  Draft   Mon Oct 31 23:51:15 2022     
  Draft   Fri Oct 28 21:39:34 2022     
  114   Thu Oct 27 22:12:21 2022 ranaGeneralOptical Contactingplotting and PID

The Arduino / AC PWM interface looks good. I recommend that you maintain the code in GitHub and post a link to the repo whenever you update the code. Use detailed commit messages so that it makes sense.

For the plotting, it would be good if you can use grid lines and markers for the data points. Then we can see the difference between the data and the fits, etc.

And to avoid the hysteresis, etc. you can record the temperature in your Arduino and use feedback to make the heater just go to whatever temperature you specify. So you would have a prescribed T(t) and the PID feedback loop would just make the heater take you there. Can your Arduino read the thermocouple?

  Draft   Thu Oct 27 19:54:20 2022 Jennifer Hritz   

Somehow I never thought of this before, but instead of increasing the "on" time of the hot plate to account for the heating drop-off, I should keep that constant and instead decrease the "off" time. That feels more logical given that I am trying to keep the temperature of the two plates as close as possible.

  112   Wed Oct 26 21:27:23 2022 JenniferGeneralOptical ContactingPWM 3.1 (more) very slow heating

[I'm (once again) behind on data processing, but I'm creating an entry on the day I actually run the tests]

  Draft   Tue Oct 25 18:55:38 2022 Jennifer HritzGeneralOptical ContactingPWM 3.1 very slow heating

[I'm behind on data processing, but I'm creating an entry on the day I actually run the tests]

  110   Mon Oct 24 21:23:22 2022 Jennifer HritzGeneralOptical ContactingPWM 3.1 longer off times in cycle

To combat the bottom plate heating up much faster than the top plate, I decided to try increasing the cycle period from 1000ms (1s) to 10000ms (10s). In other words, taking the test I today ran as an example, the hot plate will now be on for 1000ms then off for 9000ms then repeat. Hopefully this should give more time for the heat to transfer to the top plate, but even in this short test, it still appears to be a problem.

Due to the slower heating times, this will be a bit more challenging to test as each test could take hours to complete, but this is more in line with the final intended use anyways. Perhaps my cycle of 1000ms on is too much (e.g. I should do 100ms on then 9900ms off, although I think that might be so slow that it will never heat up; this also raising the question as to how I will deal with mantaining this slow heat up at the higher temperatures).

Attachment 1: PWM_plates_-_1000_ms_on_9000_ms_off.pdf
PWM_plates_-_1000_ms_on_9000_ms_off.pdf
  109   Sun Oct 23 21:01:40 2022 Jennifer HritzGeneralOptical ContactingPWM 3.0 (+280C in 5)

I decided test how fast the plates would heat up if the heat was just on constantly on for 5 minutes. In general, these tests are raising a lot of questions in regards to controlling the temperature given the hysteresis in the system. It is also apparent that the bottom plate heats up signficantly faster than the top one, which means I need to heat the samples much longer than, say 10 minutes, if I want to avoid unevenly heating both parts of the optically contacted piece.

I also have to be conscientious that I am already half way through the quarter and ideally should be devoting time to bond strength testing rather than continuing to fiddle with the hot plate.

Attachment 1: PWM_plates_-_increase_280°C_in_5_minutes.pdf
PWM_plates_-_increase_280°C_in_5_minutes.pdf
  108   Sat Oct 22 21:18:56 2022 Jennifer HritzGeneralOptical ContactingPWM v3.0

With v3.0, I took a couple steps backwards by getting rid of the feature that increases the heating rate so I can isolate the base heating rate for the two plates. In my experience, the best way to figure out how to modify the program is to try a bunch of different target temperatures and heating times and look for correlations. I started with (attempting) to increase the plates by 280°C in 10 minutes.

For a future release, I am thinking of radically (relatively speaking) changing the function parameters: the user only inputs the target heating rate and how long the plates should be heated at this rate. This is to address the hysteresis in this new set-up, which I will elaborate on if I make the change.

Attachment 1: PWM_plates_-_increase_280C_in_10_minutes.pdf
PWM_plates_-_increase_280C_in_10_minutes.pdf
  107   Fri Oct 21 22:22:37 2022 Jennifer HritzGeneralOptical ContactingPWM v2.6 on new thermocouple set up

I performed the same tests I have been doing prior (+180°C in 10 minutes) but now with the (correctly wired) thermocouples attached to the metal plates. The top plate is thermocouple #1 attached to the Fluke and the bottom plate is thermocouple #2 attached to the TPI (the lime green one).

The base heating rate for the new set up will require some tweaking to the code because the plates heat up much slower, but as I have mentioned previously, I do not think this will require a lot of extra work since I now know the tips and tricks to PWMing the hot plate. The only difficulty might come from the increase in hysteresis (i.e. the plates continue to increase in the temperature long after it turns off). For future tests, I need to remember to continue recording the temperature after program finishes its 10 min cycle.

On the positive, I think this test shows that taking the average of the two thermocouples to find the temperature in the center (where the optically contacted samples are) is a worthwhile endevor, considering how much the top plate lags behind the bottom plate in terms of heating speed.

Attachment 1: PWM_plates_-_increase_180C_in_10_minutes.pdf
PWM_plates_-_increase_180C_in_10_minutes.pdf
  106   Thu Oct 20 22:43:05 2022 Jennifer HritzGeneralOptical ContactingWired the thermocouples backwards

Since the two devices are giving different temperature readings, I would like to find out if this imprecision is linear (e.g. they are always 3°C off, so I just need to add/subtract 3°C after taking the measurements). If not, some sort of calibration is probably required. I decided to figure this out by running the heating tests I did before, but this time with the plates. This also serves as a test to see how the plates heat up.

Or rather, this is what I would have done, had I not realized that the thermometers were going down as the heat was increaing, meaning I had switched the polarity for both thermocouples. It turns out that this mix-up is a common mistake. I thought that I double checked that red was positive for thermocouples, but it is in fact not:
"red is the usual color for positive charges, whereas the red wire in thermocouple cables typically contains the negative signal. This coloration is ANSI standard for thermocouples, but it is not what most people expect."

