I realigned the beam to PMC, ACAV, RCAV, optimized gain, and find a significant coherence between ACAV_RCTRASPD and RCAV_TRANSPD.
I haven't re-aligned the beam to each cavities for awhile, and the alignment was quite bad.
PMC_RCTRANSPD: 9.4 - > 10.9 V
RCAV_RCTRANSPD: 1.7 ->2.07 V
ACAV_RCTRANSPD: 0.8 -> 1.3 V
So I need to optimize the gain setup again, see detail below.
I measured beat signal and coherence before and after realigning.
For coherence, I see nothing significant except <beat | ACAV>, see fig1, so I did not save the rest of the measurement.
After I realigned the beam, there is a big coherence between ACAV andRCAV see fig 2.
the coherence between PMC and RCAV follow the same trait as that of PMC and ACAV, but slightly less, so I show only PMC and ACAV.
However, the beat note before and after realigning the beam are still the same, see fig3.
I'll add RIN from RCAV/ACAV/PMC .
-----------------
gain setup
--------------
PMC: Gain 16
RF_ADJ 5.7 V
FSS: Common gain 21.5
FAST gain 21.2
RF_ADJ 10
These values allow maximum Transmission power and stable lock (I checked this by unlock and lock the cavity and see if the signal is stable after loss lock) |