I measured the transmission of the Coastline 1m mirror at 180. ppm (S122C).
Alignment procedure while setting location of optics:
- use 2x irises to constrain a beam path at the locations of the eventual MZ input BS and the cavity BS.
- Place the first (R~1) flat mirror at 45 degrees centered on the beam line. Use this mirror to steer into an iris at the location of the cavity's curved mirror.
- Use steering optics to direct the eventual transmission beam into the trans mon PD (PDA 20CS)
- Place the curved mirror at near normal incidence. Use this mirror to steer the beam through irises at the location of the MZ output BS and cavity BS.
- Use steering mirrors to send the beam to the refl PD (PDA255).
- Place the cavity BS and steer the reflected beam to the refl PD.
Alignment procedure subsequently:
- Misalign the cavity BS
- Use the cavity mirrors to steer the circulated reflection into refl PD.
- Use cavity BS to align prompt reflection into refl PD.
|