Attachment 1: wrong_correct_polarization_PXL_20221022_054250498_20221022_055117337.MP.jpg
wrong_correct_polarization_PXL_20221022_054250498_20221022_055117337.MP.jpg
  105   Wed Oct 19 21:51:10 2022 Jennifer HritzGeneralOptical ContactingThe trials and tribulations of the thermocouples

I intended to test the new thermocouple set up today, but when I plugged them in, both did not read a temperature. It took me a long time to figure out what went wrong: when installing the K Type thermocouple connector, the wires of the thermocouple need to be pushed in as far in as possible, otherwise the circuit would not be completed. It took a lot of trial and error to figure this out. I first created a test "circuit" with wire and a resistor to make sure that the connector itself was not broken. Then I carefully observed how moving the wires in different places affected the reading.

Once I did carefully reassemble the thermocouples, they worked perfectly, as indicated by the non-zero current. I ran tests with my three thermocouples and two devices to see how precise the temperature reading is. The results are below and pictures of the readings can be found in the zip file. I cannot explain why one of the adhered thermocouples is hotter than the other.

(°C) Fluke 17B+
Digital Multimeter
Digital Thermometer 343
Thermocouples T1 T2
Adhered to Plate #1 23.1 27.4 26.1
Plate #2 26.6 28.8 28.5
Rod placed
on
Plate #1 21.7 24.5 24.0
Plate #2 21.3 24.1 24.0

Plate #1 and 2 refers to the two different aluminum plates. T1 and T2 refers to the two ports on the Digital Thermometer 343. It cannot read two thermocouples simultaneously (as far as I can tell); it's so one can be used as a baseline/reference value for the other.

Attachment 1: thermocouple_connector_test_ciruit_PXL_20221020_035719119.MP.jpg
thermocouple_connector_test_ciruit_PXL_20221020_035719119.MP.jpg
Attachment 2: correct_assembly_PXL_20221020_042855181.MP.jpg
correct_assembly_PXL_20221020_042855181.MP.jpg
Attachment 3: resistance_through_broken_and_fixed_connector_PXL_20221020_035951538_20221020_041939698.MP.jpg
resistance_through_broken_and_fixed_connector_PXL_20221020_035951538_20221020_041939698.MP.jpg
Attachment 4: broken_and_fixed_thermocouple_PXL_20221020_042242792_20221020_042201775.MP.jpg
broken_and_fixed_thermocouple_PXL_20221020_042242792_20221020_042201775.MP.jpg
Attachment 5: thermocouples_and_thermometers_room_temp.zip
  104   Tue Oct 18 19:33:38 2022 Jennifer HritzGeneralOptical ContactingSetting up thermocouples

Firstly, last night's heating did not change the contacted surface area greatly, but there is too many factors to speculate as to why that is the case. I leave that for future testing.

I attached the thermocouples by adhering them to the two aluminum plates. I was careful to make sure that the thermocouple was in the dead center of the aluminum plate. The other end of the thermocouples—exposed positive and negative wires—were screwed into the K Type connector so they can be plugged into the thermometer/multimeter. Taking the average between the top and bottom plate will give a more precise estimate of the temperature of the samples.

Attachment 1: first_real_PWM_test_samples_before_after_PXL_20221018_080838426.MP.jpg
first_real_PWM_test_samples_before_after_PXL_20221018_080838426.MP.jpg
Attachment 2: K_Type_connector_PXL_20221019_000708969.MP.jpg
K_Type_connector_PXL_20221019_000708969.MP.jpg
Attachment 3: thermocouple_plates_PXL_20221019_003002817.MP.jpg
thermocouple_plates_PXL_20221019_003002817.MP.jpg
  103   Mon Oct 17 23:17:25 2022 Jennifer HritzGeneralOptical ContactingTesting PWM code with actual samples

Now that I have (relatively) good PWM code, I wanted to do my first real test with actual samples. Since everything went smoothly, I will now work on building the original set up for the project, which included attaching thermocouples to two plates so we could precisely measure the heat between them.

As you can see in the pictures below, I am running an Arduino off of my laptop which controls an AC/DC control replay that turns the AC power to the hot plate on and off.

Attachment 1: first_real_PWM_test_samples_before_PXL_20221018_071840794.MP.jpg
first_real_PWM_test_samples_before_PXL_20221018_071840794.MP.jpg
Attachment 2: first_real_PWM_test_set_up_PXL_20221018_071943406.MP.jpg
first_real_PWM_test_set_up_PXL_20221018_071943406.MP.jpg
Attachment 3: first_real_PWM_test_in_action_PXL_20221018_072832135.mp4
  102   Sun Oct 16 14:20:32 2022 Jennifer HritzGeneralOptical ContactingSamples after 2 weeks under pressure

Since I was focusing on the hot plate code and therefore did not need my weights, I decided to leave them on top of my samples for roughly 2 weeks.

It appears that an increased amount of time under pressure does not result in any noticable differences. A slight increase in surface area (SA) in two places, and a slight decrease in SA in another place, but overall no change. Note that "(initally)" in the picture below refers to http://nodus.ligo.caltech.edu:8081/Mariner/89.

Attachment 1: 2_weeks_pressure_PXL_20221004_190652540_20221017_074745619.jpg
2_weeks_pressure_PXL_20221004_190652540_20221017_074745619.jpg
  101   Sat Oct 15 21:59:13 2022 Jennifer HritzGeneralOptical ContactingHot plate PWM v2.6 Test #1, 2, 3

I realized that, after changing so much from v2.3 to 6, I should check that my first two tests produce correct results with the latest version. This was good because all three tests turned out to be innaccurate, as they were all short roughly 10°C. However, they were very precise. For all three, the final temperature was 193.15±1.5°C.

Attachment 1: PWM_v2.6_Test_#1_2_3_-_increase_180C_in_10_7_20_minutes.pdf
PWM_v2.6_Test_#1_2_3_-_increase_180C_in_10_7_20_minutes.pdf
Attachment 2: PWM_v2.6_tests_data.zip
  100   Fri Oct 14 21:36:52 2022 Jennifer HritzGeneralOptical ContactingHot plate PWM v2.6 Test #3

The goal of "v2.X test #3" is to heat the hot plate to 200°C over the course of 20 minutes, and with v2.6, I have effectively succeeded. There will likely be more issues once I try, for example, to heat the hot plate to 300°C over the course of 60 minutes, but for now, I want to stick with lower temps and shorter times while I work out the kinks. Now that I understand the difficulties of PWMing a hot plate, adapting the code to combat future issues should be straightforward.

