40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  PSL  Not logged in ELOG logo
Entry  Wed May 13 18:07:32 2020, anchal, DailyProgress, NoiseBudget, Bayesian Analysis CTN_Bayesian_Inference_Analysis_Of_Best_Result.pdf
    Reply  Fri May 15 12:09:17 2020, aaron, DailyProgress, NoiseBudget, Bayesian Analysis 
       Reply  Fri May 15 16:50:24 2020, anchal, DailyProgress, NoiseBudget, Bayesian Analysis 
          Reply  Fri May 22 17:22:37 2020, anchal, DailyProgress, NoiseBudget, Bayesian Analysis CTN_Bayesian_Inference_Analysis_Of_Best_Result.pdf
             Reply  Mon May 25 08:54:26 2020, anchal, DailyProgress, NoiseBudget, Bayesian Analysis with Hard Ceiling Condition CTN_Bayesian_Inference_Analysis_Of_Best_Result_Hard_Ceiling.pdf
                Reply  Tue May 26 15:45:18 2020, anchal, DailyProgress, NoiseBudget, Bayesian Analysis CTN_Bayesian_Inference_Analysis_Of_Best_Result.pdfCTN_Bayesian_Inference_Analysis_Of_Best_Result_Hard_Ceiling.pdf
                   Reply  Thu May 28 14:13:53 2020, anchal, DailyProgress, NoiseBudget, Bayesian Analysis CTN_Bayesian_Inference_Analysis_Of_Best_Result_New.pdf
                      Reply  Sun May 31 11:44:20 2020, Anchal, DailyProgress, NoiseBudget, Bayesian Analysis Finalized CTN_Bayesian_Inference_Final_Analysis.pdf
                         Reply  Mon Jun 1 11:09:09 2020, rana, DailyProgress, NoiseBudget, Bayesian Analysis Finalized 
                         Reply  Thu Jun 4 09:18:04 2020, Anchal, DailyProgress, NoiseBudget, Bayesian Analysis Finalized CTN_Bayesian_Inference_Final_Analysis.pdf
                            Reply  Thu Jun 11 14:02:26 2020, Anchal, DailyProgress, NoiseBudget, Bayesian Analysis Finalized CTN_Bayesian_Inference_Final_Analysis.pdf
                               Reply  Mon Jun 15 16:43:58 2020, Anchal, DailyProgress, NoiseBudget, Better measurement on June 14th CTN_Bayesian_Inference_Final_Analysis.pdf
                                  Reply  Tue Jun 23 17:28:36 2020, Anchal, DailyProgress, NoiseBudget, Better measurement on June 22nd (as I turned 26!) CTN_Best_Measurement_Result.pdf
                                     Reply  Wed Jun 24 21:14:58 2020, Anchal, DailyProgress, NoiseBudget, Better measurement on June 24th 
                               Reply  Fri Jun 26 12:38:34 2020, Anchal, DailyProgress, NoiseBudget, Bayesian Analysis Finalized, Adding Slope of Bulk Loss Angle as variable CTN_Bayesian_Inference_Final_Analysis_with_Slope.pdf
Message ID: 2580     Entry time: Thu Jun 4 09:18:04 2020     In reply to: 2578     Reply to this: 2582
Author: Anchal 
Type: DailyProgress 
Category: NoiseBudget 
Subject: Bayesian Analysis Finalized 

Better measurement captured today.


Final result of CTN experiment as of June 4th 9 am:

\huge \Phi_B = (8.74 \pm 1.51) \times 10^{-4}\quad,rad

with shear loss angle taken from Penn et al. which is 5.2 x 10-7. The limits are 90% confidence interval.

The analysis is attached. This result will be displayed in upcoming DAMOP conference and would be updated in paper if any lower measurement is made.


Adding Effecting Coating Loss Angle (Edit Fri Jun 5 18:23:32 2020 ):

If all layers have an effective coating loss angle, then using gwinc's calculation (Yam et al. Eq.1), we would have effective coating loss angle of:

\huge \Phi_\text{c} = (4.95 \pm 0.86) \times 10^{-4}

This is worse than both Tantala (3.6e-4) and Silica (0.4e-4) currently in use at AdvLIGO.

Also, I'm unsure now if our definition of Bulk and Shear loss angle is truly same as the definitions of Penn et al. because they seem to get an order of magnitude lower coating loss angle from their bulk loss angle.


Analysis Code

Attachment 1: CTN_Bayesian_Inference_Final_Analysis.pdf  233 kB  Uploaded Thu Jun 4 10:20:47 2020  | Show | Show all
ELOG V3.1.3-