40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  PSL  Not logged in ELOG logo
Entry  Sun Mar 6 23:47:17 2011, tara, Notes, BEAT, boths cavities are locked 
    Reply  Mon Mar 7 01:26:37 2011, Frank, Notes, BEAT, boths cavities are locked 
       Reply  Mon Mar 7 20:13:58 2011, koji, frank, tara, Notes, BEAT, boths cavities are locked beat_shake.png
Message ID: 538     Entry time: Mon Mar 7 01:26:37 2011     In reply to: 537     Reply to this: 539
Author: Frank 
Type: Notes 
Category: BEAT 
Subject: boths cavities are locked 

great, so we can check how sensitive the beat signal scatter bump is to cavity motion if we excite individual eigenmodes of both cavities before we change the suspension. This (hopefully) gives us information about which motion we are most sensitive at the moment so we can optimize the "new" suspension for that.

Quote:

I lock both cavities, setpoints are.

ACAV 37.0 C

RCAV 34.85 C

 

ELOG V3.1.3-