To summarize my code, I control the heating rate by cycling the hot plate's power on and off for some % of 1000ms. In other words, the hot plate is on 300ms then off 700ms then on 300ms etc., where the relation between target heating rate and hot plate on time is based on previously gathered data. This produces a nice, linear(ish) temperature increase up until a certain temperature, at which point it plateaus. In the previous versions, the way I compensated for this was by increasing the on time by 5ms for every cycle after 150°C. This did not work for slower heating rates, so the newer versions changed this by making the 5ms and 150°C varry depending on the target heating rate. The exact value is a linear extrapoliation from previous data. This is imperfect, but I do not think perfection will ever be possible with the current equipent, and I think I have reached something good enough that now I can finally apply it to my optically contacted samples.

Since I have finished this "stage" of work, for completeness, I am including all of the code, data*, and graphs involved so far.
*the .txt data files are in the cycle_vX_graphs folders; these folders also have the Jupyter notebooks I used for graphing the data

Attachment 1: PWM_v2_Test_#3_-_increase_180°C_in_20_minutes.pdf
PWM_v2_Test_#3_-_increase_180°C_in_20_minutes.pdf
Attachment 2: PWM_v2.6_Test_#3_-_increase_180°C_in_20_minutes.pdf
PWM_v2.6_Test_#3_-_increase_180°C_in_20_minutes.pdf
Attachment 3: cycler_v1_graphs.zip
Attachment 4: cycler_v2_graphs.zip
Attachment 5: hot_plate_cycler_v1-2.6.zip
  99   Thu Oct 13 20:56:48 2022 Jennifer HritzGeneralOptical ContactingHot plate PWM v2.4,5 Test #3

My two corrections ended up being huge overshoots. The drop off time (100°C) is correct, but the default rate increase that worked in the other cases is not working at all here.

Attachment 1: PWM_v2.3_Test_#3_-_increase_180C_in_20_minutes.pdf
PWM_v2.3_Test_#3_-_increase_180C_in_20_minutes.pdf
  98   Wed Oct 12 23:26:48 2022 Jennifer HritzGeneralOptical ContactingHot plate PWM v2.3 Test #3

I tried increasing the temperature by 180°C over 20 minutes. As suspected, it did not quite reach the target temperature because the temperature started to drop off around 100°C instead of 150°C, as the program expected. This should be an easy adjustment, since it is just a matter of increasing the duration of the cycle at an earlier time.

Attachment 1: PWM_v2.3_Test_#3_-_increase_180°C_in_20_minutes.pdf
PWM_v2.3_Test_#3_-_increase_180°C_in_20_minutes.pdf
  97   Tue Oct 11 23:59:07 2022 Jennifer HritzGeneralOptical ContactingHot plate PWM v2.3 Test #1 and 2

Here are the graphed results from yesterday's tests, both by themselves and overlayed with the previous tests. I am satisfied with my code; it has given me the (roughly) linear heat increase that I desired. The only last thing I would like to test is heating over a signficantly slower time. 

Attachment 1: PWM_v2.3_Test_#1_-_increase_180°C_in_10_minutes.pdf
PWM_v2.3_Test_#1_-_increase_180°C_in_10_minutes.pdf
Attachment 2: PWM_v2.3_Test_#2_-_increase_180°C_in_7_minutes.pdf
PWM_v2.3_Test_#2_-_increase_180°C_in_7_minutes.pdf
Attachment 3: PWM_v2_Test_#1_-_increase_180°C_in_10_minutes.pdf
PWM_v2_Test_#1_-_increase_180°C_in_10_minutes.pdf
Attachment 4: PWM_v2_Test_#2_-_increase_180°C_in_7_minutes.pdf
PWM_v2_Test_#2_-_increase_180°C_in_7_minutes.pdf
  96   Mon Oct 10 15:34:13 2022 Jennifer HritzGeneralOptical ContactingHot plate PWM v2.1,2,3 Test #1 and v2.3 Test #2

Before trying the PWM on actual samples, I wanted to make one final attempt at improving my code (labled as v2.1). This change appears to have 1) broken the code regulating the basic heat cycling process 2) caused the hot plate to heat up far, far too quick. Since the thermometer strangely turned off halfway through, I only have two pictures as evidence that this test existed: a screenshot of the Arduino program telling me that the max cycle rate had been reached (which should have not happened) and a frame from the video filming the thermometer showing the peak temperature (which is 100°C high than expected). Somehow the hot plate reached over 300°C, which I thought was impossible because the hot plate's built-in heat cycle should have kicked in around 260°C. Unrelated, but I am performing this test in my dorm room because I was quarentined due to COVID exposure, and I like using my personal fan and the house's freezer to cool down the hot plate quicker.

I made some adjustments (labled as v2.2), and I had the same failure as v2.1, except I managed to capture it on camera.

Finally, with v2.3, I managed to fix all the issues. I ran out time today to transcribe the temperatures for graphing, but this itteration of the code managed to reach 200°C in 10 and 7 minutes for test #1 and #2, respectively. I also managed to fix the problem of the hot plate not turning off after the desired heating time. The real test will be trying a slower heating time, like 20 minutes, but I am glad I postponed using actual samples because this fix has given me code that appears to work exactly as I hoped.

Attachment 1: PWM_v2.1_test1_max_rate_reached_and_peak_heat.jpg
PWM_v2.1_test1_max_rate_reached_and_peak_heat.jpg
  95   Sun Oct 9 21:55:53 2022 Jennifer HritzGeneralOptical ContactingHot plate PWM v2 test #1 & 2

For the following two graphs, I ram four tests: two using the the v1 of the PWM code and two using v2 of the PWM code. The graphs show the heating rate I was aiming for and the actual results. It turns out, my v2 does not work better than my v1. Before 150°C (which is where I believed that (assuming the rate is kept constantly) the heating rate shifted from linear to logarithmic), v1 is an overshoot and v2 is slightly less of an overshoot. The goal of v2 was to increase the rate after 150°C to compensate for this drop off, but it does not appear to have worked.

While I would still like to refine my code, I think it will be good enough to try using it to actually heat the samples.

