Entry  Wed May 18 20:51:35 2011, tara, DailyProgress, optic, optic layout for new fss setup fss_layout_2011_05_18.png
    Reply  Thu May 19 11:32:46 2011, tara, DailyProgress, optic, optic layout for new fss setup fss_2011_05_18.png
       Reply  Tue Jun 14 23:07:03 2011, tara, DailyProgress, optic, optic layout for new fss setup 2aom_fss.png
          Reply  Mon Jul 18 02:20:01 2011, tara, DailyProgress, optic, optic layout for new fss setup 2aom_fss_2011-07-15.png
             Reply  Mon Jul 18 11:41:22 2011, rana, DailyProgress, optic, optic layout for new fss setup 
                Reply  Wed Jul 20 15:54:21 2011, tara, DailyProgress, optic, optic layout for new fss setup 2011_07_20_layout.png
                   Reply  Wed Jul 20 23:00:30 2011, Frank, DailyProgress, optic, optic layout for new fss setup 
Message ID: 631     Entry time: Wed Jul 20 15:54:21 2011     In reply to: 629     Reply to this: 632
Author: tara 
Type: DailyProgress 
Category: optic 
Subject: optic layout for new fss setup 

  I edited the layout so that the spots in both AOMs are 200 um. I'll list what optics we might have to buy.

 

Most of the optics are already used on the table. I need to find:

  •  a lens with f = 343.6 m (plcx R =154.5mm)
  •  one more curve mirror with R = 0.3m for the second AOM.
  • aom adaptor plate (need to submit this to the work shop to have it done
  • periscope sets for both ACAV and RCAV (we need 4 in total, but we have only 2 sets)
  • second VCO

The optics on ACAV path have been removed, I left the optics on RCAV path for now because Raphael might want to remeasure EOM TF.

Once the measurement is done, all optics will be removed. We will clean the table, clean the optics before put them back on the table.

 

 

 

2011_07_20_layout.png