40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  PSL  Not logged in ELOG logo
Entry  Tue Mar 29 22:15:25 2011, tara, Notes, Electronics Equipment, RFPD bad connector rcavPD.pngacavPD.pngnoise_raw.pngRFPDnoise.png
    Reply  Wed Mar 30 20:23:58 2011, tara, Notes, Electronics Equipment, RFPD noise comparison D980454-00.pdfRFPDnoise.pngrfpd_psl.fil
       Reply  Thu Mar 31 19:37:54 2011, rana, Notes, Electronics Equipment, RFPD noise comparison 
          Reply  Fri Apr 1 02:18:13 2011, tara, Notes, Electronics Equipment, RFPD noise comparison 
Message ID: 558     Entry time: Thu Mar 31 19:37:54 2011     In reply to: 557     Reply to this: 561
Author: rana 
Type: Notes 
Category: Electronics Equipment 
Subject: RFPD noise comparison 

 Its basically OK, but you have to use 10 Ohms for the diode resistance instead of 90 Ohms. 

This version of RFPD is so old and bad, that we have to be a little careful - you have to measure the response with the Jenne laser to make sure its OK and tuned right. Because of the PD bias feedback, the capacitance can be mistuned for low or high light levels.

I suggest that you use Acrobat or Illustrator soon and mark up this PDF to reflect what you actually have in the box.

ELOG V3.1.3-