Today we floated the table, measured beat, measured TF between beat and horizontal, and added horizontal seismic to the noise budget.
- Replaced 2 legs, floated table: monitor pressure, 2000psi @ Wed Oct 26 18:06:41 2011
- Measure TF between beat and accelerometer signal
- measure beat signal, when table is floated, optimized optics.
- add Horizontal motion to noise budget
- ring down measurements from rubber spring, see psl:717
details coming soon.

fig1: accelerometer was mounted on the table to measure East-West motion (along beam line)

fig2: details of the accelerometer

fig3: TF between Beat signal and signal from acceleromter (East-West motion). The data is not calibrated yet. Sensitivity for beat = 7 kHz/V.

fig4: Current beat noise: floated table. Seismic from horizontal motion is calculated by applying the measured TF to the measured acceleration on the floated table.

fig5: Current beat noise: floated table. Seismic from horizontal motion is calculated by applying the measured TF to the measured seismic noise (converted to acceleration) on the floated table. |