Attachment 1: PWM_v2_Test_#1_-_increase_180°C_in_10_minutes.pdf
PWM_v2_Test_#1_-_increase_180°C_in_10_minutes.pdf
Attachment 2: PWM_v2_Test_#2_-_increase_180°C_in_7_minutes.pdf
PWM_v2_Test_#2_-_increase_180°C_in_7_minutes.pdf
  94   Sat Oct 8 23:22:25 2022 Jennifer HritzGeneralOptical ContactingHot plate PWM v2 test #1

I had some trouble with the code not working as intended (partially because it has been I while since I coded in C++). However, I was able to run two tests with the new code, although I ran out of time to type up the data for the 2nd. Graphing the 1st test's data, it appears that my improved code is an improvement, but the heating is still slowing down as it approaches 200°C. I need to re-run this test, but with v1 of the code, for better comparison.

The hot plate was supposed to increase 180°C in 10 minutes (so that I would reach 200°C), but due to an inscrutable bug, it did not exit the while loop, so it continued past 10 minutes.

Attachment 1: PWM_v2_Test_#1.pdf
PWM_v2_Test_#1.pdf
  93   Fri Oct 7 21:20:08 2022 Jennifer HritzGeneralOptical ContactingHot plate PWM v2 progress

I had a little set back regarding the non-linear portion of the heating. After about 150°C, if the heating rate is kept constant, the heating graph transitions from linear to logarithmic. I was able to show graphically that, yes, it is indeed logarithmic, but I could not think of an algorithmic way to translate this logarithmic curve into the increase in heating rate to maintain a linear heating rate. I do have some ideas which I will test tomorrow.

Attachment 1: PWM_Test_#2_(log_comparison).pdf
PWM_Test_#2_(log_comparison).pdf
Attachment 2: PWM_Test_#3_(log_comparison).pdf
PWM_Test_#3_(log_comparison).pdf
  92   Fri Oct 7 00:21:24 2022 Jennifer HritzGeneralOptical ContactingHot plate PWM Test #3

The previous test was cycled with 0.3s on follwed by 0.7s off*. This test was 0.7s on followed by 0.3s off. I intended to let it run longer, but I accidetly knocked the thermocouple over while trying to move the cable father from the hot plate so the plastic would not risk melting.

Like before, we see that it starts out relatively linear. I noticed the heating light kind of fluttering around 200°C which appeared in the data as a small decrease around 450s on the graph. I do not know the source of this issue, but I fear it may be the hot plate overriding my cycling with its own built-in cycle; something left for future testing. This is the last data I will gather using v1 of my Arduino code, as am I now working on implementing what I have learned in a smarter v2 of the code. I included v1 of the code, and the txt files for the first three tests.

*I think. Could have been 0.1 on, 0.9 off. Note to self: double check this.

Attachment 1: PWM_Test_#3.pdf
PWM_Test_#3.pdf
Attachment 2: PWM_Test_#3_(comparison).pdf
PWM_Test_#3_(comparison).pdf
Attachment 3: hot_plate_cycler_v1.zip
Attachment 4: PWM_tests_1-3.zip
  91   Wed Oct 5 23:24:08 2022 Jennifer HritzGeneralOptical ContactingHot plate PWM Test #2

I repeated the first test, but let the hot plate run longer. It revealed that the linearity for the lower temperatures completely falls apart at the higher temperatures. I think it should be fairly straightforward to modify the code to accommodate this.

Attachment 1: PWM_Test_#2.pdf
PWM_Test_#2.pdf
Attachment 2: PWM_Test_#2_(comparison).pdf
PWM_Test_#2_(comparison).pdf
  90   Tue Oct 4 22:15:23 2022 Jennifer HritzGeneralOptical ContactingHot plate PWM Test #1

I wrote a program to control the heating rate of the hot plate using Pulse Width Modulation (PWM), and it was a great success!

For roughly 6 minutes, the hot plate was power cycled with a rate of 100 ms on followed by 900 ms off. Based on my calculations, this should correspond to a 0.08°C/sec temperature increase. In other terms, we expect a 24°C increase in the span of 5 minutes. For comparision, without PWM, the hot plate heats up roughly 100°C in that same timespan. I recorded the temperature by filming a thermometer and transcribing that video into a text file, which could be analyzed and graphed. I only transcribed the first 5 minutes of the 17 minute video (I also filmed part of the cool down) because 5 minutes was enough to show clear results.

At t=0, the hot plate was 21.4°C, and at t=300, the hot plate was 49.7°C. That is a 28.3°C increase in the span of 5 minutes, only 4.3°C higher than the predicted value. The rate, 0.094°C/sec, is only slightly faster than the desired 0.08°C/sec. Further, as shown in the graph, the temperature increase was almost perfectly linear, which is ideal. Overall, using an Arduino to PWM the hot plate is looking very promising.

Attachment 1: PWM_Test_#1_(first_300_sec).pdf
PWM_Test_#1_(first_300_sec).pdf
Attachment 2: frame_from_pwm_test1_video_PXL_20221005_064240498_exported_231444.jpg
frame_from_pwm_test1_video_PXL_20221005_064240498_exported_231444.jpg
  89   Mon Oct 3 23:32:30 2022 Jennifer HritzGeneralOptical ContactingCause of improved bond: time or pressure (Update)

I was unable to check the samples because I could not get access to Bridge, so they will be checked tomorrow and the results will be added as an edit to this log.
Given that I was unable to do work in the lab, I instead began a second attempt at writing code for the Arduino to use PWM to control the hot plate temperature.


As expected, the suface area of the bond only increased for the samples under the weights. I did notice something worrying: one of the non-weighted samples actually had its surface area decrease. It is unclear if this is a one-time thing or if all of the bonds deteriorate with time. Unrelated, but I also noticed that the bonded areas always have small dots that refuse to bond. It's unclear if that is due to imperfections or contamination (I suspect the latter).
I left all 4 samples under both weights out of curiosity to see if the bonded surface area would increase further (or possibly decrese further).

Attachment 1: pressure_v_time_samples_both.jpg
pressure_v_time_samples_both.jpg
  88   Sun Oct 2 23:23:07 2022 Jennifer HritzGeneralOptical ContactingCause of improved bond: time or pressure

Before jumping to conclusions based on my previous results, I wanted to check that it was indeed heat and pressure, not time, that led to the bonds improving.

I prepared 4 samples, all with my standard pressing technique (which still leaves room for improvement). 2 samples will simply be left to sit undisturbed, and the other 2 will be left under both (rectangular and cylindrical) weights. I will check these in roughly 24 hours, just like the last test.

