Message ID: 14     Entry time: Fri May 7 17:50:31 2021
Author: Nina Vaidya & Shruti Maliakal 
Type: General 
Category: Design specs 
Subject: Arm Cavity Design 2021 update 

Here are the final slides with all the results on the Arm Cavity Design, please review. 

For RoC of 56.2 +/- 1% things are working well. Tolerance of 0.5% will be better however, 1% is still working; as long as we do not want any peaks ~50kHz away.

For length, 38+0.5% = 38.19 (with RoC 56.2) not ideal, peak is close (48.8kHz) but maybe ok? @Rana thoughts? and 38-0.5% = 37.81 (with RoC 56.2) works well.

To summarise the design:

RoC = 56.2 +/- 1%

L = 38 +/- 0.5%

Attachment 1: Arm_Cavity_Design_05072021_with_tolerances.pptx  37.538 MB  Uploaded Fri May 7 18:57:05 2021
Attachment 2: HOMhelper.py  2 kB  Uploaded Fri May 7 18:58:56 2021  | Show | Show all
Attachment 3: Arm_HOManalysis.ipynb  617 kB  Uploaded Fri May 7 18:59:57 2021  | Show | Show all
Attachment 4: HOMplot.py  5 kB  Uploaded Fri May 7 19:00:10 2021  | Show | Show all