both cavities @35 degrees +/- 5mK
slow actuator values
RCAV : 0.1890
ACAV : 0.1993
(.1890 V - .1993V) * 1180 MHz/V = 12 MHz difference
- temperature of one cavity has to be tuned about ~12MHz with 6mK/MHz, so ~70mK
- loosing lock @~+3V input to the VCO which should go up to at least 5V
anyway, the tuning range is only 10MHz in that direction, so can't lock both at the same time right now
frequency see here: VCO tuning |