Message ID: 679     Entry time: Fri Sep 16 22:15:03 2011
Author: Frank 
Type: Summary 
Category: Foam 
Subject: thermal insulation added 

when playing with the setup i've noticed that we are actually very sensitive to temperature fluctuations (more than i expected), so the absolute length (or CTE?) must be more different than we expected. So i've added the thermal insulation, drilled four new holes in the end caps, connected the temp sensors etc. Thermal control loop is running now. Made some changes to channels (mostly range limiuts, warning levels etc) and added four new software channels which i will use to generate absolute calibrated temp signals. The difference between the sensors is about 1K. I know it does not matter for the stabilization but i hate it as you never know if you have a gradient or degraded sensor or so...

I'ver started heating it up to 26.5C (you may ask why this number: laziness -  actually i've started the heater before finishing the insulation and that was the actual temperature at the time i finished it). The LO for the AOM already increased to 70.6MHz (from 64MHz). So my guess is that once we reach 30C we are back in range for the iLIGO VCO.

Will change the temperature over the weekend in a couple of steps to measure the temperature sensitivity  (differential length change with temp). This will tell us how good the thermal stability has to be to not be limited by thermal fluctuations. We then can make a decision on further things like second, external box, passive foam or metal box with active control etc..

ELOG V3.1.3-