The 2 slides on the right are the ones under the weights.

Attachment 1: before_time_on_left_pressure_on_right_PXL_20221003_061828860.MP.jpg
before_time_on_left_pressure_on_right_PXL_20221003_061828860.MP.jpg
Attachment 2: set_up_time_vs_pressure_PXL_20221003_062018249.MP.jpg
set_up_time_vs_pressure_PXL_20221003_062018249.MP.jpg
Attachment 3: in_progress_time_vs_pressure_PXL_20221003_062117715.MP.jpg
in_progress_time_vs_pressure_PXL_20221003_062117715.MP.jpg
  87   Sat Oct 1 23:00:03 2022 Jennifer HritzGeneralOptical Contacting(RESULTS!) Quick test heat and pressure test

These are the results from the previous log.

At long last, there was an improvement with pressure and heat! Pressure without heat and pressure with heat both showed a small improvement. Although the improvement was not major, it does show that pursuing this method of adding weight and heat are viable. Since this was a test, I put less weight on and heated it fast than intended, but now I feel confident to add more weight and slower/greater amounts of heat.
 

Attachment 1: samples_before_AND_after_pressure_PXL_20221002_054158446.MP.jpg
samples_before_AND_after_pressure_PXL_20221002_054158446.MP.jpg
Attachment 2: samples_before_AND_after_heat_w_pressure_PXL_20221002_053842185.MP.jpg
samples_before_AND_after_heat_w_pressure_PXL_20221002_053842185.MP.jpg
  86   Fri Sep 30 21:51:31 2022 Jennifer HritzGeneralOptical ContactingQuick test heat and pressure test

Given that these glass slides are much thinner than the ones I worked with prior, I suspected they would be more receptive to pressure. I decided to replicate the tests I performed with the larger slides: I prepared 8 samples, 4 by smushing the slides together with methanol in the middle and another 4 by cleaning the slides with methanol before pressing them together with my fingers. I put 2 of each type under the cylindrical weight, and 2 of each type under the rectangular weight with the addition of heating. The heating consisted of switching the temperature from off --> low --> med --> high with 15 minutes on each setting.

I will check the results in the morning. I need to wait until the rectangular weight is completely cooled, otherwise I cannot remove it from the hot plate in manner that does not risk cracking the glass.

The first sample picture shows the pressed slides on the top and the smushed slides on the bottom. For the second picture, this is reveresed. Correction: the order is the same for both samples.

Attachment 1: pressure_set_up_PXL_20221001_040539385.MP.jpg
pressure_set_up_PXL_20221001_040539385.MP.jpg
Attachment 2: samples_before_pressure_PXL_20221001_040441662.MP.jpg
samples_before_pressure_PXL_20221001_040441662.MP.jpg
Attachment 3: heat_w_pressure_set_up_PXL_20221001_033228622.MP.jpg
heat_w_pressure_set_up_PXL_20221001_033228622.MP.jpg
Attachment 4: samples_before_heat_w_pressure_PXL_20221001_034620991.MP.jpg
samples_before_heat_w_pressure_PXL_20221001_034620991.MP.jpg
  84   Thu Sep 29 18:36:40 2022 Jennifer HritzGeneralOptical ContactingLooked at Thor Lab slides (continued)

Continuining with my casual exploration of the Thor Lab slides, I heated them from off --> low --> med --> high, with 10 minutes on each setting. The only pressure I applied was 3 larger glass slides, and that was only to flatten out the copper that the smaller, bonded slides sat on top of (so the contact with the heating plate was even).

The heat made the bonded area smaller, but it did not break. As the slides cooled, the bond area increased slightly but not back to the original size. Next I will try this with slower heating and additional pressure.

Attachment 1: before_heat_PXL_20220930_011031420.MP.jpg
before_heat_PXL_20220930_011031420.MP.jpg
Attachment 2: set_up_PXL_20220930_011050114.MP.jpg
set_up_PXL_20220930_011050114.MP.jpg
Attachment 3: right_after_heating_PXL_20220930_014301900.MP.jpg
right_after_heating_PXL_20220930_014301900.MP.jpg
Attachment 4: after_4ish_hours_of_cooling_PXL_20220930_063533904.jpg
after_4ish_hours_of_cooling_PXL_20220930_063533904.jpg
  83   Wed Sep 28 22:11:31 2022 Jennifer HritzGeneralOptical ContactingLooked at Thor Lab slides

While finalizing my work plan for the quarter, I decided to look at the Thor Lab slides. This was instructive because they highlighted the troubles I will have with working with silicone. They are fragile and their small, thin sizes makes cleaning and manipulating them (without contamination) much more difficult compared to the glass sides from before.

I tried cleaning and bonding them the same way as the larger slides. Rubbing them together did not work like with the larger sides, but that may also be a function of being more careful, as not to break them. Once I cleaned them, it only took a tap from my finger to get the center to bond, but the bonded surface area still did not spread out like it did in the YouTube videos (http://youtu.be/se3K_MWR488?t=80). By pressing down around the bonded area, I could expand it slighty. Note that I did crack one slide in the process of doing this, as shown in the pictures.

Because the slides are so thin, I think they will benefit greatly from being left under a heavy object, although it may be difficult to put the weight on the slides without them breaking.

Attachment 1: thor_lab_slides_first_bond_PXL_20220929_045352675.MP.jpg
thor_lab_slides_first_bond_PXL_20220929_045352675.MP.jpg
Attachment 2: thor_lab_slides_after_attempting_to_expand_bond_area_PXL_20220929_045510075.MP.jpg
thor_lab_slides_after_attempting_to_expand_bond_area_PXL_20220929_045510075.MP.jpg
  82   Tue Sep 27 22:29:34 2022 Jennifer HritzGeneralOptical ContactingRazor test apparatus disassembly pictures

Pictures of the razor test apparatus before and after disassembly, to make future reassembly easier.

Attachment 1: razor_before_PXL_20220817_211123631.MP.jpg
razor_before_PXL_20220817_211123631.MP.jpg
Attachment 2: razor_after_PXL_20220911_062818511.MP.jpg
razor_after_PXL_20220911_062818511.MP.jpg
  81   Wed Sep 7 10:42:12 2022 RadhikaGeneralHeat LoadMariner TM Cooldown model

The attached plot (upper) compares the heat load delivered to the test mass from various snout lengths (end to end), as a function of test mass temperature. (At steady state, our point of interest is 123K.) Note that these curves use the original CAD snout radius of 5.08cm (2").

