40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  Engineering group Lab elog  Not logged in ELOG logo
Message ID: 226     Entry time: Fri Dec 13 13:06:36 2019
Author: Jordan  
Type: Progress 
Category: Clean and Bake 
Subject: Small ABO Controller Upgrade 

Below are the temperature v. time plots from the new Platinum series contoller/software. The profile ran was a 2.5 hour ramp to 94°C, 2.5 hour soak, and 2.5 hour ramp down to room temp. The controller had a maximum overshoot of ~4°C. The Extech datalogger was also used to track temperature, with a maximum overshoot of ~ 5°C.

Attachment 1: Controller.png  22 kB  | Hide | Hide all
Controller.png
Attachment 2: Datalogger.png  15 kB  | Hide | Hide all
Datalogger.png
Attachment 3: Profile_2-test4.xlsx  439 kB