Message ID: 858     Entry time: Fri Mar 2 23:37:36 2012     Reply to this: 860   863
Author: tara 
Type: DailyProgress 
Category: BEAT 
Subject: beat is back 

 Beat signal is back now, but we have not measured the spectrum yet since the temperature is still drifting fast (10kHz/ min). We plan to measure it tomorrow afternoon.

    The optics for RCAV are aligned. The visibility is ~95% without adjusting the lenses for mode-matching. I do not redo the mode match yet because we probably have to move the chamber for installing the air springs very soon. For ACAV, the beam is aligned, with visibility only ~ 80%. I also adjusted the cable length for PDH locking so that the error signal looks symmetric (Frank added the 4-ch splitter for demodulating RFAM, so the phase shift changed a bit.  All reflected beams  are properly blocked with razor blade blocks.

    The beat frequency is ~ 188MHz instead of the nominal value of 160 MHz because the temperature is not settled at the set point yet. It takes longer than before because of the copper shields around the cavities.  We expect it to be stable by tomorrow afternoon.