The greatest marginal reduction in heat load comes from increasing the end-to-end snout length to 1m, as concluded in the prevous ELOG. This drops the heat load from just under 0.5W (from snout length 0.5m) to 0.15W. Further increase in snout length to 1.5m drops the heat load to well under 0.1W. After this point, we get diminishing marginal benefit for increase in snout length. 

The effect on the TM cooldown curve can be seen in the lower plot. A snout length of 1m drops the steady-state TM temperature to under 100K. Then, like above, increasing the length to 1.5m makes the next non-negligible impact. 

Attachment 1: SnoutLengthCooldownTM.pdf
SnoutLengthCooldownTM.pdf
  80   Mon Aug 29 15:44:46 2022 RadhikaGeneralHeat LoadMariner TM Cooldown model

Here is a more detailed analysis of varying the length and radius of the snout.

Attachment 1 plots the heat load (W) from the snout opening as a function of temperature, for different combinations of snout length and radius. The model using the CAD snout parameters (length=0.67m end-to-end; radius=5.08cm) results in ~0.3W of heat load at steady state. The plot shows that the largest marginal reduction in heat load is achieved by doubling the length of the snout (green curve), which cuts the heat load by over a factor of 2/3. This validates the choice in snout length used in the previous ELOG entry analysis. The bottom line is that the end-to-end snout length should be on the order of 1 meter, if physically possible.

The next marginal improvement comes from reducing the radius of the snout. Attachment 1 considers reducing the radius by a half in addition to doubling the length (red curve). A snout radius of an inch is quite small and might not be feasible within system constraints, but it would reduce the snout heat load to only 25mW at steady state (along with length doubling). 

The cooldown model resulting from optimizing parameters of the snout (length=1.33m, radius=2.54cm) is shown in Attachment 2. The test mass reaches 123K in ~57hrs - only 2 hours faster than the case where only the snout length is doubled (see previous ELOG entry) - and the test mass reaches steady state at 92K - only 6K colder than in the previous case. This could discourage efforts to reduce the radius of the snout at all, since increasing the length provides the most marginal gains. 

Attachment 1: VaryingSnoutparams.pdf
VaryingSnoutparams.pdf
Attachment 2: MarinerTMCooldown_snout_optimal.pdf
MarinerTMCooldown_snout_optimal.pdf
  79   Fri Aug 26 14:24:57 2022 RadhikaGeneralHeat LoadMariner TM Cooldown model

Here I describe the current radiative cooldown model for a Mariner test mass, using parameters from the most recent CAD model. A diagram of all conductive and radiative links can be seen in Attachment 1. Below are some distilled key points:

1. The source of cooling power is an infinite reservoir at 60K - realistically there will be finite cooling power and the system will be optimized within that constraint.
2. The outer shield surrounds the suspension system and some cooling power can be delivered conductively to the outer shield to hold it at an optimal temperature.
3. The suspension cage has 4 feet that insulate the cage from the table at RT.
4. The cage is composed of vertical beams and bottom and top lids. Radiative view factors from the cage to other components have been loosely estimated.
5. Suspension wires conduct heat from the cage to the upper mass, and from the upper mass to the test mass.
6. The inner shield and snout surround the test mass. Aperature openings in the inner shield (for wires) allow the test mass to radiatively "see" surroundings at ~outer shield T.
7. The snout openings and incident laser power are additional heat loads to the test mass. 

All parameters have been taken from CAD, with the exception of:

1) snout length: originally 0.665m in CAD (end to end), but I doubled it to 1.33m following a discussion in group meeting
2) length of copper bar / conductive cooling pathway: set to the end-to-end length of snout. Though this is a conservative over-estimate
2) thermal conductivity of insulating feet: using 0.25 W/mK
3) radius of aperture in IS for suspension wires: using 1"

Attachment 2 contains the cooldown curves for the system components. With the above assumptions, the test mass takes ~59hrs to reach 123K, and the final steady-state temperature is 96K. (*This was edited - found a bug in previous iteration of code that underestimated the TM cooldown time constant and incorrectly concluded ~36hrs to reach 123K. The figures have been updated accordingly.)

Attachment 3-6 are power budgets for major components: TM, IS, Cage, OS (can produce for UM if there's interest). For each, the top plot shows the total heating and cooling power delivered to the component, and the bottom plot separates the heating into individual heat loads. I'll discuss these below:

- TM: The laser delivers 100mW of heating power to the test mass throughout the cooldown. The next most significant source of heating is snout - this warrants further optimization (see next ELOG). 
- IS: Inevitably the test mass heats the inner shield, but the other heat loads are minimal. Note that the model does not consider radiation from the snout opening to the snout/inner shield walls, and this will be added in soon.
- Cage: The only significant heat load to the cage is conduction from the RT table through the feet. The feet can be made taller, or actively held at a colder temperature.
- OS: I've arbitrarily added conductive cooling to the OS which holds it around 175K. With the current model, adding more cooling power would only help, but in reality this will divert cooling power from going to the IS. This constraint needs to be added in before the optimal OS temperature can be determined. The most significant heat loads are from the chamber walls and the cage (see above).

The next post will describe optimization of the snout length/radius for cooldown.

Attachment 1: Mariner_Heat_Load_Sketch.pdf
Mariner_Heat_Load_Sketch.pdf
Attachment 2: MarinerTMCooldown.pdf
MarinerTMCooldown.pdf
Attachment 3: TMPowerBudget.pdf
TMPowerBudget.pdf
Attachment 4: ISPowerBudget.pdf
ISPowerBudget.pdf
Attachment 5: CagePowerBudget.png
CagePowerBudget.png
Attachment 6: OSPowerBudget.pdf
OSPowerBudget.pdf
  78   Tue Aug 23 17:10:50 2022 PacoGeneralDesign specsNew corner plots

Instead of varying individual layer thicknesses using the MC sampler, I made sure both the thickness and index of refractions are varied as a global systematic error to estimate the design sensitivity. The results for ITM/ETM respectively, with 1e5 samples this time, are in Attachments 1-2 below.

