3.9 +/- 0.05 [nW] (OMC unlocked)   ==> ROMC(UNLOCKED) = 0.57 +/- 0.01 [ppm]

Note that the aperture size of Iris(B) was ~5.5mm in diameter. 


V-dump test

- Additional beam dump (CLASS A) was brought from the 40m. This allowed us to use the beam dump before and after the periscope.

- When the beam dump was placed after the periscope: P = 0.9+/-0.05nW

- When the beam dump was placed before the periscope: P=1.0+/-0.1nW

===> This basically suggests that the periscope mirrors have no contribution to the reflected power.

- When the beam dump was placed in the REFL path: P=2.1+/-0.1nW


Trial to find backward circulating beam at the output coupler

The same amount of backreflection beam can be found not only at the input side of the OMC but also transmission side.
However, this beam is expected to be blocked by the beamsplitter. It was tried to insert a sensor card between the output coupler
and the transmission BS, but nothing was found.


In order to see if the detected power is diffused light or not, the dependence of the detected light power on the aperture size was measured.
Note that the dark offset was nulled during the measurement.

IRIS B
aperture   detected
diameter   power

[mm]       [nW]
 1.0        1.1

 2.5        2.6
 4.25       4.0
 5.5        4.6
 8.0        5.3
 9.0        6.1
11.0        6.3
15.0        7.0

We can convert these numbers to calculate the power density in the each ring. 
(Differentiate the detected power and aperture area. Calculate the power density in each ring section, and plot them as a function of the aperture radius)


This means that the detected power is concentrated at the central area of the aperture.
(Note that the vertical axis is logarithmic)

If the detected power is coming from a diffused beam, the power density should be uniform.
Therefore this result strongly suggests that the detected power is not a diffused beam but
a reflected beam from the OMC.

According to this result, the aperture size of 2.6mm in raduis (5.5mm in diameter) was determined for the final reflected power measurement.

Attachment 1: OMC_backscatter.pdf  887 kB  Uploaded Tue Jul 15 04:59:53 2014  | Show | Hide all | Show all