Entry  Mon Nov 23 10:58:46 2020, anchal, Notes, Laser, North laser not switching on, power supply display not working PXL_20201123_184931888.mp4
    Reply  Mon Nov 23 14:56:59 2020, Koji, Notes, Laser, North laser not switching on, power supply display not working 
       Reply  Mon Nov 30 10:28:08 2020, Anchal, Notes, Laser, North laser not switching on, power supply display not working image-ebd6ed74-fed3-4890-b45d-07d36e3b3000.jpg
          Reply  Mon Nov 30 21:31:56 2020, Koji, Notes, Laser, North laser not switching on, power supply display not working 
Message ID: 2595     Entry time: Mon Nov 23 14:56:59 2020     In reply to: 2594     Reply to this: 2599
Author: Koji 
Type: Notes 
Category: Laser 
Subject: North laser not switching on, power supply display not working 

I suspected the interlock failure. Can you replace the interlocking wire with a piece of wire for the troubleshooting?

There probably is a spare LWE NPRO controller at the 40m.