Message ID: 324
Entry time: Tue Aug 31 15:04:51 2010
In reply to: 322
Reply to this: 330
|
Author: |
Frank |
Type: |
Notes |
Category: |
Computers |
Subject: |
dead channel, C3:PSL-ACAV_TEMPAVG |
|
|
be carefull with the data you are taking right now. it's wrong for your projection as the power fluctuations are different when locking only the ACAV using the AOM. The largest contributor might be the pointing from the AOM itself, which is different if the laser isn't locked to the other cavityat the same time.
Why don't you use the new fast channels you have hooked up last week? And don't forget to change the names of those :-)
Quote: |
Quote: |
looks like some fault of the database. /usr1/epics/psl/db/acav.db does not contain the correct entry. check rcav.db and copy the record for "C3:PSL-RCAV_TEMPAVG" into acav.db. Then simply change "RCAV" into "ACAV" everywhere for this record. Also change the setpoint for the ACAV in the startup.cmd file to 37.3 and the ACAV-heater value to 4.653. Those are the latest values when both where locked for several hours. Reset the crate again.
Quote: |
Yesterday, I reset the PSL crate behind the SUN computer, but the channel C3:PSL-ACAV_TEMPAVG is stil inactive.
|
|
Will do in a moment, I'm taking data from ACAV for now just to compare with yesterday results.
|
|