Attachment 1: ITM_corner.pdf
ITM_corner.pdf
Attachment 2: ETM_corner.pdf
ETM_corner.pdf
  77   Tue Aug 16 19:54:29 2022 Jennifer HritzGeneralOptical ContactingRazor blade test

We succeeded in setting up an apparatus for quantifiying the razor blade test. After mounting the glass slides such that the razor edge rested against the gap, we slowly turned the knob to push the blade into the gap. We started with the knob at 0.111, and at 0.757, the bond between the glass slides failed. As we approached 0.757, the interference pattern in the glass shifted, foreshadowing the break.

(Edit by Koji. This 0.757 is 0.0757 I suppose...? And the unit is in inch)

Attachment 1: PXL_20220817_023737796.MP.jpg
PXL_20220817_023737796.MP.jpg
Attachment 2: PXL_20220817_023741865.MP.jpg
PXL_20220817_023741865.MP.jpg
  76   Tue Aug 16 09:58:23 2022 PacoGeneralDesign specsHR coating update

A couple of coating stacks with better tolerance (transmission +- 10%). Attachments #1-2 show the spectral reflectivities for ETM/ITM respectively, while Attachments #3-4 show the corner plots. I think the tolerances are inflated by the fact that all the stack indices and thicknesses are varying, while in reality these two effects are degenerate because what matters is the optical thickness. I will try to reflect this in the MCMC code next. Finally, attachments # 5-6 are the hdf5 files with the optimization results.

Attachment 1: ETM_R_220816_094640..pdf
ETM_R_220816_094640..pdf
Attachment 2: ITM_R_220816_095441..pdf
ITM_R_220816_095441..pdf
Attachment 3: ETM_corner.pdf
ETM_corner.pdf
Attachment 4: ITM_corner.pdf
ITM_corner.pdf
Attachment 5: ETM_Layers_220816_094640.hdf5
Attachment 6: ITM_Layers_220816_095441.hdf5
  75   Mon Aug 15 16:37:51 2022 Juan GeneralGeneralUpdate on Sat Amp and OSEMs

Sat Amp 

- Changes to sat amp 15.8 k ohm resistors instead of 16k The change has been made on Sat Amp - S1103733 & S1103732 ONLY Channel 4 and 2 have been changed on both boards.

OSEM

- I developed a test bed for our OSEM to measure force 

I will attach images of the setup and some of the results from 3 different OSEMs.

Future Work

- For the current test bed, we are using a clear plastic bin although not ideal it manages to get the job done and works for now there could be a better solution for this,
- Next step for OSEM we want to use 40 m single pendulum to test OSEM and measure the transfer function.

Attachment 1: IMG-6458.jpg
IMG-6458.jpg
Attachment 2: IMG-6355.jpg
IMG-6355.jpg
Attachment 3: IMG-6459.jpg
IMG-6459.jpg
Attachment 4: IMG-6460.jpg
IMG-6460.jpg
Attachment 5: OSEMdata.png
OSEMdata.png
  74   Mon Aug 8 13:00:56 2022 JuanGeneralGeneralSat Amp testing of OSEMS

In the following test, a single Sat Amp chassis that holds Sat Amp Board S1106078 and S1106077

Verification of Sat Amp

First, as the test of the LED driver circuits in the chassis, 8 of various color LEDs were inserted to the appropriate output pins of the chassis. This resulted in all the LED lit and the LED mon TP was confirmed to have voltage outputs of 5V. (See my previous ELOG)

OSEM tests

Connected OSEMs to the sat amp to test the OSEM LED/PD pairs. With the first test, the PD out gave us 15V. We wondered if this was just the problem of the OSEM or circuit, or just there are too much light for the transimpedance gain of 121K Ohm.

By blocking the OSEM light by a random heat shrink tube found on the table, we saw the number got reduced. This indicates that the OSEM/Satamp outputs are working and there are just too much light.

We decided to reduce the gain: The transimpedance gain R18 was reduced to 16k, which gave us a voltage range from 5V~7V  with some outlier OSEMS at 1V (See the attached table)

There are 24 total OSEMs:

  • one apparently not functional
  • 20 fell in the range of 5~7V
  • 3 fell in the range of  ~1V

(These numbers given after the change of R18 to 16k Ohm)

Note: We originally aimed for 8~9V. However, from a previous study of OSEM at cryogenic temperature, we learned that there was about an about 30% increase in the response.
Therefore, we decided to leave a sufficient margin from 10V considering this expected increase in the response.

Attachment 1: OSEMs.xlsx
  73   Thu Aug 4 13:44:56 2022 Jennifer HritzGeneralOptical ContactingSuccess with slowly heating

Yesterday, I did two rounds of slowly heating 4 samples to the maximum hot plate temperature. This was to formally test if my success with a single sample earlier in the week was a fluke. Note that the hot plate takes about 10-15 minutes to reach a stable temperature when it is turned up one notch.

First round:
I bonded 4 samples by putting methanol in the gap between the glass slides and letting it dry to create a gap.
Starting at room temperature, I heated the slides on each setting for roughly 15 minutes, then let them cool down naturally over the course of an hour. 3 broke broke at medium heat, and 1 survived the whole process. I belive these broke because the bonds were weaker and I heated them slightly too quickly. In previous tests, I would manually switch the hot plate on and off, but I wanted to see if the hot plate could heat up slow enough on its own.

Second round:
I bonded 4 samples by scrubbing the slides with methanol, using a compressed air duster to blow off the fibers, rubbing them together with the pressure of my fingers, and repeating this whole procedure until they stuck (it took 2-4 repeats).
Starting at room temperature, I heated the slides on each setting for exactly 20 minutes, then let them cool down naturally over the course of an hour. All of them survived to the maximum temperature (the pictures show them at the start and end of the heating, note the temperature). I credit this to the stronger bonding proceedure and the extra 5 minutes for them to adjust to the temperature. I did not turn the hot plate on or off at any point, I just let it heat up at its own rate.

I cannot tell if the bonds are stronger. The size and shape of the Newtons rings did not change.

Attachment 1: PXL_20220803_232203193.jpg
PXL_20220803_232203193.jpg
Attachment 2: PXL_20220804_002433906.jpg
PXL_20220804_002433906.jpg
  72   Thu Aug 4 11:26:55 2022 JuanGeneralGeneralSat Amp

Sat amp seems to be working just fine. There does seem to be a saturation issue with one of the outputs we may need to change a resistor on the board.

