40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  PSL  Not logged in ELOG logo
Entry  Fri Nov 1 17:54:47 2019, anchal, DailyProgress, FSS, FSS Diagnostics - RFPD RF Ouput under inspection NFSS_RFPD_Output_Oscilloscope.pdfSFSS_RFPD_Output_Oscilloscope.pdfNFSS_RFPD_RF_OUT_COUPLED.pdfSFSS_RFPD_RF_OUT_COUPLED.pdf
    Reply  Tue Nov 5 18:37:59 2019, anchal, DailyProgress, FSS, FSS Diagnostics - TTFSS Testpoint Spectrum Data FSS_TP_Spectrum.pdf
       Reply  Sat Nov 9 19:28:36 2019, anchal, DailyProgress, FSS, FSS Diagnostics - Loop nonlinearity Test NFSS_NonLinearity_Test.pngSFSS_NonLinearity_Test.pngsignal-attachment-2019-11-09-195635.jpeg
          Reply  Mon Nov 11 20:42:49 2019, anchal, DailyProgress, FSS, FSS Diagnostics - Two Tone Test NFSS_RFPD_SN009_TF_11-11-2019_183131.pdfSFSS_RFPD_SN010_TF_11-11-2019_203310.pdfFSStwoToneTest.pdf
             Reply  Tue Nov 12 18:36:59 2019, anchal, DailyProgress, FSS, FSS Diagnostics - Two Tone Third Order Intermodulation Test FSStwoToneThirdOrderIM.pdf
                Reply  Thu Nov 14 12:19:32 2019, rana, DailyProgress, FSS, FSS Diagnostics - Two Tone Third Order Intermodulation Test 
                   Reply  Tue Nov 19 18:51:02 2019, anchal, DailyProgress, FSS, FSS Diagnostics - Second-order Distortion Calculation FSSDistortionCalc.pdfFSSDistortion.zip
                      Reply  Mon Dec 9 19:54:44 2019, anchal, DailyProgress, FSS, FSS Diagnostics - How much distortion affects the functioning of PDH? 
       Reply  Wed Nov 27 13:59:24 2019, anchal, DailyProgress, FSS, FSS Diagnostics - TTFSS TP1 TimeSeries NFSS_TP1_Oscilloscope.pdf
          Reply  Wed Nov 27 15:17:51 2019, anchal, DailyProgress, FSS, FSS Diagnostics - Quick HF OLTF of NFSS NFSS_OLTF_HF.pdf
             Reply  Wed Nov 27 16:31:20 2019, rana, DailyProgress, FSS, FSS Diagnostics - Quick HF OLTF of NFSS 
                Reply  Wed Dec 4 18:11:09 2019, anchal, DailyProgress, FSS, Free running laser frequency noise spectrum PMCErrorSignals.pdf
                   Reply  Thu Dec 5 18:25:11 2019, rana, DailyProgress, FSS, Free running laser frequency noise spectrum 
                   Reply  Tue Dec 17 18:29:58 2019, anchal, DailyProgress, FSS, Need for ISS? 
                      Reply  Wed Dec 18 18:40:32 2019, anchal, DailyProgress, FSS, Maybe the NPRO Laser isn't bad after all NorthLaserRINAnalysis.pdf
                         Reply  Wed Dec 18 22:03:17 2019, rana, DailyProgress, FSS, Maybe the NPRO Laser isn't bad after all 
                            Reply  Fri Dec 20 15:00:30 2019, anchal, DailyProgress, FSS, PZT, EOM to RIN TF Measurement NorthLaserPZTEOMtoRINTF.pdf
                               Reply  Fri Dec 20 18:08:44 2019, anchal, DailyProgress, FSS, PZT, EOM to RIN TF Measurement Through FSS NorthLaserPZTEOMtoRINTFThruFSS.pdf
                                  Reply  Tue Jan 7 17:59:20 2020, anchal, DailyProgress, FSS, NPMC TF NPMC_TF.pdf
       Reply  Thu Dec 12 19:15:55 2019, anchal, DailyProgress, FSS, FSS Diagnostics - Beatnote Spectrum wrt Laser Power and Modulation Index BeatnoteSpectrumwrtPowerandModindex.pdf
       Reply  Tue Dec 17 16:55:07 2019, anchal, DailyProgress, FSS, FSS Diagnostics - TTFSS North U7 Boost Stage Input-Output TimeSeries NFSS_UT_TS_Analysis.pdf
    Reply  Tue Nov 12 11:39:48 2019, anchal, DailyProgress, FSS, FSS Diagnostics - RFPD RF Ouput under inspection 
    Reply  Fri Dec 13 16:24:02 2019, anchal, DailyProgress, FSS, FSS Diagnostics - RFPD Circuit Latest Available Photos SN009.jpgSN010.jpgSN009.jpgSN010.jpg
    Reply  Fri Dec 13 17:44:05 2019, anchal, DailyProgress, FSS, FSS Diagnostics - RFPD RF Coupled Output Offset 
       Reply  Mon Jan 13 19:04:38 2020, anchal, DailyProgress, FSS, FSS Diagnostics - SN009 (North RFPD) Notch Improvement Attempt SN009.jpgSN009_D980454-00-C.pdfSN009_Notch_Improvement_Attempt.pdf
          Reply  Tue Jan 14 18:19:33 2020, anchal, DailyProgress, FSS, FSS Diagnostics - SN009 (North RFPD) Notch Improved! SN009.jpgSN009_D980454-00-C.pdfSN009_Notch_Improvement_Attempt.pdf
             Reply  Thu Jan 16 11:17:15 2020, anchal, DailyProgress, FSS, FSS Diagnostics - SN010 (South RFPD) Notch Improvement Attempt SN010.jpgSN010_D980454-00-C.pdfSN010_Notch_Improvement_Attempt.pdf
                Reply  Thu Jan 16 20:18:44 2020, anchal, DailyProgress, FSS, FSS Diagnostics - SN010 (South RFPD) Notch Improved SN010.jpgSN010_D980454-00-C.pdfSN010_Notch_Improved.pdf
                   Reply  Mon Jan 27 18:29:19 2020, anchal, DailyProgress, FSS, FSS Diagnostics - RFPD RF Out Spectrum FSS_RFPD_RFOut_Spectrum.pdf
Message ID: 2514     Entry time: Thu Jan 16 20:18:44 2020     In reply to: 2513     Reply to this: 2519
Author: anchal 
Type: DailyProgress 
Category: FSS 
Subject: FSS Diagnostics - SN010 (South RFPD) Notch Improved 

