40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  Cryo Lab eLog  Not logged in ELOG logo
Entry  Mon Nov 3 15:16:28 2014, Dmass, Noise Hunting, Noise Budget, New beat spectrum: ErrSigs.pngBeatNz.png
    Reply  Mon Nov 3 16:15:21 2014, Dmass, Noise Hunting, Noise Budget, Beat spectrum not ltd by  
       Reply  Mon Nov 3 20:27:28 2014, Dmass, Noise Hunting, Noise Budget, ISS servo fail? 
          Reply  Wed Nov 19 14:54:36 2014, nicolas, Noise Hunting, Noise Budget, ISS servo fail?  
       Reply  Tue Dec 2 22:51:52 2014, Dmass, Noise Hunting, Noise Budget, Beat spectrum not ltd by  PDH_demod_chain.png
          Reply  Sat Sep 12 14:25:05 2015, Dmass, Noise Hunting, Noise Budget, Beat spectrum not ltd by  
    Reply  Mon Nov 3 20:36:14 2014, Dmass, Noise Hunting, Noise Budget, New beat spectrum: 
       Reply  Tue Nov 25 22:45:43 2014, Dmass, Noise Hunting, Noise Budget, New beat spectrum: 
          Reply  Wed Dec 3 17:02:31 2014, Dmass, Noise Hunting, Noise Budget, Noise budget cleaning up 
             Reply  Thu Dec 4 02:45:49 2014, Dmass, Noise Hunting, Noise Budget, Noise budget cleaning up PDH_sensing_nz.pngLB1005inputNoise.png
                Reply  Sun Dec 7 12:42:55 2014, rana, Noise Hunting, Noise Budget, Scatter Noise hump is RAM 
Message ID: 1175     Entry time: Thu Dec 4 02:45:49 2014     In reply to: 1174     Reply to this: 1178
Author: Dmass 
Type: Noise Hunting 
Category: Noise Budget 
Subject: Noise budget cleaning up 

PDH sensing noise:

PDH_sensing_nz.png

New lower sensing noise with addition of transformers between LO and RF in both paths.

Cavity unlocked (detuned with temp) noise shows the scatter bump - unsure if this is interesting or should be totally obvious

 

 

 

LBservo noise:

LB1005inputNoise.png

 

The LB input noise gets to a minimum of 40 nV/rtHz if we turn the gain 100% up at high freq. This is 4x over what they claim

(shown it is 50 nV/rtHz)

 

Data is all on svn and will be put into noisebudget update