Doesn't the phase meter just read out the noise even with no locking? I thought that was going to be the magic.
For locking, the mixer readout is in units of phase and the laser current modulation gives a proportional frequency modulation with no frequency wiggles until > 1 MHz. So it should phase lock with no integrator, but I'm not sure if the free running noise will drive it out of the phase lock or not. I wonder if its possible to use the phase meter as an error signal. It would be much easier to lock frequency instead of phase via a mixer.
|