Entry  Mon May 21 02:38:48 2012, tara, Notes, NoiseBudget, thermo optic noise added to noise budget nb_short_cav.png
    Reply  Mon May 21 21:11:16 2012, tara, Notes, NoiseBudget, thermo optic noise added to noise budget beat_2012_05_18_compare_lid.png
Message ID: 970     Entry time: Mon May 21 21:11:16 2012     In reply to: 968
Author: tara 
Type: Notes 
Category: NoiseBudget 
Subject: thermo optic noise added to noise budget 

Frank corrected me that the acoustic coupling was not improved with the legs changing around 500 Hz. The measurement in comparison was not equipped with acoustic box yet. So I compared the measurement with the result from psl:955 , and there is no improvement around acoustic frequency. This means that these peaks are really from acoustic coupling through air, not legs.

beat_2012_05_18_compare_lid.png