Circuit changes (Updated Schematic)

I made the following changes to SN009:

  • R1   ->  200 Ohms
  • R2   ->  10 Ohms
  • Ln   -> 470 nH r. (Coilcraft 1206CS-471X_L_)
  • Replaced U1, MAX4107.

Updated Photo


Measurements

  • I used the setup in 40m to take light transfer function of the modified SN010.
  • Method described in CTN:2247 was used.

Inference

  • The ratio of resonance peak at 1-Omega (37 MHz) to 2-Omega (74 MHz) has been improved from about 20 to about 230.
  • Also, the measurement through test port and using lasers is still not matching up. This could mean that my estimation of DC transimpedance is wrong.
  • In the last plot, I tried multiplying the measurements of the result by a factor of 3.56 and it makes the transimpedance measurements match with test port measurements.
  • This shows there is a factoring error in my estimation of dc transimpedance or maybe the whole thing is damped down when light falls on the photodiode instead of test signal.
  • But the oscillation peak problem has been resolved. So the photodiode is good to be used. Minor tweaks might be required to see why test port estimation is different from laser measurement.

Data


Parallel relevant threads:

CTN:2516 : Notch improved on FSS RFPDs. Gain values increased as well.

CTN:2517: TTFSS OLTFs with Maximum Gain

Attachment 2: SN010_D980454-00-C.pdf  868 kB  | Hide | Hide all
SN010_D980454-00-C.pdf SN010_D980454-00-C.pdf
Attachment 3: SN010_Notch_Improved.pdf  99 kB  | Hide | Hide all
SN010_Notch_Improved.pdf SN010_Notch_Improved.pdf SN010_Notch_Improved.pdf