1) Deburr the bottom surfaces of the QPD housings
2) Aligned the QPDs
QPD# QPD1 QPD2
Housing# #004 #008
Diode# #44 #46
Shim 1.75mm 001 1.25mm 001
-------------------------------------
Power Incident 125.7 uW 126.4 uW
Sum Out 80.1 mV 78.9 mV
Vertical Out + 3.4 mV 0 mV
Horizontal Out -23.7 mV -26 mV
SEG1 -15.6 mV -13.2 mV
SEG2 -13.1 mV -13.3 mV
SEG3 -29.0 mV -26.4 mV
SEG4 -23.2 mV -26.3 mV
-------------------------------------
Spot position X -13 um - 0.8 um (positive = more power on SEG1 and SEG4)
Spot position Y +93 um +107 um (positive = more power on SEG3 and SEG4)
-------------------------------------
Responsivity[A/W] 0.64 0.62
Q.E. 0.74 0.73
-------------------------------------
Arrangement of the segments
View from the beam
/ 2 | 1 X
|---+---|
\ 3 | 4 /
---------------
I(w,x,y) = Exp[-2 (x^2 + y^2)/w^2]/(Pi w^2/2)
(SEG_A+SEG_B-SEG_C-SEG_D)/(SEG_A+SEG_B+SEG_C+SEG_D) = Erf[sqrt(2) d/w]
d: distance of the spot from the center
w: beam width
|