40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  PSL  Not logged in ELOG logo
Entry  Fri Feb 14 17:40:27 2020, anchal, DailyProgress, ISS, Installed ISS on both paths using SR560s IMG_20200214_174127.jpgISS_Effect.pdf
    Reply  Sat Feb 15 18:57:43 2020, rana, DailyProgress, ISS, Installed ISS on both paths using SR560s 
       Reply  Thu Feb 20 17:42:23 2020, rana, DailyProgress, ISS, Installed New ISS on both paths using SR560s ISS_New_Analysis.pdf
          Reply  Sun Feb 23 18:19:32 2020, rana, DailyProgress, ISS, Installed New ISS on both paths using SR560s 
             Reply  Fri Feb 28 14:00:11 2020, anchal, DailyProgress, ISS, Installed New ISS on both paths using SR560s - PROBLEM 
                Reply  Sun Mar 1 16:34:31 2020, rana, DailyProgress, ISS, Installed New ISS on both paths using SR560s - PROBLEM 
                   Reply  Mon Mar 2 16:48:12 2020, anchal, DailyProgress, ISS, Follow-up, porbalem stopped happening, OLTF plot. NISS_OLTF.pdf
Message ID: 2542     Entry time: Thu Feb 20 17:42:23 2020     In reply to: 2539     Reply to this: 2544
Author: rana 
Type: DailyProgress 
Category: ISS 
Subject: Installed New ISS on both paths using SR560s 

I've installed new ISS consisting of one SR560 per path.


ISS SR560 Details:

Path Input Coupling Invert High Pass Roll off (dB/octave) High Pass Corner (Hz) Low Pass Roll off (dB/octave) Low Pass Corner (Hz) Gain Mode Gain Output Impedance (Ohms)
North AC No 6 300 6 300 Low Noise 5x103 600
South AC Yes 6 300 6 300 Low Noise 1x104 600

Measurements:

  • I took transfer function by exciting at port B and using mode A-B on the SR560.
  • The transfer function was measured as Actuation Signal (output of SR560) over Error Signal (input to SR560).
  • This transfer function should be equal to open-loop gain over the SR560 transfer function.
  • So, I took another measurement of the transfer function of SR560 itself not connected to the loop.
  • This gave me total open-loop transfer function.
  • I used \frac{\pi}{2V_{\pi}} as the transfer function of EOAM in units of relative intensity change per V (1/V).
  • I used the coupling factor (See CTN:2528) and DC power level of the laser after transmission to get the detector transfer function in units of V.
  • Using the above, I calculated the plant transfer function for each path and then used that to calculate the closed-loop transfer function.
  • To compare, I also calculated noise suppression before and after switching on ISS.
  • These two separate methods match very well. At 300 Hz, intensity noise is suppressed by 40 dB.
  • There is no other witness detector after transmission to use as an out-of-loop photodiode, neither is there any space to put a new one.
  • There is oscillation happening at UGF  fo around 40 kHz in both paths. I can work on it if this looks like a problem in the final beatnote or if it seems to saturate FSS loops. Currently, I saw nothing alarming.

Data

Attachment 1: ISS_New_Analysis.pdf  99 kB  | Hide | Hide all
ISS_New_Analysis.pdf ISS_New_Analysis.pdf ISS_New_Analysis.pdf
ELOG V3.1.3-