 

 

Attachment 1: IMG-6153.jpg
IMG-6153.jpg
  71   Wed Jul 27 14:50:20 2022 Jennifer HritzGeneralOptical ContactingBonding without liquids and narrowing down heating issue

I have found that, after cleaning the glass with methanol (or even sometimes with just a dry lense-cleaning cloth), I can get glass slides to bond by rappidly rubbing them together until something sticks. This was inspired by watching "Wizard of Vaz" perform bonds on YoutTube. While cleaning, I now use enough strength to make the glass squeak, as advised by him.

Upon heating, I encountered the same issue as when I bonded them by putting a liquid (water, methanol, etc.) in the gap, which leads me to now believe that the broken bond is not due to the expansion of a liquid. Further, even at the low temperature of 60°C, placing the liquid-less sample on the hotplate breaks the bond in seconds, which I caught on video. In the attached video*, you can see that, before the heat, the bond is strong enough that I cannot push it appart with my fingers, but after the heat, it slides easily. Note that, outside of taking the video, I always lay the entire slide on the center of the metal so the sample is evenly heated.

*This is my first time attaching a video. If it didn't attach properly, I'll add it on to a later log. I also want to record myself performing the rubbing bonding technique.

Attachment 1: PXL_20220727_214658230.jpg
PXL_20220727_214658230.jpg
Attachment 2: PXL_20220727_214241668.mp4
  70   Tue Jul 26 14:17:44 2022 JuanGeneralGeneralOSEMS actuators

we have 23 OSEMS they look all full built and I will try and test them this week and or next week.

Attachment 1: IMG-6050.jpg
IMG-6050.jpg
Attachment 2: IMG-6045.jpg
IMG-6045.jpg
Attachment 3: IMG-6044.jpg
IMG-6044.jpg
Attachment 4: IMG-6051.jpg
IMG-6051.jpg
Attachment 5: IMG-6049.jpg
IMG-6049.jpg
Attachment 6: IMG-6046.jpg
IMG-6046.jpg
Attachment 7: IMG-6048.jpg
IMG-6048.jpg
Attachment 8: IMG-6047.jpg
IMG-6047.jpg
  69   Fri Jul 22 13:47:47 2022 JuanGeneralGeneralUpdate

Just a general update of what I have been up to deriving Lagrange for double pendulum system and also been looking at code that koji gave to me I've add comment to some of the code also working on my report.

  68   Fri Jul 22 13:36:55 2022 PacoGeneralDesign specsHR coating update

Here are some corner plots to analyze the sensitivity of the designs in the previous elog to a 1% gaussian distributed perturbation using MCMC.

Attachment #1 shows the ETM corner plot 

Attachment #2 shows the ITM corner plot.

I let the indices of both high and low index materials vary, as well as the physical thicknesses and project their covariances to the transmission for PSL and AUX wavelengths.

The result shows that for our designs it is better to undershoot in the optimization stage rather than meet the exact number. Nevertheless, 1% level perturbations in the optical thickness of the stack result in 30% deviations in our target transmission specifications. It would be nice to have a better constraint on how much each parameter is actually varying by, e.g. I don't believe we can't fix the index of refraction to better than 1%., but exactly what its value is I don't know, and what are the layer deposition tolerances? These numbers will make our perturbation analysis more precise.

Attachment 1: ETM_corner.pdf
ETM_corner.pdf
Attachment 2: ITM_corner.pdf
ITM_corner.pdf
  67   Mon Jul 18 18:34:29 2022 PacoGeneralDesign specsHR coating update

I've been running the HR coating optimization for mariner TMs. Relative to the specifications found here we now are aiming for

  • ITM HR coating of 2000 ppm @ 2050.15 nm, and 1000 ppm @ 1550 nm
  • ETM HR coating of 25 ppm @ 2050.15 nm, and 1000 ppm @ 1550 nm.

Both the PSL and AUX cavity finesses range the few couple of thousands, and the goal is not to optimize the coating stack for noise, but more importantly for the transmission values and tolerances. This way we ensure the average finesse and differential finesse requirements are met. Anyways, Attachment #1-2 shows the transmission plots for the optimized coating stacks (so far). Attachments #3-4 show the dielectric stacks. The code still lives in this repository.

I'm on the process of assessing the tolerance of this design stacks against perturbations in the layer thicknesses; to be posted in a follow-up elog.

Attachment 1: ETM_R.pdf
ETM_R.pdf
Attachment 2: ITM_R.pdf
ITM_R.pdf
Attachment 3: ETM_Layers.pdf
ETM_Layers.pdf
Attachment 4: ITM_Layers.pdf
ITM_Layers.pdf
  66   Thu Jul 14 14:55:01 2022 Jennifer HritzGeneralOptical ContactingTesting isopropanol and methanol

Note that I am just testing out different techniques, so I have not set up the thermocouples to precisely measure the temperatue.
On Tuesday, I developed a new method of putting water, isopropanol, or methanol on one slide then squishing the other slide on top of it to fill the gap with the afformentioned liquid. The slides are slippery at first, but as they dried, which took about 15 minutes, the bond forms. The bonds were strong enough that I could just barely push the slides appart by applying pressure to the side using my thumbs. I prepared 4 samples this way, 2 with iso and 2 with meth. I took one of each and heated them on Medium for 30 minutes under the brass hunk with the aluminum square on the bottom and copper foil on both sides of the samples. Earlier in the day, I tried heating them without the weight on top, but the heat just broke the bond. I took the remain two and set them aside as controls.
On Thursday, I returned to check the bonds. The heated samples had broken. I intented to check on Wednesday, but I was sick from food poisoning, so I do not know whether the bonds broke immediately after heating or due to sitting for an extra day. For the control samples, one also had a broken bond, but the other had become even stronger.
I noticed that, when the slides are successfully bonded, the shape and appearance of the Newton's rings change, which can be seen in the pictures. I speculate that the circles on the unbroken control are the bonded regions. Ideally, we want to see no Newton's rings.

Attachment 1: PXL_20220714_220953206.MP_2.jpg
PXL_20220714_220953206.MP_2.jpg
Attachment 2: PXL_20220714_220940258.MP_2.jpg
PXL_20220714_220940258.MP_2.jpg
Attachment 3: PXL_20220714_222105409.jpg
PXL_20220714_222105409.jpg
Attachment 4: PXL_20220713_003923957.jpg
PXL_20220713_003923957.jpg
ELOG V3.1.3-